Nanomaterials tend to agglomerate in aqueous media, resulting in inaccurate safety assessment of the biological response to these substances. The present study searched for suitable dispersion methods for the preparation of nanomaterial suspensions. Titanium dioxide (TiO2) and zinc oxide (ZnO) nanoparticles were dispersed in a biocompatible dispersion medium by direct probe-type sonicator and indirect cup-type sonicator. Size characterization was completed using dynamic light scattering and transmission electron microscopy. A series of dispersion time and output power, as well as two different particle concentrations were tested. Microscopic contamination of metal titanium that broke away from the tip of the probe into the suspension was found. Size of agglomerated nanoparticles decreased with increase in sonication time or output power. Particle concentration did not show obvious effect on size distribution of TiO2 nanoparticles, while significant reduction of secondary diameter of ZnO was observed at higher concentration. A practicable protocol was then adopted and sizes of well-dispersed nanoparticles increased by less than 10% at 7 d after sonication. Multi-walled carbon nanotubes were also well dispersed by the same protocol. The cup-type sonicator might be a useful alternative to the traditional bath-type sonicator or probe-type sonicator based on its effective energy delivery and assurance of suspension purity.
Nanomaterials tend to agglomerate in aqueous media, resulting in inaccurate safety assessment of the biological response to these substances. The present study searched for suitable dispersion methods for the preparation of nanomaterial suspensions. Titanium dioxide (TiO2) and zinc oxide (ZnO) nanoparticles were dispersed in a biocompatible dispersion medium by direct probe-type sonicator and indirect cup-type sonicator. Size characterization was completed using dynamic light scattering and transmission electron microscopy. A series of dispersion time and output power, as well as two different particle concentrations were tested. Microscopic contamination of metal titanium that broke away from the tip of the probe into the suspension was found. Size of agglomerated nanoparticles decreased with increase in sonication time or output power. Particle concentration did not show obvious effect on size distribution of TiO2 nanoparticles, while significant reduction of secondary diameter of ZnO was observed at higher concentration. A practicable protocol was then adopted and sizes of well-dispersed nanoparticles increased by less than 10% at 7 d after sonication. Multi-walled carbon nanotubes were also well dispersed by the same protocol. The cup-type sonicator might be a useful alternative to the traditional bath-type sonicator or probe-type sonicator based on its effective energy delivery and assurance of suspension purity.
According to the European Commission, nanomaterial is defined as a material containing
particles with one or more external dimensions in the size range of 1–100 nm1). The remarkable revolution in nanotechnology
shows promising potential applications of manufactured nanomaterials in a variety of areas
including engineering, medicine, consumer, and information technology2, 3). On the other hand,
safety evaluation of manufactured nanomaterials is urgently required because of the
considerable probability of occupational and environmental exposure throughout the product
chain during manufacture, application and waste management2, 4). Furthermore, there is
increasing concern that the novel physico-chemical properties of nanomaterials that are
different from those of bulk materials might have unpredictable health effects5, 6).Both in vivo and in vitro studies demonstrated that the higher toxicity of nanomaterials
was correlated with the relatively larger surface area derived from their smaller sizes7,8,9,10,11). However, nanomaterials tend to
agglomerate into micrometer-sized structures in aqueous media, subsequently exerting
different biological effects compared to well-dispersed ones, probably due to the
ineffective surface area delivered by the poor dispersed suspension12,13,14,15). Therefore, the
selection of suitable dispersion methods is important for accurate evaluation of
toxicological responses to nanomaterials.In safety research on nanomaterials, sonication is the most widely used procedure to
accomplish a suspension16,17,18,19,20). Cavitation from
alternating high and low pressure cycles provides energy that facilitates the disruption of
agglomerates. Agglomerates of particles bound together by relatively weak forces, such as
Van der Waals forces, can be fragmented by sonication, while aggregates created by stronger
chemical bonds are difficult to be broken down21). Instead of electrostatic stabilization, steric stabilization by
adding stabilizers into the suspension is often used to maintain homogeneous dispersion
