BACKGROUND: Intraoperatively fabricated polymethylmethacrylate (PMMA) implants based on computer-designed moulds were used to improve cosmetic results after hard tissue replacement. To assess the implant's cosmetic and functional results we performed both subjective and objective assessments.
METHODS: This retrospective analysis was performed using a cohort of 28 patients who received PMMA implants between February 2009 and March 2012. The cosmetic and functional results were assessed using a patient questionnaire. Furthermore an objective volumetric subtraction score (0-100) was applied and implant thickness, as well as gaps and tiers, were measured.
RESULTS: Patients mainly judged their cosmetic result as "good". Two of the 28 patients found their cosmetic result unfavourable. The functional result and stability was mainly judged to be good. Measurements of implant thickness showed a very high correlation with the thickness of the contralateral bone. Volumetric subtraction led to a median quality of 80 on a scale from 0 to 100. Median gaps around the margins of the implant were 1.5 mm parietally, 1.7 mm frontally and 3.5 mm fronto-orbitally, and median tiers were 1.2 mm, 0 mm and 0 mm respectively. The overall rate of surgical revisions was 10.7 % (three patients). Two patients suffered from wound healing disturbances (7.1 %). The overall complication rate was comparable to other reports in the literature.
CONCLUSIONS: Implantation of intraoperatively fabricated patient-specific moulded implants is a cost-effective and safe technique leading to good clinical results with a low complication rate.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24442601     DOI: 10.1007/s00701-013-1977-5

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  3 in total

1.  Efficacy of 3D-Printed Titanium Mesh-Type Patient-Specific Implant for Cranioplasty.

Authors:  Hong-Gyu Yoon; Yong Ko; Young-Soo Kim; Koang-Hum Bak; Hyoung-Joon Chun; Min-Kyun Na; Sook Yang; Hyeong-Joong Yi; Kyu-Sun Choi
Journal:  Korean J Neurotrauma       Date:  2021-09-09

2.  Craniofacial Reconstruction by a Cost-Efficient Template-Based Process Using 3D Printing.

Authors:  Bilal Msallem; Fabian Beiglboeck; Philipp Honigmann; Claude Jaquiéry; Florian Thieringer
Journal:  Plast Reconstr Surg Glob Open       Date:  2017-11-17

3.  Accuracy Assessment of Molded, Patient-Specific Polymethylmethacrylate Craniofacial Implants Compared to Their 3D Printed Originals.

Authors:  Dave Chamo; Bilal Msallem; Neha Sharma; Soheila Aghlmandi; Christoph Kunz; Florian M Thieringer
Journal:  J Clin Med       Date:  2020-03-19       Impact factor: 4.241

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.