Aptin Maboudi1, Hamidreza Abtahi1, Mostafa Hosseini2, Amin Tamadon3, Enayat Safavi1.
Abstract
BACKGROUND: Untrained fingertip palpation has been shown to be unable to adjust endotracheal tube (ET) cuff pressure in most intubated patients.Entities:
Keywords: Manometry; Nursing, Pressure; School
Year: 2013 PMID: 24349723 PMCID: PMC3838645 DOI: 10.5812/ircmj.4164
Source DB: PubMed Journal: Iran Red Crescent Med J ISSN: 2074-1804 Impact factor: 0.611
Figure 1.Box and whisker plot of median endotracheal cuff tube pressure (cmH2O) before and after training sessions measured by cuff manometry in the different periods of adjustment of cuff pressure by hand touching of ICU nurse training
Number (%) of endotracheal tube cuff pressure scores a obtained by fingertip palpation in seven sessions
| Measurement Time [ | Safe, No. (%) | Low, No. (%) | High, No. (%) | Very High, No. (%) | Total, No. | Mean ± SD | |
|---|---|---|---|---|---|---|---|
|
| S1 | 31 (41.3) | 21 (28.0) | 7 (9.3) | 16 (21.3) | 75 | 36.2 ± 24.2 |
| S2 | 58 (80.6) | 11 (15.3) | 3 (4.2) | 0 (0) | 72 | 25.1 ± 5.6 | |
|
| S3 | 53 (71.6) | 17 (23.0) | 3 (4.1) | 1 (1.4) | 74 | 23.9 ± 6.6 |
| S4 | 68 (90.7) | 7 (9.3) | 0 (0) | 0 (0) | 75 | 24.7 ± 3.5 | |
|
| S5 | 60 (83.3) | 10 (13.9) | 1 (1.4) | 1 (1.4) | 72 | 25.8 ± 5.7 |
|
| S6 | 43 (71.7) | 17 (28.3) | 0 (0) | 0 (0) | 60 | 22.2 ± 3.4 |
|
| S7 [ | 22 (73.3) | 7 (23.3) | 1 (3.3) | 0 (0) | 30 | 24.9 ± 6.0 |
aCuff pressure scores: Safe, > 20 to ≤35 cmH2O; Low, ≤ 20 cmH2O; High, > 35 to ≤ 50 cmH2O; Very high, > 50 cmH2O. S1-S7 as defined in the text
bOnly 10 nurses were available due to ICU staff turnover
Percentage of endotracheal tube cuff pressure scores a obtained by Cuff manometer protocols (twice a day at fixed intervals and after each intervention on the endotracheal tube) in previous studies
| Study | Measurement time | Appropriate, % | Low, % | High, % |
|---|---|---|---|---|
|
| Not defined[ | 56 | 15 | 29 |
|
| S1 [ | 17 | 6 | 77 |
| 68 | 8 | 24 | ||
| 28 | 40 | 31 | ||
|
| S1 [ | 31 [ | 28[ | 41[ |
| 64 | 27 | 9 | ||
| 57 | 23 | 20 |
aPercentage of time in each cuff pressure scores in a referral ICU with standard cuff manometer protocol. Appropriate, > 15 to ≤ 30 cmH2O; low, ≤ 15 cmH2O; high, > 30 cmH2O
bS1: Before protocol implementation; S2: One month after Protocol; S3: Three years after protocol. Appropriate, > 20 to ≤30 cmH2O; low, ≤ 20 cmH2O; high, > 30 cmH2O
cS1, S6 and S7 as defined in the text
dThe definitions of cuff pressure scores are the same as was used by Jaber et al. (2) and are different from definitions that are used in Table 1