Rosana Martins Ferreira de Carvalho1, Nilton Mazzer2, Claudio Henrique Barbieri2.
Abstract
OBJECTIVE: To evaluate the intra- and inter-examiner reliability and reproducibility of goniometry in relation to photogrammetry of hand, comparing the angles of thumb abduction, PIP joint flexion of the II finger and MCP joint flexion of the V finger.Entities:
Keywords: Goniometry; Hands; Photogrammetry; Reproducibility of tests
Year: 2012 PMID: 24453594 PMCID: PMC3718433 DOI: 10.1590/S1413-78522012000300003
Source DB: PubMed Journal: Acta Ortop Bras ISSN: 1413-7852 Impact factor: 0.513
Figure 1Wax hand in the position in which the photographic record was taken.
Figure 2Goniometer used - CARCI, Americanópolis, São Paulo, adapted at the Precision Workshop of USP.
Figure 3Angle of palmar abduction of thumb, using CorelDraw X3 software.
Figure 4Angle of flexion of the proximal interphalangeal joint of the second finger using ALCimagem 2.1 software.
ICC values for the interexaminer angular measurements obtained in the three groups 1 (scholars), 2 (physiotherapists), 3 (hand therapists).
| Group 1 | 0.995 | C.I. 95%(0.988-0.999) | Excellent |
| Group 2 | 0.995 | C.I. 95%(0.987-0.999) | Excellent |
| Group 3 | 0.992 | C.I. 95%(0.982-0.998) | Excellent |
ICC: Intraclass correlation coefficient; C.I.: Confidence interval.
ICC values for the interexaminer angular measurements obtained for each angle: 1 (palmar abduction of thumb), 2 (flexion of the proximal interphalangeal joint of the second finger), 3. (flexion of the metacarpophalangeal joint of the fifth finger).
| Angle 1 | 0.903 | C.I. 95%(0.658-0.997) | Excellent |
| Angle 2 | 0.701 | C.I. 95%(0.0-0.992) | Acceptable |
| Angle 3 | 0.869 | C.I. 95%(0.576-0.996) | Good |
ICC: Intraclass correlation coefficient; C.I.: Confidence interval.
ICC values for the interexaminer angular measurements obtained in the three methods 1 (goniometer), 2 (Corel Draw photogrammetry), 3. (ALCimagem photogrammetry).
| Method 1 | 0. 998 | C.I. 95% (0. 988-0. 999) | Excellent |
| Method 2 | 0. 998 | C.I. 95% (0. 987-0. 999) | Excellent |
| Method 3 | 0.999 | C.I. 95% (0. 982-0. 998) | Excellent |
ICC: Intraclass correlation coefficient; C.I.: Confidence interval.
ICC and standard deviation values for the intraexaminer angular measurements obtained for each group 1. (scholars), 2. (physiotherapists), 3. (hand therapists).
| Group 1 | 0.986 | 0.022 |
| Group 2 | 0.974 | 0.033 |
| Group 3 | 0.993 | 0.003 |
ICC: Intraclass correlation coefficient.
Figure 5Mean ICC values between the methods for each angle.
Figure 6Mean standard deviation values between the methods for each angle.
Comparison between the methods 1 (goniometry), 2 (Corel Draw photogrammetry) and 3 (ALCimagem photogrammetry) in angle 1 (palmar abduction of the thumb) for group 1 (scholars).
| Method 1 | Mean 44.37 |
| SD 4.61 | |
| Method 2 | Mean 39.13 |
| SD 4.61 | |
| Method 3 | Mean 37.91 |
| SD 2.91 | |
SD: Standard deviation.
Comparison between the methods 1 (goniometry), 2 (Corel Draw photogrammetry) and 3 (ALCimagem photogrammetry) in angle 1 (palmar abduction of the thumb) for group 2 (physiotherapists).
| Method 1 | Mean 41.00 |
| SD 5.90 | |
| Method 2 | Mean 38.80 |
| SD 6.28 | |
| Method 3 | Mean 38.83 |
| SD 5.49 | |
SD: Standard deviation.
Comparison between the methods 1 (goniometry), 2 (Corel Draw photogrammetry) and 3 (ALCimagem photogrammetry) in angle 1 (palmar abduction of the thumb) for group 3 (hand therapists).
| Method 1 | Mean 44.33 |
| SD 9.54 | |
| Method 2 | Mean 33.37 |
| SD 8.34 | |
| Method 3 | Mean 31.28 |
| SD 6.33 | |
SD: Standard deviation.
