Generation of virtual monochromatic CBCT from dual kV∕MV beam projections.

Hao Li1, Bo Liu, Fang-Fang Yin.   

Abstract

PURPOSE: To develop a novel on-board imaging technique which allows generation of virtual monochromatic (VM) cone-beam CT (CBCT) with a selected energy from combined kilovoltage (kV)∕megavoltage (MV) beam projections.
METHODS: With the current orthogonal kV∕MV imaging hardware equipped in modern linear accelerators, both MV projections (from gantry angle of 0°-100°) and kV projections (90°-200°) were acquired as gantry rotated a total of 110°. A selected range of overlap projections between 90° to 100° were then decomposed into two material projections using experimentally determined parameters from orthogonally stacked aluminum and acrylic step-wedges. Given attenuation coefficients of aluminum and acrylic at a predetermined energy, one set of VM projections could be synthesized from two corresponding sets of decomposed projections. Two linear functions were generated using projection information at overlap angles to convert kV and MV projections at nonoverlap angles to approximate VM projections for CBCT reconstruction. The contrast-to-noise ratios (CNRs) were calculated for different inserts in VM CBCTs of a CatPhan phantom with various selected energies and compared with those in kV and MV CBCTs. The effect of overlap projection number on CNR was evaluated. Additionally, the effect of beam orientation was studied by scanning the CatPhan sandwiched with two 5 cm solid-water phantoms on both lateral sides and an electronic density phantom with two metal bolt inserts.
RESULTS: Proper selection of VM energy [30 and 40 keV for low-density polyethylene (LDPE), polymethylpentene, 2 MeV for Delrin] provided comparable or even better CNR results as compared with kV or MV CBCT. An increased number of overlap kV and MV projection demonstrated only marginal improvements of CNR for different inserts (with the exception of LDPE) and therefore one projection overlap was found to be sufficient for the CatPhan study. It was also evident that the optimal CBCT image quality was achieved when MV beams penetrated through the heavy attenuation direction of the object.
CONCLUSIONS: A novel technique was developed to generate VM CBCTs from kV∕MV projections. This technique has the potential to improve CNR at selected VM energies and to suppress artifacts at appropriate beam orientations.

Entities:  

Mesh:

Year:  2013        PMID: 24320521     DOI: 10.1118/1.4824324

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Metal artifact correction for x-ray computed tomography using kV and selective MV imaging.

Authors:  Meng Wu; Andreas Keil; Dragos Constantin; Josh Star-Lack; Lei Zhu; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

2.  Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Phys Med Biol       Date:  2022-08-09       Impact factor: 4.174

3.  Metal artifact reduction through MVCBCT and kVCT in radiotherapy.

Authors:  Gao Liugang; Sun Hongfei; Ni Xinye; Fang Mingming; Cao Zheng; Lin Tao
Journal:  Sci Rep       Date:  2016-11-21       Impact factor: 4.379

Review 4.  Technical Principles of Dual-Energy Cone Beam Computed Tomography and Clinical Applications for Radiation Therapy.

Authors:  Shailaja Sajja; Young Lee; Markus Eriksson; Håkan Nordström; Arjun Sahgal; Masoud Hashemi; James G Mainprize; Mark Ruschin
Journal:  Adv Radiat Oncol       Date:  2019-07-30
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.