induced by sonication. Pulmonary surfactant, albumin, cell culture medium, and serum have
been reported to prevent nanomaterials from approaching each other by formation of a protein
corona, contributing to the suppression of gross cluster formation and the improvement of
suspension stability13, 22, 23).Recent studies have shown that exposure to various kinds of nanomaterials, such as metal
oxide nanoparticles, carbon nanoparticles and nanotubes is injurious to biological
systems24,25,26,27,28,29). In addition to the gastrointestinal track and skin, the
respiratory system is recognized as one of the most important portals of entry and a target
tissue, and a number of in vivo research studies have focused on the
respiratory system to investigate the harmful effects of airborne nanomaterials14, 30, 31). For the evaluation of respiratory
toxicity, intratracheal instillation/spary and pharyngeal aspiration are useful and
cost-effective exposure techniques routinely used in investigative exposure of animals to
particles32,33,34,35,36,37). The dispersion medium (DM), an artificial physiological buffer
comprised of protein and surfactant components naturally found in lung alveolar fluids,
which are known to be effective in both reduction of agglomeration and stabilization of
suspensions, was recommended in the preparation of nanomaterials suspension in pulmonary
exposure studies using intratracheal instillation/spray or pharyngeal aspiration38).The aim of the present study was to investigate the factors that influence the dispersion
status and establish a suitable and reproducible protocol for the preparation of
nanomaterial suspensions specific for pulmonary exposure studies. Two kinds of different
nanospheres and one kind of carbon nanotube were dispersed in DM by a direct probe-type
sonicator or indirect cup-type sonicator. A series of sonication times and output powers, as
well as two different dispersion concentrations of particle suspension were tested. Size
characterization was completed by dynamic light scattering (DLS) and further examined
morphologically with a transmission electron microscope (TEM). Dispersion stability over
time was also assessed.
Subjects and Methods
Nanomaterials
Titanium dioxide (TiO2) nanoparticles (AEROXIDETiO2 P25; Degussa
AG, Dusseldorf, Germany) with a primary diameter of 21 nm, zinc oxide (ZnO) nanoparticles
(MKN-ZnO-020; mkNano, Missisauga, ONT, Canada) with a primary diameter of 20 nm, and
multi-walled carbon nanotubes (MITSUI MWCNT-7; Mitsui, Tokyo, Japan) with a diameter of
100 nm and length of 27% >5 μm were used in the present study. The selection of 20 nm
for ZnO nanoparticles was based on the similarity of the diameter to that of P25
TiO2 nanoparticles. The nanomaterials were stored in 50-ml polypropylene
conical tubes until use.
Dispersion medium (DM)
Porter et al.38)
demonstrated that DM neither caused pulmonary inflammation or cytotoxicity, nor altered
the toxicity of tested silica particles, suggesting that DM is an effective, biocompatible
and economical vehicle for nanotoxicological evaluation. DM was prepared as described
previously. Briefly, Ca2+ and Mg2+ -free phosphate buffered saline
(PBS) was supplemented with 5.5 mM D-glucose, 0.6 mg/ml bovineserum albumin (BSA), and
0.01 mg/ml 1, 2-dipalmitoyl-sn-glycero-3-phosphocholine (DPPC), and sonicated for 10 min
with continuous 70 W output power by bath-type sonicator after mixing. DM was prepared
within 2 days before dispersion and stored in a refrigerator at 4°C.
Suspension
Dispersion of nanomaterial suspensions were performed in a 50-ml polypropylene conical
tube. In terms of probe-type sonicator, we used the ultrasonic system (Sonifier 450
Advances; Branson, Danbury, CT) with a horn having a tip diameter of 1/2” (13 mm).
Particles were dispersed in 20 ml of DM and the tube was immersed in an ice-salt bath
during sonication. For the cup-type sonicator (cup horn; Branson), a circulatory cooling
system was used to avoid over heating of the suspension. When the tube is immersed into
the circulatory cooling water in the sonicator, the maximum volume of suspension in the
tube to be covered by the cooling water is 8 ml, and therefore the volume of DM was
adjusted to 8 ml. Output power was determined by the sonicator dial according to the
manual provided by the manufacturer.
Size characterization
Soon after sonication, dispersed suspensions of 25 µg/ml TiO2
(absorbance: 0.010; refractive index: 2.520) and 500 µg/ml ZnO
(absorbance: 0.010; refractive index: 2.020) were prepared in DM (viscosity: 1.0;
refractive index: 1.33), and then vortexed and filtered through a 0.6 µm
isopore membrane filter (Merk Millipore, Billerica, MA) into a disposable standard
cuvette. Each sample was measured by dynamic light scattering (DLS) (Zetasizer Nano-S;
Malvern Instruments, Worcestershire, UK) at 25°C four times after 1 h on standing.
