Evaluation of the Nova StatStrip glucometer in a pediatric hospital setting.

M G Lockyer1, K Fu1, R M Edwards2, L Collymore2, J Thomas2, T Hill2, S Devaraj3.   

Abstract

CONTEXT: Routine point-of-care (POC) glucose monitoring in the pediatric setting has become increasingly important, both for assessing hypoglycemia as well as hyperglycemia. A reliable and precise system is required to monitor pediatric patients.
OBJECTIVES: The aim of this study was to evaluate the Nova Biomedical StatStrip POC glucometer against the Roche ACCU-CHEK Inform in lieu of our currently used LifeScanSureStepFlexx POC glucose analyzer. DESIGN AND METHODS: Intra-assay and inter-assay precision, linearity, correlation and interference studies were performed as per the NCCLS criteria. An analysis of 37 pediatric samples across the linearity ranges of all the meters was used to assess concordance between the systems.
RESULTS: The Nova StatStrip glucometer demonstrated an excellent coefficient of variation (<5%) for glucose across the entire analytical measurement range. The Nova StatStrip also had good concordance with the central laboratory (Bland-Altman plots r(2)=0.01), while Roche Inform had poorer correlation (Bland-Altman plots r(2)=0.46). We also evaluated the effect of hematocrit (20-60%) and maltose (100-500mg/dL) on the Nova StatStrip analyzer and demonstrated that there is little to no interference by either.
CONCLUSIONS: The Nova StatStrip system gave the best performance with acceptable imprecision, good correlation, and minimal to no interference from hematocrit levels or maltose. The Nova StatStrip is a satisfactory replacement for our POC glucometer system and, additionally, provides results in less time (just 6 s) utilizing a lower amount of blood with the advantage of being immediately interfaced to our laboratory information systems.
Copyright © 2014. Published by Elsevier Inc.

Entities:  

Keywords:  Glucometer; Glucose monitoring; Hyperglycemia; Hypoglycemia; Neonate; Newborn; Pediatrics; Point-of-care testing

Mesh:

Substances:

Year:  2014        PMID: 24457064     DOI: 10.1016/j.clinbiochem.2014.01.004

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  4 in total

1.  Comparison of 2 portable human glucometers for the measurement of blood glucose concentration in White New Zealand rabbits.

Authors:  Kassy G Silva; Isabella Rotta; Leandro B Costa; Cristina S Sotomaior
Journal:  J Vet Diagn Invest       Date:  2020-07-29       Impact factor: 1.279

Review 2.  Point-of-care diagnostics to improve maternal and neonatal health in low-resource settings.

Authors:  Catherine E Majors; Chelsey A Smith; Mary E Natoli; Kathryn A Kundrod; Rebecca Richards-Kortum
Journal:  Lab Chip       Date:  2017-10-11       Impact factor: 6.799

3.  Glucose intolerance associated with hypoxia in people living at high altitudes in the Tibetan highland.

Authors:  Kiyohito Okumiya; Ryota Sakamoto; Yasuko Ishimoto; Yumi Kimura; Eriko Fukutomi; Motonao Ishikawa; Kuniaki Suwa; Hissei Imai; Wenling Chen; Emiko Kato; Masahiro Nakatsuka; Yoriko Kasahara; Michiko Fujisawa; Taizo Wada; Hongxin Wang; Qingxiang Dai; Huining Xu; Haisheng Qiao; Ri-Li Ge; Tsering Norboo; Norboo Tsering; Yasuyuki Kosaka; Mitsuhiro Nose; Takayoshi Yamaguchi; Toshihiro Tsukihara; Kazuo Ando; Tetsuya Inamura; Shinya Takeda; Masayuki Ishine; Kuniaki Otsuka; Kozo Matsubayashi
Journal:  BMJ Open       Date:  2016-02-23       Impact factor: 2.692

4.  A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting.

Authors:  Güzin Aykal; Ayşenur Yegin; Özgür Tekeli; Necat Yilmaz
Journal:  Biochem Med (Zagreb)       Date:  2016       Impact factor: 2.313

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.