Comparison between the methods 1 (goniometry), 2 (Corel Draw photogrammetry) and 3 (ALCimagem photogrammetry) in angle 2 (flexion of the proximal interphalangeal of the second finger) for group 1 (scholars).
| Method 1 | Mean 32.33 |
| SD 2.28 | |
| Method 2 | Mean 29.87 |
| SD 1.91 | |
| Method 3 | Mean 29.51 |
| SD 1.64 | |
SD: Standard deviation.
Comparison between the methods 1 (goniometry), 2 (Corel Draw photogrammetry) and 3 (ALCimagem photogrammetry) in angle 2 (flexion of the proximal interphalangeal of the second finger) for group 2 (physiotherapists).
| Method 1 | Mean 31.37 |
| SD 3.11 | |
| Method 2 | Mean 30.03 |
| SD 3.79 | |
| Method 3 | Mean 30.67 |
| SD 3.92 | |
SD: Standard deviation.
Comparison between the methods 1 (goniometry), 2 (Corel Draw photogrammetry) and 3 (ALCimagem photogrammetry) in angle 2 (flexion of the proximal interphalangeal of the second finger) for group 3 (hand therapists).
| Method 1 | Mean 30.13 |
| SD 2.12 | |
| Method 2 | Mean 29.03 |
| SD 2.6 | |
| Method 3 | Mean 28.98 |
| SD 2.78 | |
SD: Standard deviation.
Comparison between the methods 1 (goniometry), 2 (Corel Draw photogrammetry), 3 (ALCimagem photogrammetry) in angle 3 (flexion of the metacarpophalangeal of the fifth finger) for group 1 (scholars).
| Method 1 | Mean (69.37 |
| SD 6.00 | |
| Method 2 | Mean 71.73 |
| SD 2.71 | |
| Method 3 | Mean 75.14 |
| SD 6.22 | |
SD: Standard deviation.
Comparison between the methods 1 (goniometry), 2 (Corel Draw photogrammetry), 3 (ALCimagem photogrammetry) in angle 3 (flexion of the metacarpophalangeal of the fifth finger) for group 2 (physiotherapists).
| Method 1 | Mean 73.40 |
| SD 4.20 | |
| Method 2 | Mean 75.13 |
| SD 4.35 | |
| Method 3 | Mean 76.83 |
| SD 4.97 | |
SD: Standard deviation.
Comparison between the methods 1 (goniometry), 2 (Corel Draw photogrammetry), 3 (ALCimagem photogrammetry) in angle 3 (flexion of the metacarpophalangeal of the fifth finger) for group 3 (hand therapists).
| Method 1 | Mean 71.83 |
| SD 5.67 | |
| Method 2 | Mean 72.67 |
| SD 4.57 | |
| Method 3 | Mean 77.18 |
| SD 5.72 | |
SD: Standard deviation.
Comparison between the groups 1 (scholars), 2 (physiotherapists), 3 (hand therapists) in angle 1 (palmar abduction of the thumb) for method 1 (goniometry).
| Group 1 | Mean 44.37 |
| SD 4.61 | |
| Group 2 | Mean 41.00 |
| SD 5.90 | |
| Group 3 | Mean 44.33 |
| SD 9.54 | |
SD: Standard deviation.
Comparison between the groups 1 (scholars), 2 (physiotherapists) and 3 (hand therapists) in angle 1 (palmar abduction of the thumb) for method 2 (Corel Draw photogrammetry).
| Group 1 | Mean 39.13 |
| SD 5.97 | |
| Group 2 | Mean 38.80 |
| SD 6.28 | |
| Group 3 | Mean 33.37 |
| SD 8.34 | |
SD: Standard deviation.
Comparison between the groups 1 (scholars), 2 (physiotherapists) and 3 (hand therapists) in angle 1 (palmar abduction of the thumb) for method 3 (ALCimagem photogrammetry).
| Group 1 | Mean 37.91 |
| SD 2.91 | |
| Group 2 | Mean 38.83 |
| SD 5.49 | |
| Group 3 | Mean 31.28 |
| SD 6.3 | |
SD: Standard deviation.
Comparison between the groups 1 (scholars), 2 (physiotherapists) and 3 (hand therapists) in angle 2 (flexion of the proximal interphalangeal of the second finger) for method 1 (goniometry).
| Group 1 | Mean 32.33 |
| SD 2.28 | |
| Group 2 | Mean 31.37 |
| SD 3.11 | |
| Group 3 | Mean 30.13 |
| SD 2.12 | |
SD: Standard deviation.
