Motion-map constrained image reconstruction (MCIR): application to four-dimensional cone-beam computed tomography.

Justin C Park1, Jin Sung Kim, Sung Ho Park, Zhaowei Liu, Bongyong Song, William Y Song.   

Abstract

PURPOSE: Utilization of respiratory correlated four-dimensional cone-beam computed tomography (4DCBCT) has enabled verification of internal target motion and volume immediately prior to treatment. However, with current standard CBCT scan, 4DCBCT poses challenge for reconstruction due to the fact that multiple phase binning leads to insufficient number of projection data to reconstruct and thus cause streaking artifacts. The purpose of this study is to develop a novel 4DCBCT reconstruction algorithm framework called motion-map constrained image reconstruction (MCIR), that allows reconstruction of high quality and high phase resolution 4DCBCT images with no more than the imaging dose as well as projections used in a standard free breathing 3DCBCT (FB-3DCBCT) scan.
METHODS: The unknown 4DCBCT volume at each phase was mathematically modeled as a combination of FB-3DCBCT and phase-specific update vector which has an associated motion-map matrix. The motion-map matrix, which is the key innovation of the MCIR algorithm, was defined as the matrix that distinguishes voxels that are moving from stationary ones. This 4DCBCT model was then reconstructed with compressed sensing (CS) reconstruction framework such that the voxels with high motion would be aggressively updated by the phase-wise sorted projections and the voxels with less motion would be minimally updated to preserve the FB-3DCBCT. To evaluate the performance of our proposed MCIR algorithm, we evaluated both numerical phantoms and a lung cancer patient. The results were then compared with the (1) clinical FB-3DCBCT reconstructed using the FDK, (2) 4DCBCT reconstructed using the FDK, and (3) 4DCBCT reconstructed using the well-known prior image constrained compressed sensing (PICCS).
RESULTS: Examination of the MCIR algorithm showed that high phase-resolved 4DCBCT with sets of up to 20 phases using a typical FB-3DCBCT scan could be reconstructed without compromising the image quality. Moreover, in comparison with other published algorithms, the image quality of the MCIR algorithm is shown to be excellent.
CONCLUSIONS: This work demonstrates the potential for providing high-quality 4DCBCT during on-line image-guided radiation therapy (IGRT), without increasing the imaging dose. The results showed that (at least) 20 phase images could be reconstructed using the same projections data, used to reconstruct a single FB-3DCBCT, without streak artifacts that are caused by insufficient projections.

Entities:  

Mesh:

Year:  2013        PMID: 24320496     DOI: 10.1118/1.4829504

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  SPARE: Sparse-view reconstruction challenge for 4D cone-beam CT from a 1-min scan.

Authors:  Chun-Chien Shieh; Yesenia Gonzalez; Bin Li; Xun Jia; Simon Rit; Cyril Mory; Matthew Riblett; Geoffrey Hugo; Yawei Zhang; Zhuoran Jiang; Xiaoning Liu; Lei Ren; Paul Keall
Journal:  Med Phys       Date:  2019-07-19       Impact factor: 4.071

2.  A new scheme for real-time high-contrast imaging in lung cancer radiotherapy: a proof-of-concept study.

Authors:  Hao Yan; Zhen Tian; Yiping Shao; Steve B Jiang; Xun Jia
Journal:  Phys Med Biol       Date:  2016-03-04       Impact factor: 3.609

3.  Four dimensional digital tomosynthesis using on-board imager for the verification of respiratory motion.

Authors:  Justin C Park; Jin Sung Kim; Sung Ho Park; Matthew J Webster; Soyoung Lee; William Y Song; Youngyih Han
Journal:  PLoS One       Date:  2014-12-26       Impact factor: 3.240

4.  An efficient iterative CBCT reconstruction approach using gradient projection sparse reconstruction algorithm.

Authors:  Heui Chang Lee; Bongyong Song; Jin Sung Kim; James J Jung; H. Harold Li; Sasa Mutic; Justin C Park
Journal:  Oncotarget       Date:  2016-12-27

5.  Technical Note: 4D cone-beam CT reconstruction from sparse-view CBCT data for daily motion assessment in pencil beam scanned proton therapy (PBS-PT).

Authors:  Lydia A den Otter; Kuanling Chen; Guillaume Janssens; Arturs Meijers; Stefan Both; Johannes A Langendijk; Lane R Rosen; Hsinshun T Wu; Antje-Christin Knopf
Journal:  Med Phys       Date:  2020-10-24       Impact factor: 4.071

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.