Minoru Tomita1, Miyuki Watabe2, Satoshi Yukawa2, Nobuo Nakamura2, Tadayuki Nakamura2, Thomas Magnago2. 1. From Shinagawa LASIK Center (Tomita, Watabe, Yukawa, N. Nakamura, T. Nakamura), Chiyoda-ku, Tokyo, Japan; Wenzhou University (Tomita), Wenzhou, China; Schwind eye-tech-solutions GmbH (Magnano), Kleinostheim, Germany. Electronic address: tomita@shinagawa-lasik.com. 2. From Shinagawa LASIK Center (Tomita, Watabe, Yukawa, N. Nakamura, T. Nakamura), Chiyoda-ku, Tokyo, Japan; Wenzhou University (Tomita), Wenzhou, China; Schwind eye-tech-solutions GmbH (Magnano), Kleinostheim, Germany.
Abstract
PURPOSE: To evaluate the clinical outcomes of laser in situ keratomileusis (LASIK) to correct myopia or myopic astigmatism using the Amaris 750S 750 Hz excimer laser. SETTING: Private LASIK center, Tokyo, Japan. DESIGN: Case series. METHODS: Patients with myopia or myopic astigmatism (spherical equivalent -0.50 to -11.63 diopters [D]), a corrected distance visual acuity (CDVA) of 20/20 or better, and an estimated residual bed thickness of 300 μm or more had LASIK using the aspheric aberration-free ablation profile of the 750 Hz scanning-spot laser and the Femto LDV Crystal Line femtosecond laser for flap creation. Study parameters included uncorrected distance visual acuity (UDVA), CDVA, manifest refraction, astigmatism, and higher-order aberrations (HOAs). RESULTS: The study included 1280 eyes (685 patients). At 3 months, 96.6% of eyes had a UDVA of 20/20 or better and 99.1% had 20/32 or better; 94.1% of eyes were within ± 0.50 D of the intended correction and 98.9% were within ± 1.00 D; 89.7% of eyes had no residual cylinder and 96.0% had a postoperative astigmatism of less than 0.50 D. All eyes had a postoperative CDVA of 20/20 or better. The HOAs increased postoperatively (P<.001), with mean total postoperative corneal and ocular HOAs of 0.66 μm ± 0.20 (SD) and 0.56 ± 0.23 μm, respectively. The efficacy index and safety index were 1.02 and 1.06, respectively. CONCLUSION: Laser in situ keratomileusis with the 750 Hz scanning-spot laser was safe, effective, and predictable. No specific clinical side effects that might be associated with a high repetition rate occurred. FINANCIAL DISCLOSURE: Mr. Magnago is an employee of Schwind eye-tech-solutions GmbH. No other author has a financial or proprietary interest in any material or method mentioned.
PURPOSE: To evaluate the clinical outcomes of laser in situ keratomileusis (LASIK) to correct myopia or myopic astigmatism using the Amaris 750S 750 Hz excimer laser. SETTING: Private LASIK center, Tokyo, Japan. DESIGN: Case series. METHODS:Patients with myopia or myopic astigmatism (spherical equivalent -0.50 to -11.63 diopters [D]), a corrected distance visual acuity (CDVA) of 20/20 or better, and an estimated residual bed thickness of 300 μm or more had LASIK using the aspheric aberration-free ablation profile of the 750 Hz scanning-spot laser and the Femto LDV Crystal Line femtosecond laser for flap creation. Study parameters included uncorrected distance visual acuity (UDVA), CDVA, manifest refraction, astigmatism, and higher-order aberrations (HOAs). RESULTS: The study included 1280 eyes (685 patients). At 3 months, 96.6% of eyes had a UDVA of 20/20 or better and 99.1% had 20/32 or better; 94.1% of eyes were within ± 0.50 D of the intended correction and 98.9% were within ± 1.00 D; 89.7% of eyes had no residual cylinder and 96.0% had a postoperative astigmatism of less than 0.50 D. All eyes had a postoperative CDVA of 20/20 or better. The HOAs increased postoperatively (P<.001), with mean total postoperative corneal and ocular HOAs of 0.66 μm ± 0.20 (SD) and 0.56 ± 0.23 μm, respectively. The efficacy index and safety index were 1.02 and 1.06, respectively. CONCLUSION: Laser in situ keratomileusis with the 750 Hz scanning-spot laser was safe, effective, and predictable. No specific clinical side effects that might be associated with a high repetition rate occurred. FINANCIAL DISCLOSURE: Mr. Magnago is an employee of Schwind eye-tech-solutions GmbH. No other author has a financial or proprietary interest in any material or method mentioned.