Takashi Satomi1, Toshiaki Tanaka2, Takehito Kobayashi3, Mituyoshi Iino3. 1. Department of Dentistry, Oral and Maxillofacial Plastic and Reconstructive Surgery, Yamagata Japan. ; The first two authors contributed equally to this work. 2. Department of Dentistry, Oral and Maxillofacial Plastic and Reconstructive Surgery, Yamagata Japan. ; Department of Anatomy and Cell biology, School of Medicine, Faculty of Medicine, Yamagata University, Yamagata Department of Anatomy and Cell biology, School of Medicine, Faculty of Medicine, Yamagata University, Yamagata Japan. ; The first two authors contributed equally to this work. 3. Department of Dentistry, Oral and Maxillofacial Plastic and Reconstructive Surgery, School of Medicine, Faculty of Medicine, Yamagata University, Yamagata Japan.
Abstract
OBJECTIVES: The principle of leverage to superpose the convex surfaces of two shells was applied to develop a device for treating limitation of mouth opening and called it the "shell-shaped mouth opener" and analyzed pressure on the teeth with the TheraBite(®) appliance and the shell-shaped mouth opening appliance. MATERIAL AND METHODS: To compare the TheraBite(®) appliance and the shell-shaped mouth opening appliance, pressure on the teeth in the dentition model with both devices was analyzed using the Inastomer(®) flexible conductive sensor. RESULTS: The load was better dispersed to each tooth in the shell-shaped mouth opening appliance in the all quadrants compared to the TheraBite(®) appliance. CONCLUSIONS: The present study revealed that the shell-shaped mouth opening appliance which was originally invented in our lab, dissipated the mechanical load on teeth more evenly than the TheraBite(®) appliance.
OBJECTIVES: The principle of leverage to superpose the convex surfaces of two shells was applied to develop a device for treating limitation of mouth opening and called it the "shell-shaped mouth opener" and analyzed pressure on the teeth with the TheraBite(®) appliance and the shell-shaped mouth opening appliance. MATERIAL AND METHODS: To compare the TheraBite(®) appliance and the shell-shaped mouth opening appliance, pressure on the teeth in the dentition model with both devices was analyzed using the Inastomer(®) flexible conductive sensor. RESULTS: The load was better dispersed to each tooth in the shell-shaped mouth opening appliance in the all quadrants compared to the TheraBite(®) appliance. CONCLUSIONS: The present study revealed that the shell-shaped mouth opening appliance which was originally invented in our lab, dissipated the mechanical load on teeth more evenly than the TheraBite(®) appliance.
The limitation of mouth opening is a condition marked by orofacial pain and a
limited ability to open the mouth and can result due to tumours, trauma, inflammation,
intra- or extracapsular disorders of the temporomandibular joint (TMJ) [1-3].
Patients with limitation of mouth opening experience multiple problems related to
exhibit restricted chewing, swallowing, and speaking [2].
They also require extra care for maintaining oral hygiene [2].
A variety of mouth opening appliances including rubber plugs, wooden tongue blades,
and the TheraBite® Jaw Motion Rehabilitation System™ (Atos Medical Inc.,
West Allis, WI, USA) have been used to treat the condition of mouth opening limitation
[4]. The TheraBite® appliance
has shown greater efficacy than any other treatments [5,6].
It is a useful appliance for patients with sustained trismus particularly for those
having undergone treatments for head and neck cancers [5,7].
It consists of two mouthpieces and the attached plastic handles. The mouthpieces
are inserted between teeth of the upper and lower jaws. The patient's mouth can
be opened by pressing together the plastic handles that force the mouthpieces to
separate [2]. Some researchers revealed
that muscular problems cause the limitation of mouth opening [8,9].
Other group implied that use of the TheraBite® appliance results in pain
because of the consequent rebounding spasms of the masticatory muscles [4]
and this is because the TheraBite® appliance is a high-torque appliance
that provides a short-duration passive stretch [4].
