Literature DB >> 24346586

Design characteristics of a three-dimensional geometric aortic valve annuloplasty ring.

Philip S Crooke1, L Alan Beavan, Charles D Griffin, Domenico Mazzitelli, J Scott Rankin.   

Abstract

OBJECTIVE: A full geometric annuloplasty ring could facilitate aortic valve repair. The purpose of this report was to document the design of such a ring using mathematical analyses of normal human aortic valve computed tomographic angiograms.
METHODS: One-millimeter axial slices of high-resolution computed tomographic angiograms from 11 normal aortic roots were used to generate high-density x, y, and z coordinates of valve structures in Mathematica. Three-dimensional least squares regression analyses of leaflet-sinus coordinates were used to assess geometry of aortic valve and root structures.
RESULTS: Normal valve geometry could be represented as three leaflet-sinus general ellipsoids nested within an elliptical aortic root. Minor-major diameter ratio of the valve base was 0.60 ± 0.07, and elliptical geometry extended vertically up the commissures. By contrast, leaflet-sinus horizontal circumferences were fairly circular (diameter ratios, 0.82-0.87), and the left coronary/noncoronary commissural post was located at the posterior base minor diameter-circumference junction, with the center of the right coronary leaflet opposite. Post location on the circumference was symmetrical, with a deviation of only ±2% to ±3% from 33.3% symmetry. Commissural posts flared outward by 5 to 10 degrees, and leaflet areas were statistically equivalent (P > 0.10). From end diastole to midsystole, the aortic root became less elliptical (diameter ratio increased by 0.15), but root area expanded minimally (less than +5%). A one-piece rigid annuloplasty ring was designed with 2:3 base ellipticality, three 10-degree outwardly flaring symmetrical posts, and post height = base circumference/2π.
CONCLUSIONS: A three-dimensional aortic annuloplasty ring was designed that could prove useful for enhancing applicability and stability of aortic valve repair.

Entities:  

Mesh:

Year:  2013        PMID: 24346586     DOI: 10.1097/IMI.0000000000000026

Source DB:  PubMed          Journal:  Innovations (Phila)        ISSN: 1556-9845


  2 in total

1.  First uses of HAART 300 rings for aortic valve repair in Poland - 4 case studies.

Authors:  Jacek H Juściński; Andrzej Koprowski; Magdalena Kołaczkowska; Maciej M Kowalik; Jan A Rogowski; James S Rankin
Journal:  Kardiochir Torakochirurgia Pol       Date:  2018-03-28

2.  Bicuspid aortic valve repair using geometric ring annuloplasty: A first-in-humans pilot trial.

Authors:  J Scott Rankin; Domenico Mazzitelli; Theodor J M Fischlein; Yeong-Hoon Choi; Diana Aicher; Lawrence M Wei; Vinay Badhwar
Journal:  JTCVS Tech       Date:  2020-01-23