Outcomes of torsional microcoaxial phacoemulsification using tips with 30-degree and 45-degree aperture angles.

Firat Helvacioglu1, Sadik Sencan2, Celal Yeter2, Zeki Tunc2, O Murat Uyar2.   

Abstract

PURPOSE: To compare the safety and efficacy of Ozil Intelligent Phaco torsional microcoaxial phacoemulsification surgeries performed using 30-degree and 45-degree aperture angled tips.
SETTING: Maltepe University School of Medicine Department of Ophthalmology, Istanbul, Turkey.
DESIGN: Comparative case series.
METHODS: Eyes were assigned to 2.2 mm microcoaxial phacoemulsification using the torsional mode with a 45-degree (Group 1) or 30-degree (Group 2) aperture angled tip. A quick-chop surgical technique was used. The primary outcome measures were ultrasound time (UST), cumulative dissipated energy (CDE), longitudinal and torsional ultrasound (US) amplitudes, mean surgical time, mean balanced salt solution volume used, and mean central corneal thickness (CCT) changes.
RESULTS: The mean UST, CDE, and longitudinal and torsional US amplitudes were 58.21 seconds ± 33.81 (SD), 7.74 ± 6.23, 0.45 ± 0.30, and 26.30 ± 12.60 in Group 1 and 63.83 ± 23.42 seconds, 12.36 ± 6.75, 0.23 ± 0.26, and 44.65 ± 14.38 in Group 2, respectively. The mean CDE and torsional amplitudes were significantly lower in Group 1 (P=.002 and P=.001, respectively). The mean balanced salt solution volume was 73.30 ± 19.87 cc in Group 1 and 74.30 ± 19.44 cc in Group 2 (P=.821). The mean CCT change was 52.40 ± 38.08 μm and 99.35 ± 47.14 μm, respectively (P<.001).
CONCLUSION: Torsional phacoemulsification performed with a 45-degree aperture angled tip provided more effective lens removal with a lower CDE and less CCT change than a 30-degree aperture angled tip.
Copyright © 2014 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24440105     DOI: 10.1016/j.jcrs.2013.07.051

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

1.  Comparison of 45-degree Kelman and 45-degree balanced phaco tip designs in torsional microcoaxial phacoemulsification.

Authors:  Süleyman Demircan; Mustafa Ataş; Emre Göktaş; Burhan Başkan
Journal:  Int J Ophthalmol       Date:  2015-12-18       Impact factor: 1.779

2.  Efficacy of Balanced Torsional Phacoemulsification Tip for Cataract Surgery.

Authors:  Huri Sabur; Sayime Aydin Eroglu; Salih Sertac Azarsiz
Journal:  J Curr Ophthalmol       Date:  2022-04-16

3.  Results of endocapsular phacofracture debulking of hard cataracts.

Authors:  James A Davison
Journal:  Clin Ophthalmol       Date:  2015-07-10

4.  Effect of IOP based infusion system with and without balanced phaco tip on cumulative dissipated energy and estimated fluid usage in comparison to gravity fed infusion in torsional phacoemulsification.

Authors:  Praveen K Malik; Taru Dewan; Arun Kr Patidar; Ekta Sain
Journal:  Eye Vis (Lond)       Date:  2017-09-17

5.  Comparisons of the energy efficiency and intraocular safety of two torsional phacoemulsification tips.

Authors:  Yan-Xiu Sun; Rong Cao; Zi-Yuan Liu; Hua-Qin Xia; Yu-Jie Cen; Lu Gao; Dan-Dan Shi
Journal:  BMC Ophthalmol       Date:  2022-10-03       Impact factor: 2.086

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.