The use of computed radiography plates to determine light and radiation field coincidence.

James R Kerns1, Aman Anand.   

Abstract

PURPOSE: Photo-stimulable phosphor computed radiography (CR) has characteristics that allow the output to be manipulated by both radiation and optical light. The authors have developed a method that uses these characteristics to carry out radiation field and light field coincidence quality assurance on linear accelerators.
METHODS: CR detectors from Kodak were used outside their cassettes to measure both radiation and light field edges from a Varian linear accelerator. The CR detector was first exposed to a radiation field and then to a slightly smaller light field. The light impinged on the detector's latent image, removing to an extent the portion exposed to the light field. The detector was then digitally scanned. A MATLAB-based algorithm was developed to automatically analyze the images and determine the edges of the light and radiation fields, the vector between the field centers, and the crosshair center. Radiographic film was also used as a control to confirm the radiation field size.
RESULTS: Analysis showed a high degree of repeatability with the proposed method. Results between the proposed method and radiographic film showed excellent agreement of the radiation field. The effect of varying monitor units and light exposure time was tested and found to be very small. Radiation and light field sizes were determined with an uncertainty of less than 1 mm, and light and crosshair centers were determined within 0.1 mm.
CONCLUSIONS: A new method was developed to digitally determine the radiation and light field size using CR photo-stimulable phosphor plates. The method is quick and reproducible, allowing for the streamlined and robust assessment of light and radiation field coincidence, with no observer interpretation needed.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24320415      PMCID: PMC3808487          DOI: 10.1118/1.4823775

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  A test tool for the visual verification of light and radiation fields using film or an electronic portal imaging device.

Authors:  P Dunscombe; S Humphreys; K Leszczynski
Journal:  Med Phys       Date:  1999-02       Impact factor: 4.071

2.  A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery.

Authors:  Paul A Jursinic; Ben E Nelms
Journal:  Med Phys       Date:  2003-05       Impact factor: 4.071

3.  Task Group 142 report: quality assurance of medical accelerators.

Authors:  Eric E Klein; Joseph Hanley; John Bayouth; Fang-Fang Yin; William Simon; Sean Dresser; Christopher Serago; Francisco Aguirre; Lijun Ma; Bijan Arjomandy; Chihray Liu; Carlos Sandin; Todd Holmes
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

4.  Assessing radiation and light field congruence with a video based electronic portal imaging device.

Authors:  K Luchka; D Chen; S Shalev; G Gluhchev; R Rajapakshe
Journal:  Med Phys       Date:  1996-07       Impact factor: 4.071

5.  Automated x-ray/light field congruence using the LINAC EPID panel.

Authors:  Wojciech Polak; Jim O'Doherty; Matt Jones
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

6.  On visual interpretation of light localization/radiation field coincidence films.

Authors:  A Ho; B Thomadsen; B Paliwal
Journal:  Med Phys       Date:  1995-02       Impact factor: 4.071

7.  A simple quality assurance test tool for the visual verification of light and radiation field congruent using electronic portal images device and computed radiography.

Authors:  Christopher F Njeh; Blas Caroprese; Pushkar Desai
Journal:  Radiat Oncol       Date:  2012-03-27       Impact factor: 3.481

8.  Utilizing an electronic portal imaging device to monitor light and radiation field congruence.

Authors:  J I Prisciandaro; M G Herman; J J Kruse
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

9.  Feasibility study of performing IGRT system daily QA using a commercial QA device.

Authors:  Jean L Peng; Darren Kahler; Jonathan G Li; Robert J Amdur; Kenneth N Vanek; Chihray Liu
Journal:  J Appl Clin Med Phys       Date:  2011-06-01       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.