In some cases, the fracture of a titanium reconstruction plates during the TheraBite®
appliance exercises is an adverse effect that can be explained by non-union of the
fibula to the mandible, by the recurrence of carcinoma, or by excessive stretching
forces [10]. When prescribing exercises
with the TheraBite® appliance in cases of bony reconstructions of the
mandible, exercise forces should be limited until consolidation of the reconstruction
is completed [10]. However, it was
not previously reported how the high-torque using by the TheraBite® appliance
affected the pressure balance to teeth. In general, all mouth opening appliances
only assist in simple jaw opening movements. To overcome these drawbacks, a new
appliance was developed for treating limitation in mouth opening. Moreover, the
pressure balance to the teeth using by the TheraBite® appliance or a
new appliance was investigated by using the Inastomer® (Inaba Rubber
Co. Ltd., Osaka, Japan) flexible conductive sensor.
MATERIAL AND METHODS
Developing a new deviceThe principle of leverage to superpose the convex surfaces of two shells was
applied to develop a device for treating limitation of opening and called it the
"shell-shaped mouth opener" and further improved it to adapt to complex mandibular
movements. Figures 1 shows this new
device that consists of two main bodies and bases, two rods, two bite-parts and
bases, mouth pads, and a rubber band. The main bodies superpose at their convex
surfaces and the bite-part is attached to the body base. A stainless steel rod (2
mm diameter) penetrates through the bite-part and body base, contributing to its
flexibility. The base of the bite-part overlaps the body base and the two main bodies
are held together with a rubber band. The bite-parts are fitted with mouth pads
made of polyethylene foam sheets, which act as a buffer and counterbalance the mechanical
stress and torque against teeth. A new device is made of acrylonitrile-butadiene-styrene
(ABS) resins.
Figure 1
The shell-shaped mouth opening appliance consists of
two main bodies, rods, bite-parts, mouth pads, and a rubber band. The main
bodies superpose at their convex surfaces. The bite-part is attached to
the base of the main body. A stainless steel rod (diameter: 2 mm) penetrates
the bite-part and base of the main body. The base of the bite-part overlaps
the base of the main body and the two main bodies are held.
The shell-shaped mouth opening appliance consists of
two main bodies, rods, bite-parts, mouth pads, and a rubber band. The main
bodies superpose at their convex surfaces. The bite-part is attached to
the base of the main body. A stainless steel rod (diameter: 2 mm) penetrates
the bite-part and base of the main body. The base of the bite-part overlaps
the base of the main body and the two main bodies are held.The mechanism of the "shell-shaped mouth opener" involves a fulcrum that superposes
the convex surfaces of the main bodies, a power point that is formed when the patient
grasps the edges of the bodies, and a point of action that involves the bite-parts.
The leverage at the main bodies and bite-parts and the hinge movement at the main
bodies act together adjusting the complex three-dimensional movements. The combination
of leverage and hinge movements of the "shell-shaped mouth opener" enables it to
adapt to the left-right asymmetric mandibular movements (Figure
2A - C). Moreover, when the patient grasps and eases the edge of the
main bodies, the rubber band at the base of the bite-part expands and contracts
gently to prevent added torque to the teeth and masticatory muscles.
Figure 2
(A) When the patient grasps the right edges of the main
bodies, the shell-shaped mouth opening appliance opens on the left side;
(B) when the patient grasps both edges of the main bodies, the appliance
opens on both sides; (C) when the patient grasps the left edges of the main
bodies, the appliance opens on the right side. Arrows indicate the direction
of action.