Dispersion status was described as the peak value of the principal peak of the
intensity-weighted hydrodynamic diameter distribution (peak 1), in conjunction with
polydispersity index (PdI), which reflected the broadness of the size distribution (scale
range from 0 to 1, with 0 being monodispersion and 1 being polydispersion)22).To check the stability of the suspension, size characterization was conducted soon after,
and at 1, 3, and 7 days post-sonication (ZnO was not measured at 3 days post-sonication
because of inadequacy of samples). The prepared dispersions were stored in a refrigerator
at 4°C until measured, and after vortexing for several seconds they were subsequently
analyzed by DLS as mentioned previously.
Morphological characterization
A drop (approximately 0.1 ml) of each sample was deposited onto an elastic carbon-coated
copper grid or high-resolution carbon-coated copper grid to observe the dispersed
TiO2 or MWCNT, respectively, and allowed to air dry. A transmission electron
microscope (TEM, JEM-1011; JEOL, Tokyo) was used to visualize the dispersed nanomaterials.
MWCNT suspensions were also viewed using an Olympus BXJ1 optical microscope (OLYMPUS,
Tokyo) equipped with a digital camera DP70, to capture images with the DP controller
software (OLYMPUS).
Results
Direct probe-type sonicator
Particle concentration for DLS measurement
Suspensions of TiO2 and ZnO nanoparticles at a concentration of 0.5 mg/ml
were dispersed by probe-type sonicator at 20 W, 80% pulse mode, for 5, 10, 20, or
30 min. The sample dispersed for 30 min was diluted with DM into concentrations of 10,
25, 50, 100, 200 and 500 µg/ml (ZnO nanoparticles were tested at
concentrations of 100, 200 and 500 µg/ml), and peak1 and PdI values
were determined for each concentration sample (Fig.
1
).
Fig. 1.
Size characterization of TiO2 and ZnO nanoparticle suspensions at
concentration of 0.5 mg/ml dispersed by a probe-type sonicator at 20 W, 80% pulse
mode, for 30 min, being measured by DLS at concentrations of 10, 25, 50, 100, 200,
and 500 g/ml: (a) peak 1 and (b) PdI. Data are mean ±
SD.
Size characterization of TiO2 and ZnO nanoparticle suspensions at
concentration of 0.5 mg/ml dispersed by a probe-type sonicator at 20 W, 80% pulse
mode, for 30 min, being measured by DLS at concentrations of 10, 25, 50, 100, 200,
and 500 g/ml: (a) peak 1 and (b) PdI. Data are mean ±
SD.TiO2 nanoparticles showed the smallest peak1 and PdI at the concentration of
25 µg/ml. At particle concentrations ranging from 25 to 500
µg/ml, the secondary hydrodynamic diameter increased with elevated
concentration. The coefficient of variation of peak1 was 12.63, 5.18, 19.26, 25.50,
42.24 and 34.82% at concentrations of 10, 25, 50, 100, 200 and 500
µg/ml, respectively, indicating the least deviation of size
distribution at the concentration of 25 µg/ml.On the contrary, peak 1 and its standard deviation of ZnO nanoparticles slightly
decreased with elevated particle concentration, and PdI of ZnO also decreased depending
on the particle concentration to a greater extent than the peak 1 of ZnO. The size
distribution provided by DLS (Fig. 2) showed
unimodal distribution at the concentration of 500 µg/ml, while there
was a small sub-peak around 8 nm at 100 µg/ml. The size distribution of
DM was measured to confirm that the peak around 8 nm was derived from DM (data not
shown).
Fig. 2.
Report of size distribution of ZnO nanoparticles suspensions dispersed by a
probe-type sonicator at 20 W, 80% pulse mode, for 30 min, at measure concentrations
of: (a) 100 and (b) 500 g/ml.
Report of size distribution of ZnO nanoparticles suspensions dispersed by a
probe-type sonicator at 20 W, 80% pulse mode, for 30 min, at measure concentrations
of: (a) 100 and (b) 500 g/ml.
