Postural change in intraocular pressure: a comparison of measurement with a Goldmann tonometer, Tonopen XL, pneumatonometer, and HA-2.

Yaniv Barkana1.   

Abstract

PURPOSE: To compare directly the results of 4 tonometers in measuring postural change in the intraocular pressure (IOP) between sitting and lying body positions.
METHODS: In 19 healthy subjects, the IOP was measured both eyes with a Goldmann applanation tonometer (GAT) while sitting, then using the same GAT with the subject in the left lateral decubitus position after lying for 15 minutes, using a novel system comprising a motorized bed and a modified slit-lamp table. On the next day, the sitting IOP was measured in 1 eye using GAT, and additionally in random order Tonopen XL and pneumatonometer. Then subjects were asked to lie down, and the IOP was measured with these tonometers and also with a hand-held applanation tonometer (HA-2) after lying for 15 minutes and for 45 minutes; measurements were made with subjects in the supine position except with GAT.
RESULTS: There were 10 male and 9 female subjects, with a mean age of 33.0±12.4 years. On day 1, the sitting GAT IOP (mm Hg) was 13.7±3.0 right eye and 13.6±2.8 left eye (P=0.7) and the lying GAT IOP was 17.8±3.5 right eye and 18.1±3.2 left eye (P=0.3 for inter-eye postural IOP change). Postural change in the right and the left eyes was moderately correlated, with a coefficient of 0.453. On day 2, postural changes after lying for 15 minutes as measured with a GAT, a Tonopen, a pneumatonometer, and a HA-2 were 4.9±2.6, 1.6±1.8, 4.2±2.0, and 3.1±2.7, respectively. After lying for 45 minutes, it decreased to 3.3±2.1, 1.1±4.3, 3.1±2.5, and 2.6±3.3, respectively; this IOP decrease was statistically significant for the GAT and the pneumatonometer. The inter-subject variability was large, as in some subjects there was nearly no postural change in the IOP, whereas in some the IOP increased by 8 to 10 mm Hg. Ninety-five percent limits of agreement showed a poor agreement between GAT and the other 3 tonometers in the sitting and lying IOPs and postural IOP changes measured.
CONCLUSIONS: The average postural IOP change was similar when measured with GAT and a pneumatonometer, smaller with a hand-held GAT, and much smaller with a Tonopen XL. This change decreased significantly when lying between the 15- and 45-minute time points. The inter-subject variability in postural IOP was remarkable, consistent with previous reports, suggesting the importance of this parameter in clinical glaucoma practice. Inter-device agreement was poor and similar to previous reports.

Entities:  

Mesh:

Year:  2014        PMID: 24370809     DOI: 10.1097/IJG.0b013e3182a0762f

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  8 in total

1.  Development and validation of a machine learning, smartphone-based tonometer.

Authors:  Aaron Y Lee; Joanne C Wen; Yue Wu; Ian Luttrell; Shu Feng; Philip P Chen; Ted Spaide
Journal:  Br J Ophthalmol       Date:  2019-12-23       Impact factor: 4.638

2.  Ocular changes over 60 min in supine and prone postures.

Authors:  Allison P Anderson; Gautam Babu; Jacob G Swan; Scott D Phillips; Darin A Knaus; Christine M Toutain-Kidd; Michael E Zegans; Abigail M Fellows; Jiang Gui; Jay C Buckey
Journal:  J Appl Physiol (1985)       Date:  2017-05-25

3.  The effect of lateral decubitus position on nocturnal intraocular pressure over a habitual 24-hour period in healthy adults.

Authors:  Jie Hao; Yi Zhen; Hao Wang; Diya Yang; Ningli Wang
Journal:  PLoS One       Date:  2014-11-25       Impact factor: 3.240

4.  Comparison of Intraocular Pressure Measurements in Healthy Pediatric Patients using Three Types of Tonometers.

Authors:  Muhsin Eraslan; Eren Çerman; Sena Sümmen
Journal:  Turk J Ophthalmol       Date:  2017-01-17

5.  Evaluation of the Effect of Body Position on Intraocular Pressure Measured with Rebound Tonometer

Authors:  Hüseyin Mayalı; Beyza Tekin; Özcan Rasim Kayıkçıoğlu; Emin Kurt; Süleyman Sami İlker
Journal:  Turk J Ophthalmol       Date:  2019-02-28

6.  Intraocular Pressure Measurements in Standing Position with a Rebound Tonometer.

Authors:  Maddalena De Bernardo; Maria Borrelli; Giovanni Cembalo; Nicola Rosa
Journal:  Medicina (Kaunas)       Date:  2019-10-18       Impact factor: 2.430

7.  Acute effects of posture on intraocular pressure.

Authors:  Emily S Nelson; Jerry G Myers; Beth E Lewandowski; C Ross Ethier; Brian C Samuels
Journal:  PLoS One       Date:  2020-02-06       Impact factor: 3.240

8.  Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty.

Authors:  Oded Ohana; David Varssano; Gabi Shemesh
Journal:  Indian J Ophthalmol       Date:  2017-07       Impact factor: 1.848

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.