| Literature DB >> 32204321 |
Dave Chamo1,2, Bilal Msallem1,2, Neha Sharma1,2, Soheila Aghlmandi3, Christoph Kunz1, Florian M Thieringer1,2.
Abstract
The use of patient-specific implants (PSIs) in craniofacial surgery is often limited due to a lack of expertise and/or production costs. Therefore, a simple and cost-efficient template-based fabrication workflow has been developed to overcome these disadvantages. The aim of this study is to assess the accuracy of PSIs made from their original templates. For a representative cranial defect (CRD) and a temporo-orbital defect (TOD), ten PSIs were made from polymethylmethacrylate (PMMA) using computer-aided design (CAD) and three-dimensional (3D) printing technology. These customized implants were measured and compared with their original 3D printed templates. The implants for the CRD revealed a root mean square (RMS) value ranging from 1.128 to 0.469 mm with a median RMS (Q1 to Q3) of 0.574 (0.528 to 0.701) mm. Those for the TOD revealed an RMS value ranging from 1.079 to 0.630 mm with a median RMS (Q1 to Q3) of 0.843 (0.635 to 0.943) mm. This study demonstrates that a highly precise duplication of PSIs can be achieved using this template-molding workflow. Thus, virtually planned implants can be accurately transferred into haptic PSIs. This workflow appears to offer a sophisticated solution for craniofacial reconstruction and continues to prove itself in daily clinical practice.Entities:
Keywords: 3D printing; PMMA; accuracy; additive manufacturing; craniofacial reconstruction; cranioplasty; molding; patient-specific implant
Year: 2020 PMID: 32204321 PMCID: PMC7141183 DOI: 10.3390/jcm9030832
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.241
Figure 13D printed skull models with the defect area shown. (a) cranial defect (CRD); (b) temporo-orbital defect (TOD).
Cone-beam computed tomography settings.
| Implant | Volume | kV | mA | Voxel |
|---|---|---|---|---|
| CRD-PSI | 17 × 13.5 cm3 | 90 | 5.0 | 300 µm |
| TOD-PSI | 10 × 10 cm3 | 90 | 5.0 | 180 µm |
Figure 2Comparison of the 3D printed templates (a,b, beige) with the patient-specific implants according to the n-point registration with five manually placed control points (c,d, purple), and the superimposition (e,f). Left: cranial template and PSI; right: temporo-orbital template and PSI.
Figure 3Descriptive data distribution illustrating the difference between the CRD-PSIs (models 1 to 10) and the CRD 3D printed template. (a) Mean difference ± SD (mm); (b) Median difference (Q1 to Q3) (mm).
Figure 4Descriptive data distribution illustrating the difference between the TOD-PSIs (models 1 to 10) and the TOD 3D printed template. (a) Mean difference ± SD (mm); (b) Median difference (Q1 to Q3) (mm).
Overall results (mm) of the CRD-PSIs and TOD-PSIs compared to the corresponding 3D printed template.
| Implant | Mean Difference ± SD | Median Difference (Q1 to Q3) |
|---|---|---|
| CRD-PSIs | 0.383 ± 0.547 | 0.334 (0.122 to 0.722) |
| TOD-PSIs | 0.228 ± 0.790 | −0.159 (−0.662 to 0.165) |
Quantitative evaluation of the CRD-PSIs and TOD-PSIs accuracy with respect to the root mean square (RMS) values (mm).
| Implant | RMS (mm) | Implant | RMS (mm) |
|---|---|---|---|
| CRD-PSI 01 | 1.128 | TOD-PSI 01 | 0.630 |
| CRD-PSI 02 | 0.701 | TOD-PSI 02 | 0.630 |
| CRD-PSI 03 | 0.562 | TOD-PSI 03 | 1.079 |
| CRD-PSI 04 | 0.469 | TOD-PSI 04 | 0.635 |
| CRD-PSI 05 | 0.836 | TOD-PSI 05 | 0.983 |
| CRD-PSI 06 | 0.644 | TOD-PSI 06 | 0.875 |
| CRD-PSI 07 | 0.510 | TOD-PSI 07 | 0.747 |
| CRD-PSI 08 | 0.535 | TOD-PSI 08 | 0.811 |
| CRD-PSI 09 | 0.528 | TOD-PSI 09 | 0.878 |
| CRD-PSI 10 | 0.585 | TOD-PSI 10 | 0.943 |
Left: CRD-PSIs vs. CRD 3D printed template; right: TOD-PSIs vs. TOD 3D printed template.
Figure 5Box plot illustrating the accuracy comparison with respect to the root mean square (RMS) values (mm) between the polymethylmethacrylate (PMMA) CRD-PSI and TOD-PSI test groups (● describes the statistical outlier, CRD-PSI 01).
Quantitative accuracy assessment for the median RMS values (mm) in the CRD-PSIs and TOD-PSIs.
| Implant | Median (Q1 to Q3) | |
|---|---|---|
| CRD-PSIs | 0.574 (0.528 to 0.701) | 0.028 |
| TOD-PSIs | 0.843 (0.635 to 0.943) |
* p-value calculated using Mann–Whitney U test; * p < 0.05 is considered statistically significant.
Figure 6Color-coded deviation maps within each test group after applying the best-fit method and generating a 3D comparison to evaluate the accuracy. CRD-PSI: (a) squamous (outer) surface; (c) cerebral (inner) surface; TOD-PSI: (b) squamous (outer) surface; (d) cerebral (inner) surface.