Dispersion time
Suspensions of TiO2 and ZnO nanoparticles at concentration of 0.5 mg/ml were
dispersed by probe-type sonicator at 20 W, 80% pulse mode, for 5, 10, 20, or 30 min.
With increased dispersion time, DLS measurement showed a decrease in hydrodynamic
diameter of both TiO2 and ZnO nanoparticles, as well as tendency toward
monodispersion of TiO2 nanoparticles and decreased standard deviation of PdI
of ZnO nanoparticles (Fig. 3). TEM showed the
presence of smaller TiO2 nanoparticle agglomerates in 30-min sonicated
samples (Fig. 4 (b)) compared with the large
agglomerates found after sonication for 10 min (Fig.
4 (a)). These results suggest that longer dispersion time allowed the delivery
of more energy to break the bonds within the agglomerates. After filtration, poor
dispersion suspension (5 or 10 min) appeared more limpid, probably due to the retention
of larger number of micro-sized agglomerates on the filter membrane. In comparison,
samples sonicated for 30 min appeared turbid and this indicated that the majority of
small agglomerates distributed well in the dispersed suspension (Fig. 5).
Fig. 3.
Size characterization of TiO2 and ZnO nanoparticle suspensions at
concentration of 0.5 mg/ml dispersed by a probe-type sonicator at 20 W, 80% pulse
mode for, 5, 10, 20, or 30 min: (a) peak 1 and (b) PdI. Data are mean ± SD.
Fig. 4.
TEM micrographs of TiO2 nanoparticle suspensions at concentration of
0.5 mg/ml dispersed by (A) a probe-type sonicator at 20 W, 80% pulse mode, for (a)
10 min and (b) 30 min; (B) a cup-type sonicator at (c) 50 W and (d) 100W, 80% pulse
mode, for 10 min.
Fig. 5.
Change in the turbidity of filtered TiO2 nanoparticle suspensions
dispersed by a probe-type sonicator at 20 W, 80% pulse mode, for 5, 10, 20, or
30 min at concentration of 0.5 mg/ml.
Size characterization of TiO2 and ZnO nanoparticle suspensions at
concentration of 0.5 mg/ml dispersed by a probe-type sonicator at 20 W, 80% pulse
mode for, 5, 10, 20, or 30 min: (a) peak 1 and (b) PdI. Data are mean ± SD.TEM micrographs of TiO2 nanoparticle suspensions at concentration of
0.5 mg/ml dispersed by (A) a probe-type sonicator at 20 W, 80% pulse mode, for (a)
10 min and (b) 30 min; (B) a cup-type sonicator at (c) 50 W and (d) 100W, 80% pulse
mode, for 10 min.Change in the turbidity of filtered TiO2 nanoparticle suspensions
dispersed by a probe-type sonicator at 20 W, 80% pulse mode, for 5, 10, 20, or
30 min at concentration of 0.5 mg/ml.
Contamination from the probe
Black sediment was found at the bottom of the samples sonicated for 30 min after 2 h on
standing (Fig. 6 (a)). Inspection of the
surface of the probe tip showed abrasion, which was probably due to the intense
cavitation (Fig. 6 (b)). These microscopic tip
residues of titanium were presumed to be broken away from the tip into the
suspension.
Fig. 6.
(a) Black sediment at the bottom of a sample sonicated by a probe-type sonicator at
20 W, 80% pulse mode, for 30 min after 2 hours on standing and (b) the tip surface
of unused probe (left) and abrased probe (right).
(a) Black sediment at the bottom of a sample sonicated by a probe-type sonicator at
20 W, 80% pulse mode, for 30 min after 2 hours on standing and (b) the tip surface
of unused probe (left) and abrased probe (right).
Indirect cup-type sonicator
Temperature of cooling water
Suspensions of TiO2 nanoparticles at concentration of 0.5 mg/ml were
dispersed by cup-type sonicator at 220 W (max), 80% pulse mode. For the purpose of
suppression of sample heating, a circulatory cooling system consisting of a congealer
that circulated 30% ethanol with a pump at the flow rate of 400 ml/min was utilized. At
the beginning, we attempted to lower the temperature of cooling water to −10°C so that
sonication was able to be carried out with long dispersion time. When sonicated for
40 min, agglomerated TiO2 nanoparticles were dispersed into 371.8 ± 12.5 nm
with PdI of 0.24 ± 0.01 when the circulation temperature was set at −10°C, and the
sample temperature was measured lower than the room temperature (25°C) after dispersion.