(A) When the patient grasps the right edges of the main
bodies, the shell-shaped mouth opening appliance opens on the left side;
(B) when the patient grasps both edges of the main bodies, the appliance
opens on both sides; (C) when the patient grasps the left edges of the main
bodies, the appliance opens on the right side. Arrows indicate the direction
of action.Analysis of pressure on teethTo compare the efficacy of the TheraBite® appliance and the shell-shaped
mouth opening appliance, pressure on the teeth with either of the appliances in
situ was analyzed using the Inastomer® flexible conductive sensor (Figure 3A). The Inastomer® sensors were attached at roots of each
tooth in the dentition model and connected microcomputer. Electric resistance was
measured by microcomputer. The Inastomer® sensor is made with insulating
rubber and electroconductive materials. When pressure is applied, the electroconductive
materials make contact and conductive routes are formed and electric resistance
is decreased. This is because the relationship between electric resistance and mechanical
loading is inversely proportional. F-R property data of the Inastomer®
sensor (Figure 3B) which was provided
by Inaba Rubber Co. Ltd. (Osaka, Japan) was adjusted using the Lagrange interpolating
polynomial under the following conditions: temperature at 24.7 °C; loading at 3
kgf; and loading speed at 200 mm/min. Corrected F-R property data (Figure 3C) was used for analysis. Load value on each tooth was measured three
times and the average was adopted as data. Statistical analysis was performed by
Welch's t test and the number at significance level of P < 0.05 is red colour (Figures 4 - 7).
Figure 3
The mouth pads of both the TheraBite® and
the shell-shaped mouth opening appliance are fitted over the patients' teeth.
Each Inastomer® is attached to the surface of the roots of each
tooth. When the load is applied to the TheraBite® or the shell-shaped
mouth opening appliances, Inastomer® analyzes the electric resistance
on the roots of each tooth (A). F-R property data of the Inastomer
® sensor is shown (B). F-R property data was corrected by the Lagrange
interpolating polynomial. Corrected F-R property data is shown (C).
Figure 4
Results of pressure monitoring on teeth by mouth opening
appliances (11 - 15). When the distance of the interincisor opening reached
13.3 mm, the central incisor was loaded at 3 kgf, which was greater than
that for other teeth in the TheraBite® appliance. On the other hand, as the distance of interincisor opening
was gradually increased, the canine and the first and second premolars were
loaded gradually in the shell-shaped mouth opening appliance. In addition,
the central and lateral incisors were loaded at all times in the shell-shaped
mouth opening appliance.
The mouth pads of both the TheraBite® and
the shell-shaped mouth opening appliance are fitted over the patients' teeth.
Each Inastomer® is attached to the surface of the roots of each
tooth. When the load is applied to the TheraBite® or the shell-shaped
mouth opening appliances, Inastomer® analyzes the electric resistance
on the roots of each tooth (A). F-R property data of the Inastomer
® sensor is shown (B). F-R property data was corrected by the Lagrange
interpolating polynomial. Corrected F-R property data is shown (C).The two-digit system for the dental formula was used:The Lagrange interpolating polynomial:
RESULTS
Pressure on teeth with the TheraBiteIn the upper right quadrant, compared to the shell-shaped mouth opening appliance,
the central incisor was loaded (Shell vs TheraBite® 11 at 21.6 mm P =
0.0056) and other teeth were not loaded diffusely in the TheraBite® appliance
(Shell vs TheraBite® 12 at 21.6 mm P = 0.00000086, 13 at 21.6 mm P =
0.018, 14 at 21.6 mm P = 0.026, 15 at 21.6 mm P = 0.00045). However, the canine
was partially loaded and other teeth were loaded diffusely in the shell-shaped mouth
opening appliance. Moreover, the canine was gradually loaded as the interincisor
opening was increased in the shell-shaped mouth opening appliance (Shell vs TheraBite®
13 at 12.3 mm P = 0.019, 13 at 13.3 mm P = 0.036, 13 at 15.7 mm P = 0.024, 13 at
17.6 mm P = 0.003, 13 at 21.6 mm P = 0.018). Furthermore, compared to the TheraBite®
appliance, all teeth in the shell-shaped mouth opening appliance were loaded (Figure 4).Results of pressure monitoring on teeth by mouth opening
appliances (11 - 15). When the distance of the interincisor opening reached
13.3 mm, the central incisor was loaded at 3 kgf, which was greater than
that for other teeth in the TheraBite® appliance. On the other hand, as the distance of interincisor opening
was gradually increased, the canine and the first and second premolars were
loaded gradually in the shell-shaped mouth opening appliance. In addition,
the central and lateral incisors were loaded at all times in the shell-shaped
mouth opening appliance.In the upper left quadrant, the canine was mainly loaded in TheraBite®
appliance and the shell-shaped mouth opening appliance. The canine and the first
premolar were partially loaded and the second premolar was gradually loaded as the
interincisor opening was increased in the shell-shaped mouth opening appliance (Shell
vs TheraBite® 25 at 15.7 mm P = 0.0067, 25 at 17.6 mm P = 0.0054, 25
at 21.6 mm P = 0.041). Additionally, compared to the TheraBite® appliance,
the lateral incisors were loaded in the shell-shaped mouth opening appliance (Shell
vs TheraBite® 22 at 10.4 mm P = 0.0045, 22 at 10.8 mm P = 0.001, 22 at
12.3 mm P = 0.021, 22 at 13.3 mm P = 0.018, 22 at 15.7 mm P = 0.015, 22 at 17.6
mm P = 0.0024, 22 at 21.6 mm P = 0.001) (Figure
5).
