Reproducibility and variability of transepidermal water loss measurement. Studies on the Servo Med Evaporimeter.

C W Blichmann, J Serup.   

Abstract

Recording technique and accuracy of the Servo Med EP1 evaporimeter were studied. The technical accuracy of this equipment was found high with a coefficient of variation of 3.3%. Measurements of transepidermal water loss (TEWL) were performed on the palm of the hand and on the flexor side of the forearm in 10 healthy volunteers. Continuous registrations were undertaken for a three-minute period and repeated immediately thereafter. Values reached a linear state 30 sec after application of the probe. The variation of TEWL in relation to time was determined in the same individuals. Measurements were repeated after 1 hour and 24 hours. In conclusion, though the TEWL as measured by the Servo Med evaporimeter shows a considerable interindividual variation, the intraindividual variation and the variation by time are relatively minor. We recommend that more detailed studies are based on paired observations, with recordings read 30 sec after application of the probe onto the skin.

Entities:  

Mesh:

Year:  1987        PMID: 2442932

Source DB:  PubMed          Journal:  Acta Derm Venereol        ISSN: 0001-5555            Impact factor:   4.437


  4 in total

1.  Variability in transepidermal water loss of the skin: evaluation of a method to assess susceptibility to contact dermatitis in epidemiological studies.

Authors:  H A Smit; J Pinnagoda; R A Tupker; J Burema; P J Coenraads; J P Nater
Journal:  Int Arch Occup Environ Health       Date:  1990       Impact factor: 3.015

2.  International guidelines for the in vivo assessment of skin properties in non-clinical settings: Part 2. transepidermal water loss and skin hydration.

Authors:  Johan du Plessis; Aleksandr Stefaniak; Fritz Eloff; Swen John; Tove Agner; Tzu-Chieh Chou; Rosemary Nixon; Markus Steiner; Anja Franken; Irena Kudla; Linn Holness
Journal:  Skin Res Technol       Date:  2013-01-19       Impact factor: 2.365

3.  Occlusive properties of monolayer patches: in vitro and in vivo evaluation.

Authors:  Antonella Casiraghi; Paola Minghetti; Francesco Cilurzo; Luisa Montanari; Aarti Naik
Journal:  Pharm Res       Date:  2002-04       Impact factor: 4.200

4.  The aged epidermal permeability barrier. Structural, functional, and lipid biochemical abnormalities in humans and a senescent murine model.

Authors:  R Ghadially; B E Brown; S M Sequeira-Martin; K R Feingold; P M Elias
Journal:  J Clin Invest       Date:  1995-05       Impact factor: 14.808

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.