In contrast, when the circulation temperature was set at 5°C, particles were dispersed
into 173.9 ± 12.8 nm with PdI of 0.15 ± 0.01 after 10-minute sonication, and the sample
was heated to 37.6°C. In spite of the shorter dispersion time, sonication at a
relatively higher temperature (5°C) produced better monodispersion of small
agglomerates, suggesting that the cup-type sonication system should be cooled to
moderate temperatures only. Since the sample was extremely calefacient at dispersion
time longer than 10 min, we chose a dispersion time of 10 min for the suspension using
the cup-type sonicator.
Output power of sonicator
Suspensions of TiO2 and ZnO nanoparticles at concentration of 0.5 mg/ml were
dispersed by the cup-type sonicator at 10, 20, 50, 100, or 220 W, 80% pulse mode, for
10 min (Fig. 7
). The size of the sonicated
TiO2 nanoparticles appeared to decrease with increased output power,
although the maximum reduction reached a plateau level at output power of 100 W, and a
similar trend was noted for PdI. More energy delivered with higher output power
facilitated disruption of agglomeration and output power of 100W was deemed likely to be
adequate for dispersion of TiO2 nanoparticles. TEM micrographs of the samples
sonicated at 50 or 100 W confirmed the results of DLS (Fig. 4).
Fig. 7.
Size characterization of TiO2 and ZnO nanoparticle suspensions at
concentration of 0.5 mg/ml dispersed by a cup-type sonicator at 10, 20, 50, 100, or
220 W, 80% pulse mode, for 10 min: (a) peak 1 and (b) PdI. Data are mean ± SD.
Size characterization of TiO2 and ZnO nanoparticle suspensions at
concentration of 0.5 mg/ml dispersed by a cup-type sonicator at 10, 20, 50, 100, or
220 W, 80% pulse mode, for 10 min: (a) peak 1 and (b) PdI. Data are mean ± SD.Surprisingly, output power ranging from 10 to 220 W had little effect on sizes of
agglomerated ZnO nanoparticles, while PdI showed an output power-dependent decrease from
10 to 50 W. The energy delivered with output power of 10 W might be enough for
agglomerate breakage, but output power less than 50 W was insufficient to accomplish
monodispersion.
Determination of dispersion time for nanoparticles
We also compared the size distribution between samples sonicated at 220 W, 80% pulse
mode for 10 min and that of samples sonicated several times each for 10 min, with sample
cooling in between the sonications. The size of agglomerated TiO2
nanoparticles sonicated four times each for 10 min was 182.2 ± 0.6 nm, which was similar
to 173.9 ± 12.8 nm obtained after a single sonication. Regarding ZnO nanoparticles, one
single sonication for 10 min dispersed the agglomerates into 180.2 ± 5.0 nm and
three-time sonication also yielded a similar result of 177.5 ± 1.5 nm. This phenomenon
was deduced to be the result of the balance between deagglomeration and reagglomeration
taking place during the process of delivered energy elevation21). It seemed that sonication at 220 W, 80% pulse mode,
for 10 min was favorable for the dispersion of these two nanoparticles.
Applicability
Particle concentration for dispersion
Suspensions of TiO2 and ZnO nanoparticles at concentrations of 0.5 or
2.5 mg/ml were dispersed by the cup-type sonicator at 220 W, 80% pulse mode, for 10 min.
At the concentration of 2.5 mg/ml, TiO2 nanoparticles agglomerated to a
slightly larger size with larger standard deviation, while significant reduction of the
secondary diameter of ZnO nanoparticles was found with comparison to the outcome of
0.5 mg/ml samples (Table 1).
Table 1.
Effect of particle concentration for dispersion
Nanoparticle
Concentration (mg/ml)
0.5
2.5
Peak 1 (nm)
PdI
Peak 1 (nm)
PdI
TiO2
173.9 ± 12.8
0.12 ± 0.02
180.8 ± 31.4
0.13 ± 0.02
ZnO
180.2 ± 5.0
0.18 ± 0.01
169.8 ± 3.1*
0.17 ± 0.02
Data are mean ± SD; * p<0.05.