Figure 5
Results of pressure monitoring on teeth with mouth opening
appliances (21 - 25). When the distance of the interincisor opening was
increased, the canine was gradually loaded, but the load on the first and
second premolars decreased using the TheraBite® appliance. Conversely, as the distance of the interincisor opening increased, the canine and the first and second premolars were loaded gradually in the shell-shaped mouth opening appliance. Additionally, the central and lateral incisors were loaded at all times in the shell-shaped mouth opening appliance.
Results of pressure monitoring on teeth with mouth opening
appliances (21 - 25). When the distance of the interincisor opening was
increased, the canine was gradually loaded, but the load on the first and
second premolars decreased using the TheraBite® appliance. Conversely, as the distance of the interincisor opening increased, the canine and the first and second premolars were loaded gradually in the shell-shaped mouth opening appliance. Additionally, the central and lateral incisors were loaded at all times in the shell-shaped mouth opening appliance.In the lower left quadrant, the first premolar was slightly loaded, but all teeth
were weakly loaded in TheraBite® appliance. As the distance of the interincisor
opening was gradually increased, the canine, the first premolar, and the second
premolar were gradually loaded in the shell-shaped mouth opening appliance. Moreover,
the lateral incisors were loaded in the shell-shaped mouth opening appliance (Shell
vs TheraBite® 32 at 10.4 mm P = 0.0073, 32 at 10.8 mm P = 0.0054, 32
at 12.3 mm P = 0.0086, 32 at 13.3 mm P = 0.006, 32 at 15.7 mm P = 0.000096, 32 at
17.6 mm P = 0.000055, 32 at 21.6 mm P = 0.00015). Compared to the shell-shaped mouth
opening appliance, all teeth in TheraBite® appliance were not loaded
effectively in the lower left quadrant (Figure
6).
Figure 6
Results of pressure monitoring on teeth with mouth opening
appliances (31 - 35). All teeth were loaded weakly at all times using the
TheraBite® appliance. As the distance of the interincisor opening was gradually increased, the canine was mainly loaded and the first and second premolars
were loaded gradually in the shell-shaped mouth opening appliance. Moreover,
the central and lateral incisors were loaded at all times in the shell-shaped
mouth opening appliance.
