Literature DB >> 24450393

Impact of contrast agent viscosity on coronary balloon deflation times: bench testing results.

Owen Mogabgab1, Vishal G Patel, Tesfaldet T Michael, Anna Kotsia, George Christopoulos, Subhash Banerjee, Emmanouil S Brilakis.   

Abstract

OBJECTIVES: To assess the impact of viscosity on angioplasty balloon deflation times.
BACKGROUND: Lower contrast viscosity could result in more rapid coronary balloon deflation times.
METHODS: We performed a bench comparison of coronary balloon deflation times using 2 contrast agents with different viscosity (ioxaglate and iodixanol), 3 contrast dilutions, and 2 inflation syringe filling volumes. Ten identical pairs of coronary angioplasty balloons were used to conduct each comparison after balloon inflation to 12 atmospheres. Simultaneous deflations were performed under cineangiography. The time to full contrast extraction and the area of contrast remaining after 5 seconds of deflation (quantified by opaque pixel count) were compared between groups.
RESULTS: The mean time to full contrast extraction during balloon deflation was 8.3 ± 2.5 seconds for ioxaglate (lower viscosity) versus 10.1 ± 2.9 seconds for iodixanol (higher viscosity) (17.4% decrease, P = 0.005), with a 35.6% (P = 0.004) reduction in contrast area at 5 seconds. Compared to 1:1 ioxaglate-saline mixture, 1:2 and 1:3 ioxaglate/saline mixes resulted in 26.7% (P < 0.001) and 39.0% (P < 0.001) reduction in mean balloon deflation time, respectively, but at the expense of decreased balloon opacity. Filling the inflation syringe with 5 versus 15 ml of contrast/saline solution was associated with 7.5% decrease in balloon deflation time (P = 0.005), but no difference in contrast area at 5 seconds (P = 0.749).
CONCLUSIONS: Use of a lower viscosity contrast agent and higher contrast dilution significantly reduced coronary balloon deflation times, whereas use of lower syringe filling volume had a modest effect. Rapid coronary balloon deflation could improve the safety of interventional procedures.
© 2014 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24450393      PMCID: PMC3976709          DOI: 10.1111/joic.12097

Source DB:  PubMed          Journal:  J Interv Cardiol        ISSN: 0896-4327            Impact factor:   2.279


  12 in total

1.  Influence of radiographic contrast media viscosity to flow through coronary angiographic catheters.

Authors:  R Roth; M Akin; U Deligonul; M J Kern
Journal:  Cathet Cardiovasc Diagn       Date:  1991-04

2.  Iodixanol versus low-osmolar contrast media for prevention of contrast induced nephropathy: meta-analysis of randomized, controlled trials.

Authors:  Aaron M From; Firas J Al Badarin; Furman S McDonald; Brian J Bartholmai; Stephen S Cha; Charanjit S Rihal
Journal:  Circ Cardiovasc Interv       Date:  2010-07-20       Impact factor: 6.546

3.  Contrast-Induced Nephropathy (CIN) Consensus Working Panel: executive summary.

Authors:  Peter A McCullough; Fulvio Stacul; Christoph R Becker; Andy Adam; Norbert Lameire; James A Tumlin; Charles J Davidson
Journal:  Rev Cardiovasc Med       Date:  2006       Impact factor: 2.930

4.  Coronary intervention with 4-French catheters.

Authors:  Satoshi Takeshita; Shinji Tanaka; Shigeru Saito
Journal:  Catheter Cardiovasc Interv       Date:  2010-04-01       Impact factor: 2.692

5.  Effect of viscosity and iodine concentration of nonionic radiographic contrast media on coronary arteriography in patients.

Authors:  M J Kern; R A Roth; F V Aguirre; G Beauman; R Vogel
Journal:  Am Heart J       Date:  1992-01       Impact factor: 4.749

6.  Heating contrast media: role in contemporary angiography.

Authors:  R D Halsell
Journal:  Radiology       Date:  1987-07       Impact factor: 11.105

Review 7.  Structure-function relationships of iodinated contrast media and risk of nephrotoxicity.

Authors:  P Stratta; M Quaglia; A Airoldi; S Aime
Journal:  Curr Med Chem       Date:  2012       Impact factor: 4.530

8.  Percutaneous coronary intervention using a virtual 3-Fr guiding catheter.

Authors:  Shingo Mizuno; Satoshi Takeshita; Yoshio Taketani; Shigeru Saito
Journal:  Catheter Cardiovasc Interv       Date:  2010-06-01       Impact factor: 2.692

9.  Nephrotoxicity of iso-osmolar iodixanol compared with nonionic low-osmolar contrast media: meta-analysis of randomized controlled trials.

Authors:  Marc C Heinrich; Lothar Häberle; Volker Müller; Werner Bautz; Michael Uder
Journal:  Radiology       Date:  2009-01       Impact factor: 11.105

10.  Comparative rheology of low- and iso-osmolarity contrast agents at different temperatures.

Authors:  Jean Brunette; Rosaire Mongrain; Josep Rodés-Cabau; Eric Larose; Richard Leask; Olivier F Bertrand
Journal:  Catheter Cardiovasc Interv       Date:  2008-01-01       Impact factor: 2.692

View more
  1 in total

1.  Transcutaneous Puncture of an Undeflatable Coronary Angioplasty Balloon Catheter.

Authors:  Gregor Leibundgut; Christian Degen; Florian Riede
Journal:  Case Rep Cardiol       Date:  2018-09-03