The properties of an improvised piston pump for the rapid delivery of intravenous fluids.

C M Smart1, C W Primrose, A L Peters, E J Speirits.   

Abstract

To maximise the effect of a small fluid load, it is occasionally desirable to bolus manually with multiple depressions of a large-capacity syringe. This is usually achieved by placing the syringe on the side port of a three-way tap. We modified this technique by placing two-one-way valves in line with the three-way tap, effectively creating a piston pump, the infusion rates via which we compared with those achieved by an inflatable pressure-infuser in a simulated resuscitation. Fluid flow was faster using the piston pump than with the pressure-infuser (mean (SD) time to infuse 2000 ml saline 0.9% via a 16-G cannula 352 (10) s vs 495 (19) s, respectively, p < 0.0001). The piston pump appears to have potential for both tight control of fluid delivery and major high-volume resuscitation. The lightweight nature of the pump and its lack of reliance on gravity may also make it suitable for the pre-hospital setting.
© 2014 The Association of Anaesthetists of Great Britain and Ireland.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24383681     DOI: 10.1111/anae.12463

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  3 in total

1.  In-line pressure within a HOTLINE® Fluid Warmer, under various flow conditions.

Authors:  Midoriko Higashi; Ken Yamaura; Yukie Matsubara; Takuya Fukudome; Sumio Hoka
Journal:  J Clin Monit Comput       Date:  2014-08-03       Impact factor: 2.502

2.  Comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: A bench study.

Authors:  Wataru Hashimoto; Ichiro Takenaka; Keisuke Yasunami; Tomoko Minami; Haruhiko Sano
Journal:  Indian J Anaesth       Date:  2020-12-12

3.  A pressure-resistant peripherally inserted central catheter is as useful as a central venous catheter for rapid fluid infusion: an in vitro study.

Authors:  Jun Maki; Makoto Sumie; Tomoko Ide; Masako Nagamatsu; Katsuyuki Matsushita; Kazuhiro Shirozu; Midoriko Higashi; Ken Yamaura
Journal:  BMC Anesthesiol       Date:  2022-07-04       Impact factor: 2.376

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.