Literature DB >> 24405392

Comparison of a Schmidt and Haensch refractometer and an Atago PAL-USG Cat refractometer for determination of urine specific gravity in dogs and cats.

Harold W Tvedten1, Aså Norén.   

Abstract

BACKGROUND: The performance of a digital Atago PAL-USG Cat refractometer (Atago) was compared with a Schmidt and Haensch, Goldberg type refractometer (S+H).
MATERIALS AND METHODS: Specific gravity of 47 canine and feline urine samples was determined with both refractometers and the results were compared with Passing-Bablok and Bland-Altman plots. In addition, the specific gravity of dilutions of 10% glucose, 10% NaCl, and 3% albumin solutions was determined and compared with expected values.
RESULTS: Both refractometers consistently reported 1.000 with distilled water. The correlation between both refractometers based on Passing-Bablok plots of 47 urine samples was excellent (r = .99), but, in the Bland-Altman plots, there was a significant, proportional, negative error for the Atago readouts. This was also illustrated by the fact that 10 urine samples with an S+H result of > 1.030 were read out between 1.023 and 1.028 by Atago. Schmidt and Haensch results of various glucose solutions matched exactly expected values, but Atago results were lower. Likewise, S+H results with diluted NaCl solutions were closer to expected results than Atago results. In contrast, Atago results with dilutions of 3% albumin were closer to expected results than S+H results. DISCUSSION: The Atago refractometer reported lower specific gravity results in urine and standard solutions of glucose and NaCl, which could adversely affect clinical decisions concerning normal renal function based on solute concentrations determined in canine and feline urine samples.
© 2014 American Society for Veterinary Clinical Pathology and European Society for Veterinary Clinical Pathology.

Entities:  

Keywords:  Cat; dog; refractometer; urine specific gravity

Mesh:

Substances:

Year:  2014        PMID: 24405392     DOI: 10.1111/vcp.12110

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  5 in total

Review 1.  Label-Free Physical Techniques and Methodologies for Proteins Detection in Microfluidic Biosensor Structures.

Authors:  Georgii Konoplev; Darina Agafonova; Liubov Bakhchova; Nikolay Mukhin; Marharyta Kurachkina; Marc-Peter Schmidt; Nikolay Verlov; Alexander Sidorov; Aleksandr Oseev; Oksana Stepanova; Andrey Kozyrev; Alexander Dmitriev; Soeren Hirsch
Journal:  Biomedicines       Date:  2022-01-18

2.  Reliability of Three Urine Specific Gravity Meters Measuring Brix and Urine Solutions at Different Temperatures.

Authors:  Floris Wardenaar; Carmen P Ortega-Santos; Kaila Vento; Stephanie Olzinski; Jason Olig; Stavros Kavouras; Carol Johnston
Journal:  J Athl Train       Date:  2021-03-31       Impact factor: 2.860

3.  Interobserver reliability of canine urine specific gravity assessed by analog or digital refractometers.

Authors:  John C Rowe; Jessica A Hokamp; Jessica N Braatz; John R Freitag-Engstrom; Nicole L Stephens; Dennis J Chew; Catherine Langston; Adam J Rudinsky
Journal:  J Vet Diagn Invest       Date:  2021-03-01       Impact factor: 1.279

4.  Evaluation and analytical validation of a handheld digital refractometer for urine specific gravity measurement.

Authors:  Sara P Wyness; Joshua J H Hunsaker; Taylor M Snow; Jonathan R Genzen
Journal:  Pract Lab Med       Date:  2016-06-02

5.  Clinical performance evaluation of total protein measurement by digital refractometry and characterization of non-protein solute interferences.

Authors:  Joshua J H Hunsaker; Sara P Wyness; Taylor M Snow; Jonathan R Genzen
Journal:  Pract Lab Med       Date:  2016-08-17