Data are mean ± SD; * p<0.05.
Stability of suspension
Suspensions of TiO2 and ZnO nanoparticles at concentration of 0.5 mg/ml were
dispersed by a cup-type sonicator at 100 or 220 W, 80% pulse mode, for 10 min. The size
distributions conducted soon after, and at 1, 3, and 7 days post-sonication are listed
in Table 2. The secondary diameter of
TiO2 nanoparticle agglomerates increased by 4.7%, 7.8% and 9.2% when
dispersed at 100 W, and by 1.5%, 5.9% and 4.5% at 220 W, respectively. As to ZnO
nanoparticles, the secondary diameter increased by 1.9% and 3.4% (1 and 7 days
post-sonication, respectively) at 100 W, and by 1.3% and 0.6% at 220 W.
Table 2.
Stability of nanoparticle suspensions
Particles
Output power
Peak 1 (nm)
0 d
1 d
3 d
7 d
TiO2
100 W
174.0 ± 12.3
182.1 ± 14.6
187.5 ± 19.2
190.0 ± 20.6
220 W
173.9 ± 12.8
176.5 ± 13.6
184.1 ± 11.8
181.7 ± 16.2
ZnO
100 W
177.2 ± 4.7
180.6 ± 4.7
/
183.3 ± 4.9
220 W
180.2 ± 5.0
182.5 ± 4.1
/
179.2 ± 4.5
Data are mean ± SD.
Data are mean ± SD.
Dispersion of MWCNTs
Dispersion of suspensions of MWCNTs at a concentration of 2.0 mg/ml was operated using
the protocol: cup-type sonicator at 100 W, 80% pulse mode, for 10 min. The dispersion
status was assessed using optical microscopy and TEM. Highly agglomerated masses of
MWCNTs were dispersed into smaller size clusters after a single dispersion process
(Fig. 8 (a) and (b)) and MWCNTs bundles were
separated homogeneously by sonication for another 10 min (Fig. 8 (c)). Individual and bundled nanotubes were observed in
samples sonicated twice in the TEM micrographs (Fig.
8 (d)).
Fig. 8.
Optical microscope micrographs of MWCNT suspensions at concentration of 2.0 mg/ml:
(a) un-dispersed, (b) dispersed by a cup-type sonicator at 100 W, 80% pulse mode,
for 10 min, and (c) for another 10 min; and (d) TEM micrograph of MWCNT sonicated
twice.
Optical microscope micrographs of MWCNT suspensions at concentration of 2.0 mg/ml:
(a) un-dispersed, (b) dispersed by a cup-type sonicator at 100 W, 80% pulse mode,
for 10 min, and (c) for another 10 min; and (d) TEM micrograph of MWCNT sonicated
twice.
Discussion
The present study used direct probe-type and indirect cup-type sonicator to prepare
nanomaterials suspensions and subsequent size characterization was compeleted by DLS
instrument and TEM. The effects of factors including particle concentration, dispersion time
and output power of sonication on the dispersion status were investigated. Then a protocol
was established and demonstrated to be suitable and reproducible for nanomaterials
dispersion in terms of both size distribution and suspension stability.Prior to the evaluation of dispersion condition, adjustment of appropriate particle
concentration for DLS measurement was performed following the recommendations of the
instrument instructions. According to the instructions, particle concentration is a factor
for the accuracy of measurement, and sufficient opaqueness is suggested to accommodate the
optical requirements of DLS. TiO2 nanoparticles showed the smallest hydrodynamic
diameter and deviation at the concentration of 25 µg/ml. At higher or lower
concentrations than 25 µg/m, variability of hydrodynamic diameter became
greater, probably due to instable translational diffusion coefficient of Brownian motion
resulting from changes in particle interaction. Although further study is needed to ensure
whether the measured diameter was concentration dependent or not, the concentration of 25
µg/ml was optimal for size characterization of TiO2
nanoparticles with regard to the least deviation. By contrast, smaller peak 1 and especially
lower PdI were observed in samples of higher concentration of ZnO nanoparticles. The higher
PdI value derived from lower concentrations can be explained by the bimodal size
distribution. Because of the similar refractive index between ZnO nanoparticals and DM, the
light scattering intensity of DM turned out to be a certain proportion of the total recorded
intensity and the quality of resulting DLS data was too poor to be interpreted correctly. In
this situation, the light scattering intensity of nanoparticles could be enhanced by using
elevated particle concentration, and good quality of DLS data was obtained at the
concentration of 500 µg/ml in the present study. In case of size
distribution with more than one peak, z-average, which is calculated from total detected
intensity, is not suitable for description of particle size distribution. It is therefore
advisable to use peak 1, the peak value of the principal peak of the intensity-weighted
hydrodynamic diameter distribution, to compare the outcomes of DLS. It is worth noting that
determination of proper concentration should be conducted for each nanomaterial and the
selected concentration should be reported.Direct probe-type sonicator is usually recommended over indirect bath/cup-type sonicator
based on the higher efficacy in energy delivery into the suspension, without energy loss
generated from the process that ultrasonic waves pass through both the bath/cup liquid and
the wall of the sample container, and thus is widely used in safety research on
nanomaterials4, 10, 39, 40). However, large variance is often encountered due to technical
problems such as uncertainty in probe immersion position. Moreover, direct immersion of the
probe into the suspension could introduce contamination of the sample at the same time.
