The evaluation and comparison of consecutive high speed centrifugation and LipoClear® reagent for lipemia removal.

Andrea Saracevic1, Nora Nikolac2, Ana-Maria Simundic2.   

Abstract

BACKGROUND: The aim of this study was to evaluate and compare the efficiency of high speed centrifugation and LipoClear® reagent for lipemia removal in plasma samples spiked with Intralipid®, for 26 biochemistry analytes.
MATERIALS AND METHODS: A plasma pool was collected. Aliquots of the pool were spiked with Intralipid® (final concentrations of 300mg/dL and 500mg/dL Intralipid®). The lipemia was removed from the aliquots by high speed centrifugation or LipoClear® reagent. 26 analytes were determined in native, lipemic plasma and in samples after lipemia removal. The bias from the concentration in the native sample was calculated for each parameter for Intralipid® concentrations, 300 and 500mg/dL of Intralipid®, respectively. Also, the recovery for each parameter after processing the samples using high speed centrifugation and LipoClear® was calculated. The biases and test recoveries were compared with the desirable specification for imprecision (DSI) according to Ricos available at the Wesgard's website. The bias and recovery for procalcitonin were compared with DSI according to Barassi and colleagues.
RESULTS: The bias of the spiked samples exceeded the DSI at 300mg/L Intralipid® for creatinine, glucose, total protein, iron and albumin; and for all previously mentioned parameters including CK-MB, sodium, potassium, chlorides, magnesium and ALP at concentration of 500mg/L Intralipid®. For the test recovery the DSI criteria were not met for calcium, total protein, sodium and chlorides after high speed centrifugation and for glucose, calcium, phosphates, magnesium, sodium, potassium, chlorides, ALP, GGT, CK-MB, total protein, albumin and troponin T after using LipoClear®.
CONCLUSIONS: LipoClear® is not suitable for lipemia removal from samples designated for glucose, sodium, potassium, chlorides, phosphates, magnesium, CK-MB, ALP, GGT, total protein, albumin, CRP and troponin T measurements. High speed centrifugation should be used for lipemia removal instead for glucose, potassium, phosphates, magnesium, CK-MB, ALP, GGT, albumin, CRP and TnT measurements.
Copyright © 2014 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bias; Centrifugation; Interference; Lipemia; Quality improvement

Mesh:

Substances:

Year:  2014        PMID: 24434304     DOI: 10.1016/j.clinbiochem.2014.01.001

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  9 in total

Review 1.  Pre-analytic variability in cardiovascular biomarker testing.

Authors:  Roberto Cemin; Massimo Daves
Journal:  J Thorac Dis       Date:  2015-10       Impact factor: 2.895

2.  Effects of marked hypertriglyceridemia and lipid clearance techniques on canine biochemistry testing.

Authors:  Carolina N Azevedo; Jonathan A Lidbury; Unity Jeffery
Journal:  J Vet Diagn Invest       Date:  2019-06-25       Impact factor: 1.279

3. 

Authors:  Imene Ben Jdidia; Kaouther Zribi; Meriam Boubaker; Amira Brahem; Mouna Sayadi; Marwa Tlijani; Zahra Saidani; Amani Cherif
Journal:  Can J Hosp Pharm       Date:  2021

Review 4.  Laboratory Diagnostics and Quality of Blood Collection.

Authors:  Gabriel Lima-Oliveira; Giuseppe Lippi; Gian Luca Salvagno; Geraldo Picheth; Gian Cesare Guidi
Journal:  J Med Biochem       Date:  2015-07-14       Impact factor: 3.402

5.  Preanalytical external quality assessment of the Croatian Society of Medical Biochemistry and Laboratory Medicine and CROQALM: finding undetected weak spots.

Authors:  Nora Nikolac; Jasna Lenicek Krleza; Ana-Maria Simundic
Journal:  Biochem Med (Zagreb)       Date:  2017-02-15       Impact factor: 2.313

6.  Removing Lipemia in Serum/Plasma Samples: A Multicenter Study.

Authors:  María José Castro-Castro; Beatriz Candás-Estébanez; Margarita Esteban-Salán; Pilar Calmarza; Teresa Arrobas-Velilla; Carlos Romero-Román; Miguel Pocoví-Mieras; José Ángel Aguilar-Doreste
Journal:  Ann Lab Med       Date:  2018-11       Impact factor: 3.464

7.  Comparative analysis of high CRP-levels in human blood using point-of-care and laboratory-based methods.

Authors:  Christiane Eckschlager; Karin Schwenoha; Caroline Roth; Barbara Bogner; Gertie Janneke Oostingh
Journal:  Pract Lab Med       Date:  2019-09-19

Review 8.  Lipemia: causes, interference mechanisms, detection and management.

Authors:  Nora Nikolac
Journal:  Biochem Med (Zagreb)       Date:  2014-02-15       Impact factor: 2.313

9.  Frequency and causes of lipemia interference of clinical chemistry laboratory tests.

Authors:  Sandhya Mainali; Scott R Davis; Matthew D Krasowski
Journal:  Pract Lab Med       Date:  2017-02-03
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.