The effectiveness of different functional fibrinogen polymerization assays in eliminating platelet contribution to clot strength in thromboelastometry.

Christoph J Schlimp1, Cristina Solomon, Marco Ranucci, Gerald Hochleitner, Heinz Redl, Herbert Schöchl.   

Abstract

BACKGROUND: Viscoelastic tests such as functional fibrinogen polymerization assays (FFPAs) in thrombelastography (TEG®) or thromboelastometry (ROTEM®) measure clot elasticity under platelet inhibition. Incomplete platelet inhibition influences maximum clot firmness (MCF) of FFPAs. We compared the ability of existing and newly developed FFPAs to eliminate the platelet contribution to clot strength.
METHODS: MCF of whole blood (WB), platelet-rich plasma (PRP), and platelet-poor plasma samples was recorded using a ROTEM device with different FFPAs, including the TEG functional fibrinogen test (FFTEG) and different ROTEM-based assays: the standard fib-tem reagent (FIBTEM), a lyophilized single-portion reagent fib-tem S (FIBTEM-S), a newly developed reagent FIBTEM PLUS, as well as FIBTEM or the standard extrinsic activation reagent ex-tem® (EXTEM) combined with 10-μg abciximab (FIBTEM-ABC/EXTEM-ABC).
RESULTS: In WB (platelet count [mean ± SD], 183 ± 37 × 10/μL; plasma fibrinogen concentration, 2.49 ± 0.58 g/L), FFTEG and EXTEM-ABC showed higher MCF (15.7 ± 2.8 mm) than FIBTEM or FIBTEM-S (11.4 ± 3.3 mm, P < 0.001), whereas FIBTEM-ABC and FIBTEM PLUS resulted in lower MCF (9.3 ± 2.8 mm, P < 0.001). In 2 different PRP samples, with platelet counts of 407 ± 80 × 10/μL and 609 ± 127 × 10/μL, FIBTEM-ABC and FIBTEM PLUS reduced platelet contribution to clot strength within 95% confidence interval limits of -1.4 to 0.1 mm and -1.2 to 0.4 mm, respectively. Using all FFPAs it was observed that the Pearson correlation coefficient between plasma fibrinogen concentration and WB MCF was high (range, 0.75-0.93) and significant, regardless of the underlying platelet inhibiting component. Evaluating differences in the interception of regression lines by using analysis of covariance, we compared platelet-poor plasma and both PRP samples within the same assays and found that in contrast to the FIBTEM-ABC and FIBTEM PLUS assays, the FFTEG, EXTEM-ABC, FIBTEM, and FIBTEM-S methods still detected residual platelet activity and grossly overestimated fibrin clot strength in samples with high platelet counts.
CONCLUSIONS: FFPAs based solely on glycoprotein-IIb/IIIa inhibition, such as FFTEG or EXTEM-ABC, are less effective than cytochalasin D-based assays, such as FIBTEM or FIBTEM-S, at inhibiting the platelet component of clot strength. The FIBTEM PLUS assay, and the combination of FIBTEM and abciximab, sufficiently inhibits platelet contribution to clot elasticity. The combination of a glycoprotein-IIb/IIIa receptor blocker and cytochalasin D allows evaluation of functional fibrinogen polymerization without platelet "noise." In a clinical setting, the significance of potent platelet inhibition ensures a more accurate assessment of MCF and therefore the need for fibrinogen supplementation therapy. Further studies are necessary to investigate the application and impact of these tests in a clinical situation.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24445628     DOI: 10.1213/ANE.0000000000000058

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  17 in total

Review 1.  Platelet-mimetic strategies for modulating the wound environment and inflammatory responses.

Authors:  Seema Nandi; Ashley C Brown
Journal:  Exp Biol Med (Maywood)       Date:  2016-05-02

2.  Multiplate and TEG platelet mapping in a population of severely injured trauma patients.

Authors:  M J George; J Burchfield; B MacFarlane; Y-W W Wang; J C Cardenas; N J White; B S Gill; C E Wade
Journal:  Transfus Med       Date:  2017-09-15       Impact factor: 2.019

3.  Fibrinogen measurement in liver disease: validation of the functional fibrinogen thromboelastography assay and a novel mathematical predictive model.

Authors:  Kirsty Rizzo; Kevin Vella; Daniel Zammit; Peter Gatt; Charlie Grima; Monique Borg Inguanez; Jurgen Gerada; Pierre Ellul; Mario Vassallo; Neville Azzopardi; James Pocock; Alex Gatt
Journal:  Blood Transfus       Date:  2018-11-09       Impact factor: 3.443

4.  Viscoelastic measurements of platelet function, not fibrinogen function, predicts sensitivity to tissue-type plasminogen activator in trauma patients.

Authors:  H B Moore; E E Moore; M P Chapman; E Gonzalez; A L Slaughter; A P Morton; A D'Alessandro; K C Hansen; A Sauaia; A Banerjee; C C Silliman
Journal:  J Thromb Haemost       Date:  2015-09-22       Impact factor: 5.824

5.  Utility of Viscoelastic Assays Beyond Coagulation: Can Preoperative Thrombelastography Indices Predict Tumor Histology, Nodal Disease, and Resectability in Patients Undergoing Pancreatectomy?

Authors:  Hunter B Moore; Alessandro Paniccia; Peter J Lawson; Robert J Torphy; Trevor L Nydam; Ernest E Moore; Martin D McCarter; Richard D Schulick; Barish H Edil
Journal:  J Am Coll Surg       Date:  2018-03-30       Impact factor: 6.113

6.  Assays of different aspects of haemostasis - what do they measure?

Authors:  Nahreen Tynngård; Tomas L Lindahl; Sofia Ramström
Journal:  Thromb J       Date:  2015-02-05

7.  Comparison of fibrin-based clot elasticity parameters measured by free oscillation rheometry (ReoRox ®) versus thromboelastometry (ROTEM ®).

Authors:  Cristina Solomon; Herbert Schöchl; Marco Ranucci; Ulf Schött; Christoph J Schlimp
Journal:  Scand J Clin Lab Invest       Date:  2015-01-19       Impact factor: 1.713

8.  Rapid measurement of fibrinogen concentration in whole blood using a steel ball coagulometer.

Authors:  Christoph J Schlimp; Anna Khadem; Anton Klotz; Cristina Solomon; Gerald Hochleitner; Martin Ponschab; Heinz Redl; Herbert Schöchl
Journal:  J Trauma Acute Care Surg       Date:  2015-04       Impact factor: 3.313

9.  The use of frozen plasma samples in thromboelastometry.

Authors:  Christian Schoergenhofer; Nina Buchtele; Michael Schwameis; Johann Bartko; Bernd Jilma; Petra Jilma-Stohlawetz
Journal:  Clin Exp Med       Date:  2017-02-16       Impact factor: 3.984

Review 10.  Questions about COVID-19 associated coagulopathy: possible answers from the viscoelastic tests.

Authors:  Vittorio Pavoni; Lara Gianesello; Maddalena Pazzi; Pietro Dattolo; Domenico Prisco
Journal:  J Clin Monit Comput       Date:  2021-07-15       Impact factor: 1.977