Although such contamination could be avoided, an unavoidable side effect of the probe-type
is tip erosion, which was observed in our study. In addition, tip erosion also induces
reduction of energy21), resulting in
subsequent alteration of dispersion condition that is critical to data reproducibility.
Furthermore, samples are usually left on the bench in uncovered containers. Thus, the
evaporative loss of liquid content and deposition of dust should be taken into consideration
as well.On account of the aforementioned problems with probe-type sonicator, the cup-type sonicator
was utilized for further analysis. Effective energy delivery was achieved by minimal
interspace between the sample container and the internal layer of the cup, which contributed
to significant minimization of energy loss. Simultaneously, due to the massive collapse of
the bubbles and the high local energy generated, excessive heating cycles occur at the
interface of the explosion. A rapid rise in liquid temperature is an inherent effect in both
direct and indirect sonication21). In
order to avoid overheating of the suspension along with consequent liquid evaporation or
degradation of the medium components, a circulatory cooling system was used in this study.
Smaller agglomerates and better monodispersion of TiO2 nanoparticles was produced
within shorter dispersion time by sonication at cooling water temperature of 5°C rather than
−10°C. It is speculated that if the cooling water is set at an extremely low temperature,
the cup-type sonicator may be frozen by the circulation, and this may induce inefficiency of
sonication, resulting in subsequent poor dispersion of TiO2 nanoparticle
suspension. Moreover, effect of medium temperature on the particle interaction and size
distribution needs to be studied further.Besides the impact of temperature of cooling water, dispersion time and output power of
sonicator play important roles in nanomaterials dispersion as well. Decrease in the
secondary dynamic diameters of TiO2 and ZnO nanoparticles resulted from increased
dispersion time or output power using probe-type or cup-type sonicator, respectively. This
suggested that more delivered energy which was derived from increased dispersion time or
output power of sonicator contributed to break the bonds within agglomerates and disrupt the
formation of agglomeration. The asymptotic behavior of the size change with output power of
sonicator observed in the present study is consistent with the findings of previous
studies13, 41). The secondary diameters ceased to change after a critical output
power value, indicating that nanomaterials in suspensions cannot be dispersed to their
primary sizes and that the presence of at least a minimal number of agglomerates should
always be expected in safety research21, 38).In actual nanosafety research, samples with different concentrations of nanomaterials are
required for dose-response experiments. Thus, we tested the effect of particle concentration
on size distribution. TiO2 and ZnO nanoparticles were sonicated at concentration
of 0.5 or 2.5 mg/ml, and the size characterization was conducted. As a result, particle
concentration at 0.5 or 2.5 mg/ml did not show an obvious effect on size distribution of
TiO2 nanoparticles, while reduction of the secondary diameter of ZnO
nanoparticles was observed at higher concentration. In principle, higher particle
concentration leads to increased particle collision frequency which can enhance particle
breakage but also induce agglomerate formation at the same time. The discrepancy between
TiO2 and ZnO nanoparticles is due to the different physiochemical properties of
each material, including different breakage behavior and/or surface interaction with
components of the medium21, 22). Such issues need to be investigated in future studies.
