An audit of half-count myocardial perfusion imaging using resolution recovery software.

Richard S Lawson1, Duncan White, Kuldip Nijran, Sarah C Cade, David O Hall, Bob Kenny, Andy Knight, Lefteris Livieratos, Anthony Murray, David Towey.   

Abstract

INTRODUCTION: The Nuclear Medicine Software Quality Group of the Institute of Physics and Engineering in Medicine has conducted a multicentre, multivendor audit to evaluate the use of resolution recovery software from several manufacturers when applied to myocardial perfusion data with half the normal counts acquired under a variety of clinical protocols in a range of departments. The objective was to determine whether centres could obtain clinical results with half-count data processed with resolution recovery software that were equivalent to those obtained using their normal protocols.
MATERIALS AND METHODS: Sixteen centres selected 50 routine myocardial perfusion studies each, from which the Nuclear Medicine Software Quality Group generated simulated half-count studies using Poisson resampling. These half-count studies were reconstructed using resolution recovery and the clinical reports compared with the original reports from the full-count data. A total of 769 patient studies were processed and compared.
RESULTS: Eight centres found only a small number of clinically relevant discrepancies between the two reports, whereas eight had an unacceptably high number of discrepancies. There were no significant differences in acquisition parameters between the two groups, although centres finding a high number of discrepancies were more likely to perform both rest and stress scans on normal studies.
CONCLUSION: Half of the participating centres could potentially make use of resolution recovery to reduce the administered activity for myocardial perfusion scans without changing their routine acquisition protocols. The other half could consider adjusting the reconstruction parameters used with their resolution recovery software if they wish to use reduced activity successfully.

Entities:  

Mesh:

Year:  2014        PMID: 24448215     DOI: 10.1097/MNM.0000000000000078

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  3 in total

1.  Simulating dose reduction for myocardial perfusion SPECT using a Poisson resampling method.

Authors:  Il-Hyun Kim; Su Jin Lee; Young-Sil An; So-Yeon Choi; Joon-Kee Yoon
Journal:  Nucl Med Mol Imaging       Date:  2021-08-13

2.  The usefulness of SwiftScan technology for bone scintigraphy using a novel anthropomorphic phantom.

Authors:  Takayuki Shibutani; Masahisa Onoguchi; Yuka Naoi; Hiroto Yoneyama; Takahiro Konishi; Ringo Tatami; Kenichi Nakajima
Journal:  Sci Rep       Date:  2021-01-29       Impact factor: 4.379

3.  Technical note: Partitioning of gated single photon emission computed tomography raw data for protocols optimization.

Authors:  Cleiton Cavalcante Queiroz; Marcos Antonio Dorea Machado; Antonio Augusto Brito Ximenes; Andre Gustavo Silva Pino; Eduardo Martins Netto
Journal:  J Appl Clin Med Phys       Date:  2021-12-17       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.