Comparison between the groups 1 (scholars), 2 (physiotherapists) and 3 (hand therapists) in angle 2 (flexion of the proximal interphalangeal of the second finger) for method 2 (Corel Draw photogrammetry®)..
| Group 1 | Mean 29.87 |
| SD 1.91 | |
| Group 2 | Mean 30.03 |
| SD 3.79 | |
| Group 3 | Mean 29.03 |
| SD 2.76 | |
SD: Standard deviation.
Comparison between the groups 1 (scholars), 2 (physiotherapists) and 3 (hand therapists) in angle 2 (flexion of the proximal interphalangeal of the second finger) for method 3 (ALCimagem photogrammetry®).
| Group 1 | Mean 29.51 |
| SD 1.64 | |
| Group 2 | Mean 30.67 |
| SD 3.92 | |
| Group 3 | Mean 28.98 |
| SD 2.78 | |
SD: Standard deviation.
Comparison between the groups 1 (scholars), 2 (physiotherapists) and 3 (hand therapists) in angle 3 (flexion of the metacarpophalangeal of the fifth finger) for method 1 (goniometry).
| Group 1 | Mean 69.37 |
| SD 6.00 | |
| Group 2 | Mean 73.40 |
| SD 4.20 | |
| Group 3 | Mean 71.83 |
| SD 5.67 | |
SD: Standard deviation.
Comparison between the groups 1 (scholars), 2 (physiotherapists) and 3 (hand therapists) in angle 3 (flexion of the metacarpophalangeal of the fifth finger) for method 2 (Corel Draw photogrammetry®).
| Group 1 | Mean 71.73 |
| SD 2.71 | |
| Group 2 | Mean 75.13 |
| SD 4.35 | |
| Group 3 | Mean 72.67 |
| SD 4.57 | |
SD - Standard deviation.
Comparison between the groups 1 (scholars), 2 (physiotherapists) and 3 (hand therapists) in angle 3 (flexion of the metacarpophalangeal of the fifth finger) for method 3 (ALCimagem photogrammetry®).
| Group 1 | Mean 75.14 |
| SD 6.22 | |
| Group 2 | Mean 76.83 |
| SD 4.97 | |
| Group 3 | Mean 77.18 |
| SD 5.72 | |
SD: Standard deviation.
Mean ICC and standard deviation values for the intraexaminer angular measurements obtained for each angle 1 (palmar abduction of thumb), 2 (flexion of the proximal interphalangeal joint of the second finger), 3 (flexion of the metacarpophalangeal joint of the fifth finger) in each group 1 (scholars), 2 (physiotherapists), 3 (hand therapists).
| Angle 1 | ICC 0.724 | ICC 0.558 | ICC 0.871 | ICC 0.717 |
| SD 0.290 | SD 0.435 | SD 0.203 | SD 0.338 | |
| Angle 2 | ICC 0.494 | ICC 0.610 | ICC 0.446 | ICC 0.517 |
| SD 0.414 | SD 0.317 | SD 0. 360 | SD 0. 360 | |
| Angle 3 | ICC 0.660 | ICC 0.500 | ICC 0.598 | ICC 0.586 |
| SD 0.322 | SD 0.363 | SD 0. 322 | SD 0.331 | |
ICC: Intraclass correlation coefficient; SD: Standard deviation.
ICC and standard deviation values for the intraexaminer angular measurements obtained for each method 1 (goniometry), 2 (Corel Draw photogrammetry), 3 (ALCimagem photogrammetry) in each group 1 (scholars), 2 (physiotherapists), 3 (hand therapists) and mean ICC of each method for each group.
| Method 1 | ICC 0.996 | ICC 0. 925 | ICC 0.995 | ICC 0.972 |
| SD 0.003 | SD 0. 213 | SD 0.003 | SD 0.124 | |
| Method 2 | ICC 0.992 | ICC 0.993 | ICC 0.991 | ICC 0.992 |
| SD 0.009 | SD 0.006 | SD 0. 006 | SD 0. 007 | |
| Method 3 | ICC 0.987 | ICC 0.978 | ICC 0.997 | ICC 0.987 |
| SD 0.033 | SD 0.039 | SD 0. 002 | SD 0.030 | |
ICC: Intraclass correlation coefficient; SD: Standard deviation.