The results highlight the demand to characterize the size distribution when dispersion of
specific particles is operated at different concentrations. It is also important to report
the dispersion concentration to allow comparisons between studies from different
laboratories.Stable dispersion status is usually desirable for studies spanning over days. Therefore,
the stability of TiO2 and ZnO nanoparticles suspension was assessed until the 7th
days after sonication. Agglomerates of TiO2 and ZnO nanoparticles increased
slightly by only 10% and 5%, respectively, which was acceptable for safety evaluation of
nanomaterials. Furthermore, alteration of output power from 100 to 220 W did not change the
size distribution of TiO2 and ZnO nanoparticles soon after sonication and both
prevented the reagglomeration of nanoparticles for at least 7 d after dispersion. Since the
sample was extremely heated when sonicated for more than 10 min, dispersion time was limited
to 10 min. Thus sonication can be operated with the maximum output power of 220 W for more
energy delivery, but the results of stability demonstrated that desired degree of particle
dispersion can be achieved with less output power of 100 W. In order to avoid overloading
the sonicator and to minimize unwanted side effects21), we established the best dispersion protocol as followed: cup-type
sonicator, with output power of 100 W, 80% pulse mode, for 10 min.The protocol was set up based on TiO2 and ZnO nanoparticles and further
application to dispersion of MWCNTs was checked. Size characterization of nanotubes by DLS
is not accurate21), and thus optical
microscopy and TEM were used for morphological observation. MWCNTs clusters were not able to
be disrupted after one single sonication following this protocol. After cooling, another
10-min sonication was performed and MWCNTs were dispersed into smaller bundles consequently.
As previously demonstrated in the section that described the effect of dispersion time,
higher energy was delivered into the suspension by prolonged dispersion time to break the
bonds within the agglomerates of nanoparticles, indicating that this procedure can also be
applied to nanotubes. With regard to the side effect of temperature, the dispersion time
should be extended by repeating the 10-min sonication rather than selecting continuous
operation in the present cup-type sonicator protocol.It is worth mentioning that this protocol is not for all dispersion systems for different
nanomaterials, media, sonicators, and sample containers. We here outlined the procedure and
key points for a dispersion optimization study that is essential to start nanotoxicological
research. Application of this protocol to other dispersion systems is feasible by adjustment
of dispersion time, output power of sonicator or particle concentration.
Conclusions
Our study described a practical protocol using an indirect cup-type sonicator to prepare
nanomaterial suspensions specifically for in vivo pulmonary exposure
studies. TiO2 and ZnO nanoparticles, together with MWCNTs were well dispersed in
a biocompatible medium and were stable for 7 days after sonication following this protocol.
The size distribution results were influenced by particle concentration for both sonication
and DLS measurement. Dispersion time and output power affected dispersion status with regard
to the delivered energy. Additionally, appropriate temperature of cooling system appeared to
be critical to sufficient dispersion with the cup-type sonicator. The cup-type sonicator
might be recommended as a useful alternative to conventional bath-type sonicator or
probe-type sonicator based on its effective energy delivery and assurance of suspension
purity.
Authors: Michael P Holsapple; William H Farland; Timothy D Landry; Nancy A Monteiro-Riviere; Janet M Carter; Nigel J Walker; Karluss V Thomas Journal: Toxicol Sci Date: 2005-08-24 Impact factor: 4.849
Authors: Peter Wick; Pius Manser; Ludwig K Limbach; Ursula Dettlaff-Weglikowska; Frank Krumeich; Siegmar Roth; Wendelin J Stark; Arie Bruinink Journal: Toxicol Lett Date: 2006-11-19 Impact factor: 4.372
Authors: Peter Bihari; Minnamari Vippola; Stephan Schultes; Marc Praetner; Alexander G Khandoga; Christoph A Reichel; Conrad Coester; Timo Tuomi; Markus Rehberg; Fritz Krombach Journal: Part Fibre Toxicol Date: 2008-11-06 Impact factor: 9.400
Authors: Ikjot Singh Sohal; Kevin S O'Fallon; Peter Gaines; Philip Demokritou; Dhimiter Bello Journal: Part Fibre Toxicol Date: 2018-07-03 Impact factor: 9.400