| Literature DB >> 24354736 |
Bozena Kostek1, Tomasz Poremski.
Abstract
BACKGROUND: This study investigates the usefulness and effectiveness of a new way of tinnitus screening and diagnosing. The authors believe that in order to arrive at relevant diagnostic information, select the tinnitus treatment and quantitatively substantiate its effects, the measurement of the Tinnitus psychoacoustic parameters should be made an inherent part of the Tinnitus therapy.Entities:
Mesh:
Year: 2013 PMID: 24354736 PMCID: PMC3924908 DOI: 10.1186/1746-1596-8-209
Source DB: PubMed Journal: Diagn Pathol ISSN: 1746-1596 Impact factor: 2.644
Figure 1The user interface for Tinnitus sound synthesis.
Comparison of the results of determining tinnitus acoustic parameters using the audiometer and the synthesizer
| Participant no. | Gender | Age | Type and location of tinnitus (info from interview) | Tinnitus audiometric parameters [frequency|dB SPL] | Evaluation of the resemblance of the generated sound to the participant’s tinnitus in scale of [0…10] | Test duration [min] | Tinnitus parameters obtained with the synthesizer [frequency|dB SPL] | Evaluation of the resemblance of the generated sound to the participant’s tinnitus in scale of [0…10] | Test duration [min] |
| 1 | M | 51 | Squeal, in the head | RE: 6 kHz|65 | RE: 2724 Hz|35.5 | ||||
| LE: 4 kHz|59.5 | LE: 2655 Hz|32.4 | ||||||||
| 2 | M | 67 | High whistle, in LE and RE | RE: 4 kHz|34.5 | RE: 6172 Hz|79 | ||||
| LE: 4 kHz|34.5 | LE: 6251 Hz|76.7 | ||||||||
| 3 | M | 77 | Hum of bees, in the head | RE: 3 kHz|70 | RE: 1929 Hz|56.7 | ||||
| LE: 3 kHz|75 | LE: 1954 Hz|57.3 | ||||||||
| 4 | F | 79 | Low-frequency hum and squeal, in LE | LE: 2 kHz|49 | LE: 82 Hz|88.4 | ||||
| 361 Hz|65.5 | |||||||||
| 1281 Hz|39 | |||||||||
| 5 | M | 36 | Squeal, in RE | RE: 4 kHz|79.5 | RE: 5572 Hz|72 | ||||
| 6 | F | 61 | Murmur, squeal, in RE and LE | RE: 8 kHz|68 | LE: Noise(8.9-10.4) | ||||
| LE: 8 kHz|63 | | | kHz ; Favg = 9682 Hz|44.5 | ||||||
| 7 | F | 61 | High metallic sound | RE: 8 kHz|33 | RE: 7103 Hz|37.2 | ||||
| LE: 8 kHz|43 | LE: 7103 Hz|44.9 | ||||||||
| 8 | F | 65 | Constant noise in head | RE: undefined | RE: undefined | ||||
| LE: 1 kHz NB|38 | | | LE: Noise(0.35-2.1) | ||||||
| kHz; Favg = 1229 Hz|31.6 | |||||||||
| 9 | M | 42 | Squeal | RE: 4 kHz|59.5 | RE: 3216 Hz|61.5 | ||||
| LE: 3 kHz|60 | LE: 3216 Hz|61.5 | ||||||||
| 10 | F | 61 | Constant noise in RE | RE: 2 kHz NB|86 | RE: Noise(0.9-2.4) | ||||
| kHz; Favg:1655 Hz|52.2 | |||||||||
| 11 | F | 37 | Squeal or hiss in LE | LE: 3 kHz|35 | LE: 2398 Hz|32.7 | ||||
| 12 | F | 37 | Squeal or hiss in LE | LE: 6 kHz NB|25 | LE: Noise(6.2-6.6) | ||||
| kHz; Favg = 6424 Hz|22.1 | |||||||||
| 13 | M | 24 | High squeal or noise of working TV in head | RE: 6 kHz|20 | RE: 6093 Hz|28 | ||||
| LE: 6 kHz|30 | LE: 6093 Hz|38.7 | ||||||||
| 14 | M | 24 | High squeal or noise of working TV in head | RE: 6 kHz NB|20 | RE: Noise(3.6-6.1) | ||||
| kHz; Favg = 4874 Hz|14.3 | |||||||||
| LE: 6 kHz NB|35 | LE: Noise(3.6-6.1) | ||||||||
| kHz; Favg = 4874 Hz|28.2 | |||||||||
Denotations: RE–right ear, LE–left ear, Favg–frequency average of noise, NB–audiometric narrow band noise, F–female, M–male.
Results of the statistical evaluation of tests
| | ||||
|---|---|---|---|---|
| 1 | 7 | 9 | 4 | 2 |
| 2 | 6 | 9 | 6 | 2 |
| 3 | 3 | 8 | 4 | 2 |
| 4 | 2 | 9 | 5 | 4 |
| 5 | 3 | 3 | 8 | 6 |
| 6 | 6 | 9 | 10 | 8 |
| 7 | 7 | 9 | 6 | 2 |
| 8 | 9 | 10 | 6 | 1 |
| 9 | 8 | 8 | 5 | 2 |
| 10 | 8 | 5 | 3 | 4 |
| 11 | 5 | 5 | 3 | 2 |
| 12 | 7 | 8 | 5 | 3 |
| 13 | 7 | 8 | 9 | 4 |
| 14 | 1 | 9 | 2 | 6 |
| Results of the Shapiro-Wilk test | ||||
| Results of the Wilcoxon test | ||||
Key: W–Results of the Shapiro-Wilk test, T, Z–Results of Wilcoxon test, p–statistical significance (p-value), r–effect size.
Figure 2Evaluation of the resemblance of the generated noise to the noise perceived by the patient.
Figure 3Comparison of the length of the test using an audiometer and synthesizer. Wilcoxon test description (http://www.uwe.ac.uk/hlss/llas/statistics-in-linguistics/chapter%2008.pdf, link from Nov. 26).