Literature DB >> 24453573

Morphological evaluation of eccentric sets guide-plates of dcp-l 4.5 mm.

José Humberto de Souza Borges1, Antonio Lisboa da Silva Filho1, Francisco Pereira Neto1, Walter Rodrigo Daher1, Alessandro Queiroz de Mesquita1, Anderson Freitas1.   

Abstract

OBJECTIVE: To carry out isolated and comparative evaluations of the measurements of the set eccentric guide plates used in 4.5mm surgical implants, and to determine the effect of these measurements on compression strength.
METHODS: Four eccentric guides, four large dynamic compression plates (L-DCP) from four local manufacturers, and a Vonder(®) 200 mm caliper brand were used. Five standard parameter measurements were created for the set eccentric guide-plate, which were identified as A to E. Four sets were made, using materials of the same factory, and identified as groups I to IV. The analyses were performed by measuring all the parameters from a ventral view of the plate, with the eccentric guide placed in the plate hole.
RESULTS: Groups I and II showed the same values for all the parameters. All the groups showed the same measurements for E = 8.15 e B = 3.60. Group III: A = 8.10mm, C = 3.25mm, D = 1.25mm. Group IV: A = 7.00mm, C = 3.10mm, D = 0.30mm. Maximum compression force was (F Max.): Group I 80.58 N, Group. II: F Max. 81.63 N, Group. III: F Max. 36.32N, Group. IV: F Max. 37.52N.
CONCLUSION: The measurements evaluated show a lack of standardization in the manufacture of orthopedic instruments and its effects on the values for compression strength. LEVEL OF EVIDENCE: Level III, analytical study.

Entities:  

Keywords:  Biomechanics; Bone plates; Bone screws; Fracture fixation, internal/methods

Year:  2012        PMID: 24453573      PMCID: PMC3718412          DOI: 10.1590/S1413-78522012000100002

Source DB:  PubMed          Journal:  Acta Ortop Bras        ISSN: 1413-7852            Impact factor:   0.513


  8 in total

1.  The limited contact dynamic compression plate (LC-DCP).

Authors:  S M Perren; K Klaue; O Pohler; M Predieri; S Steinemann; E Gautier
Journal:  Arch Orthop Trauma Surg       Date:  1990       Impact factor: 3.067

Review 2.  Has locked plating completely replaced conventional plating?

Authors:  Michael J Gardner; David L Helfet; Dean G Lorich
Journal:  Am J Orthop (Belle Mead NJ)       Date:  2004-09

3.  The compressive properties of single osteons.

Authors:  A Ascenzi; E Bonucci
Journal:  Anat Rec       Date:  1968-07

4.  A dynamic compression plate.

Authors:  S M Perren; M Russenberger; S Steinemann; M E Müller; M Allgöwer
Journal:  Acta Orthop Scand Suppl       Date:  1969

5.  Effects of compression and cyclical loading on fracture healing--a quantitative biomechanical study.

Authors:  A A White; M M Panjabi; W O Southwick
Journal:  J Biomech       Date:  1977       Impact factor: 2.712

6.  The healing of experimental fractures by compression osteosynthesis. II. Morphometric and chemical analysis.

Authors:  P Paavolainen; R Penttinen; P Slätis; E Karaharju
Journal:  Acta Orthop Scand       Date:  1979-08

7.  The healing of experimental fractures by compression osteosynthesis. I. Torsional strength.

Authors:  P Paavolainen; P Slätis; E Karaharju; T Holmström
Journal:  Acta Orthop Scand       Date:  1979-08

8.  A mechanical comparison of the dynamic compression plate, limited contact-dynamic compression plate, and point contact fixator.

Authors:  T Miclau; A Remiger; S Tepic; R Lindsey; T McIff
Journal:  J Orthop Trauma       Date:  1995-02       Impact factor: 2.512

  8 in total