Results of pressure monitoring on teeth with mouth opening
appliances (31 - 35). All teeth were loaded weakly at all times using the
TheraBite® appliance. As the distance of the interincisor opening was gradually increased, the canine was mainly loaded and the first and second premolars
were loaded gradually in the shell-shaped mouth opening appliance. Moreover,
the central and lateral incisors were loaded at all times in the shell-shaped
mouth opening appliance.In lower right quadrant, as the distance of the interincisor opening was gradually
increased, the canine was loaded in TheraBite® appliance and the shell-shaped
mouth opening appliance. However, other teeth were not loaded effectively in the
TheraBite® appliance. Compared to the TheraBite® appliance,
the central (Shell vs TheraBite® 41 at 10.4 mm P = 0.00019, 41 at 10.8
mm P = 0.012, 41 at 12.3 mm P = 0.004, 41 at 13.3 mm P = 0.013, 41 at 15.7 mm P
= 0.014, 41 at 17.6 mm P = 0.01, 41 at 21.6 mm P = 0.0000026) and lateral incisors
(Shell vs TheraBite® 42 at 10.4 mm P = 0.00062, 42 at 10.8 mm P = 0.0089,
42 at 12.3 mm P = 0.000037, 42 at 13.3 mm P = 0.0044, 42 at 15.7 mm P = 0.0019,
42 at 17.6 mm P = 0.000097, 42 at 21.6 mm P = 0.00013) were loaded in the shell-shaped
mouth opening appliance (Figure 7).
Figure 7
Results of pressure monitoring on teeth by mouth opening
appliance (41 - 45). As the distance of the interincisor opening was gradually
increased, the canine was strongly loaded in case of the TheraBite
® appliance. When the interincisor opening distance reached 10.4 mm,
the central and lateral incisors were loaded mainly in the shell-shaped
mouth opening appliance. Furthermore, as the distance of the interincisor
opening was gradually increased, the canine and first premolars were gradually
loaded.
Results of pressure monitoring on teeth by mouth opening
appliance (41 - 45). As the distance of the interincisor opening was gradually
increased, the canine was strongly loaded in case of the TheraBite
® appliance. When the interincisor opening distance reached 10.4 mm,
the central and lateral incisors were loaded mainly in the shell-shaped
mouth opening appliance. Furthermore, as the distance of the interincisor
opening was gradually increased, the canine and first premolars were gradually
loaded.These results indicated that the load was better dispersed to each tooth in the
shell-shaped mouth opening appliance in the all quadrants compared to the TheraBite®
appliance.
DISCUSSION
The limitation of mouth opening occurs frequently due to tumours, trauma, inflammation,
TMJ disorders, and their treatment. Regardless of the relative commonness of this
condition and its harmful effects on daily functioning and quality of life, there
are only few studies concerning the treatment of limitation of mouth opening and
no standard care exists [4]. Although
physical therapy, hyperbaric oxygen [11],
pentoxifylline [12], botulinum toxin
injection [13], and surgical coronoidectomy
[3] have been used for treating the
condition. However, mouth opening appliances provide a conservative approach to
therapy in addition to being easy to use and having better efficacy than physiotherapy
and surgery [14,15]. Several mouth
opening appliances have been developed for the treatment of limitation of mouth
opening.In the present study, the Inastomer® flexible conductive sensor was
used to analyze the pressure on teeth. To our knowledge, there were no reports that
the Inastomer® flexible conductive sensor was used to analyze the pressure
on teeth. However, it is useful tool to measure the pressure on teeth. Analysis
of pressure on teeth due to mouth opening appliances revealed that the shell-shaped
mouth opening appliance dissipated the mechanical load on teeth more evenly than
the TheraBite® appliance (Figures
4 - 7). Well-balanced load on teeth was relied on the constitution of
the shell-shaped mouth opening appliance. The combination of leverage and hinge
movements enabled the shell-shaped mouth opening appliance to adapt to complex mandibular
movements, and the mouth pads helped to evenly distribute the mechanical stress
and torque. In this regard, the shell-shaped mouth opening appliance may be superior
in reducing the overconcentration of particular tooth during the treatment of limitation
of mouth opening, which is paramount for patient compliance and positive therapy
outcomes [16] and contribute to
avoid the fracture of a titanium reconstruction plates [10].
However, to confirm whether the shell-shaped mouth opening appliance is superior
to the TheraBite® appliance, further research need to compare the TheraBite®
appliance and the shell-shaped mouth opening appliance and to establish the best
way to use the shell-shaped mouth opening appliance for the treatment of limitation
of opening in vivo study.
CONCLUSIONS
The present study revealed that the shell-shaped mouth opening appliance dissipated
the mechanical load on teeth more evenly than the TheraBite® appliance.