Literature DB >> 24440016

Effect of valve design on the stent internal diameter of a bioprosthetic valve: a concept of true internal diameter and its implications for the valve-in-valve procedure.

Vinayak N Bapat1, Rizwan Attia2, Martyn Thomas2.   

Abstract

The goal of this study was to provide a measurement of the true internal diameter (ID) of various surgical heart valves (SHV) to facilitate the valve-in-valve (VIV) procedure. During a VIV procedure, it is important to choose the right of the transcatheter heart valve (THV). Most users use the stent ID of an SHV to select the appropriate THV size. Echocardiography and computed tomography measurements are not yet standardized for measuring the ID of a variety of SHVs. Hence, we measured the true ID of SHV to assess the effect of valve design on the stent ID. Thirteen types of stented and 3 types of stentless valves were evaluated. True ID measurements were obtained using calipers and Hegar dilators. These were compared with the stent ID measurements. Fluoroscopy was used to confirm the impact of SHV designs on the true ID. Caliper measurements were found to be inaccurate and are hence not recommended. Hegar dilator measurements revealed a trend of reduction in stent ID. Porcine valves were most affected by their design, with reduction in the stent ID by at least 2 mm; pericardial valves with leaflets sutured inside the stent had the stent ID reduced by at least 1 mm, and SHV with leaflets sutured outside the stent had no effect on stent ID. In the majority of SHV designs, there is a reduction in the stent ID as a result of leaflet tissue. This is important in borderline sizes to avoid problems associated with oversizing and also to confirm suitability for the VIV procedure in the smaller label sizes of SHV.
Copyright © 2014 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Keywords:  heart valve; valve replacement; valve-in-valve

Mesh:

Year:  2014        PMID: 24440016     DOI: 10.1016/j.jcin.2013.10.012

Source DB:  PubMed          Journal:  JACC Cardiovasc Interv        ISSN: 1936-8798            Impact factor:   11.195


  25 in total

1.  Valve-in-Valve Transcatheter Valve Implantation in the Nonaortic Position.

Authors:  David N Ranney; Judson B Williams; Andrew Wang; Jeffrey G Gaca
Journal:  J Card Surg       Date:  2016-04-05       Impact factor: 1.620

Review 2.  On the Mechanics of Transcatheter Aortic Valve Replacement.

Authors:  Lakshmi P Dasi; Hoda Hatoum; Arash Kheradvar; Ramin Zareian; S Hamed Alavi; Wei Sun; Caitlin Martin; Thuy Pham; Qian Wang; Prem A Midha; Vrishank Raghav; Ajit P Yoganathan
Journal:  Ann Biomed Eng       Date:  2016-11-21       Impact factor: 3.934

3.  Bioprosthetic Valve Fracture During Valve-in-valve TAVR: Bench to Bedside.

Authors:  John T Saxon; Keith B Allen; David J Cohen; Adnan K Chhatriwalla
Journal:  Interv Cardiol       Date:  2018-01

4.  Efficacy and safety of transcatheter aortic valve implantation in patients with severe bicuspid aortic stenosis.

Authors:  Bo Fu; Qingliang Chen; Feng Zhao; Zhigang Guo; Nan Jiang; Xu Wang; Wei Wang; Jiange Han; Li Yang; Yanbo Zhu; Yanhe Ma
Journal:  Ann Transl Med       Date:  2020-07

5.  A Preliminary Study on the Usage of a Data-Driven Probabilistic Approach to Predict Valve Performance Under Different Physiological Conditions.

Authors:  Brennan J Vogl; Yousef M Darestani; Juan A Crestanello; Brian R Lindman; Mohamad A Alkhouli; Hoda Hatoum
Journal:  Ann Biomed Eng       Date:  2022-04-26       Impact factor: 3.934

Review 6.  Current state of transcatheter mitral valve implantation in bioprosthetic mitral valve and in mitral ring as a treatment approach for failed mitral prosthesis.

Authors:  Vratika Agarwal; Ryan K Kaple; Hetal H Mehta; Prabhjot Singh; Vinayak N Bapat
Journal:  Ann Cardiothorac Surg       Date:  2021-09

Review 7.  Technical pitfalls and tips for the valve-in-valve procedure.

Authors:  Vinayak Bapat
Journal:  Ann Cardiothorac Surg       Date:  2017-09

Review 8.  Challenges in valve-in-valve therapy.

Authors:  Alia Noorani; Rahee Radia; Vinayak Bapat
Journal:  J Thorac Dis       Date:  2015-09       Impact factor: 2.895

9.  Radiopaque preventive landmarks' placement during stentless bioprosthesis implantation.

Authors:  Niki Lama; Vasileios Patris; Dimitrios Fagkrezos; Petros Moschouris; Konstantinos Giakoumidakis; Charikleia Triantopoulou; Petros Maniatis; Mihalis Argiriou
Journal:  J Thorac Dis       Date:  2018-06       Impact factor: 2.895

10.  Transcatheter valve-in-valve implantation in degenerated aortic bioprostheses: are patients with small surgical bioprostheses at higher risk for unfavourable mid-term outcomes?

Authors:  Clarence Pingpoh; Holger Schroefel; Tanja Franz; Martin Czerny; Maximilian Kreibich; Martin Moser; Christoph Bode; Friedhelm Beyersdorf; Franz-Josef Neumann; Willibald Hochholzer; Matthias Siepe
Journal:  Ann Cardiothorac Surg       Date:  2020-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.