| Literature DB >> 24383576 |
Randy Jackson1, Robert P Oda, Raj K Bhandari, Sari B Mahon, Matthew Brenner, Gary A Rockwood, Brian A Logue.
Abstract
Although commonly known as a highly toxic chemical, cyanide is also an essential reagent for many industrial processes in areas such as mining, electroplating, and synthetic fiber production. The "heavy" use of cyanide in these industries, along with its necessary transportation, increases the possibility of human exposure. Because the onset of cyanide toxicity is fast, a rapid, sensitive, and accurate method for the diagnosis of cyanide exposure is necessary. Therefore, a field sensor for the diagnosis of cyanide exposure was developed based on the reaction of naphthalene dialdehyde, taurine, and cyanide, yielding a fluorescent β-isoindole. An integrated cyanide capture "apparatus", consisting of sample and cyanide capture chambers, allowed rapid separation of cyanide from blood samples. Rabbit whole blood was added to the sample chamber, acidified, and the HCN gas evolved was actively transferred through a stainless steel channel to the capture chamber containing a basic solution of naphthalene dialdehyde (NDA) and taurine. The overall analysis time (including the addition of the sample) was <3 min, the linear range was 3.13-200 μM, and the limit of detection was 0.78 μM. None of the potential interferents investigated (NaHS, NH4OH, NaSCN, and human serum albumin) produced a signal that could be interpreted as a false positive or a false negative for cyanide exposure. Most importantly, the sensor was 100% accurate in diagnosing cyanide poisoning for acutely exposed rabbits.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24383576 PMCID: PMC3983020 DOI: 10.1021/ac403846s
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986
Endogenous Levels of Cyanide Thiocyanate and ATCA in the Blood of Smokers and Non-Smokers
| marker of CN exposure | typical biological matrix analyzed | nonsmoker (μM) | smoker (μM) | refs |
|---|---|---|---|---|
| cyanide | whole blood or RBCs | 0.02–10 | 0.03–10 | ( |
| thiocyanate | plasma | 4.6–130 | 1.7–290 | ( |
| ATCA | urine or plasma | 0.08–0.27 | 0.12–0.45 | ( |
Concentrations compiled for nonsmokers ranged from 0.02 to 3 μM for Logue et al.[1] and 3–10 μM for Minakata et al.[17]
Logue et al.[1] compiled endogenous concentrations of cyanide, thiocyanate, and ATCA for smokers and nonsmokers from studies prior to 2010.
Comparison of Recently Proposed Rapid Analysis Methods and/or Portable Technologies for the Diagnosis of Cyanide Exposure
| investigators | core technology | sample prep method | analysis time (min) | LOD | notes |
|---|---|---|---|---|---|
| Ma et al., 2011[ | hydroxoaquocobinamide | microdiffusion | ∼2 | 0.5 | H2S is an interferent. |
| Ma and Dasgupta,
2010[ | hydroxoaquocobinamide | microdiffusion | ∼1.5 | 0.030 | H2S is an interferent,
and the NaOH mobile phase
is necessary. |
| Tian
et al., 2013[ | hydroxocyanocobinamide | microdiffusion | <4 | 2.2 | Potential interferents were not evaluated, but H2S likely interferes. |
| Lee et al., 2012[ | turn on fluorescence Cu2+–c-mpg-C3N4 | Isolate serum. Follow on sample prep not described. | 40 | 0.080 | The analysis time reported (10 min) likely did not include
the time needed to clot blood and separate the serum. |
| Kumar et al., 2013[ | fluorescence of 1-(4′-nitrophenyl)benzimidazolium | Isolate serum then add HEPES buffer and DMSO solution. | 31 | 0.030 | The analysis time reported (<60 s) did not include the time needed to clot blood and
separate serum. |
LOD, limit of detection.
The listed LODs are for rabbit whole blood.
These techniques were verified using CN exposed rabbits.
Method not verified in an animal model.
c-mpg-C3N4 is cubic mesoporous graphitic carbon nitride.
Thirty minutes was added to the reported analysis time to account for the estimated time neseccary to clot blood and separate serum from blood.
The listed LODs are for human blood serum.
The sample preparation to obtain serum from blood requires extra equipment.
Figure 1Schematic of the stacked cyanide capture apparatus.
Figure 2The proposed reaction schemes for the possible reactions of NDA, taurine, and cyanide. Pathways A → B and A → C both yield H2O as a byproduct.
Figure 3Calibration curves obtained for the PE frit (20 mL of air), the silicone septa with forward flow tubing (50 mL of air shown), and rabbit whole blood (50 mL of air). Aqueous standards were used for the PE frit (○) and the silicone septa (□). A silicone septa, with forward flow tubing as the chamber separation material, was used for analysis of rabbit whole blood (△). Error bars represent standard deviation.
Figure 4Assessment of the short- and long-term stability of the capture solution reagents. The long-term stability of the reagents (up to 70 days) is presented in the inset. Error bars represent standard deviation.
Figure 5Assessment of potential interferents to the sensor technology present in cyanide spiked blood. Error bars represent standard deviation.
Figure 6(A) Comparison of the cyanide concentrations found in the whole blood of cyanide exposed rabbits at 15, 25, and 35 min into the infusion period (5, 8.3, and 11.7 mg NaCN exposure, respectively). The dashed line represents the LLOQ (3.12 μM) and the solid line represents 10 μM cyanide, the threshold considered “cyanide exposure” for this study. Standard deviation values (±3 s) for the lines were not presented because they were negligible compared to the scale of the x axis. Note that for the 15 min time point n = 3 because three animals did not have blood drawn at that time interval. (B) Dose–response curves for three different doses of NaCN (5, 8.3, and 11.7 mg) intravenously administered to rabbits. Error bars represent standard deviation.