| Literature DB >> 24279414 |
Xiance Jin1, Ce Han, Yongqiang Zhou, Jinling Yi, Huawei Yan, Congying Xie.
Abstract
BACKGROUND: To investigate the feasibility and benefits of a modified adaptive radiotherapy (ART) by replanning in the initial CT (iCT) with new contours from a repeat CT (rCT) based on CT-CT image fusion for nasopharyngeal cancer (NPC) patients underwent volumetric modulated arc radiotherapy (VMAT).Entities:
Mesh:
Year: 2013 PMID: 24279414 PMCID: PMC4222034 DOI: 10.1186/1748-717X-8-277
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Flowchart for the whole study design. iCT is the initial CT; rCT is the repeat CT; SOiCT is the structures on the initial CT; SOrCT is the structures on the repeat CT; SOr-to-iCT is the structures propagated from the repeat CT on the initial CT.
Patients characteristics
| 1 | T2N2Mx | 59.61 | 617.99 |
| 2 | T2N1Mx | 85.49 | 757.70 |
| 3 | T3N0M0 | 62.18 | 589.08 |
| 4 | T4N0Mx | 170.96 | 546.76 |
| 5 | T3N3Mx | 92.39 | 504.18 |
| 6 | T2N2Mx | 64.38 | 416.45 |
| 7 | T4N0Mx | 83.41 | 739.74 |
| 8 | T2N2Mx | 45.27 | 498.05 |
| 9 | T2N3Mx | 45.79 | 462.43 |
Figure 2Volumetric changes between iCT and rCT. Volume changes between the initial planning CT and the rCT at the 23rd fraction; a) contours on the initial planning CT; b) contours on the rCT; c) contours on the initial planning CT projected on the rCT after image fusion.
Volume changes between the initial planning CT (iCT) and the repeat CT (rCT) at 23rd fraction
| GTV | 78.83 ± 38.42 | 71.44 ± 37.46 | 0.42 |
| Brainstem | 28.77 ± 2.89 | 26.43 ± 3.03 | 0.22 |
| Cord | 25.52 ± 2.44 | 24.57 ± 2.30 | 0.44 |
| Mandible | 89.49 ± 14.33 | 89.23 ± 12.32 | 0.58 |
| Left parotid | 19.91 ± 4.89 | 11.80 ± 2.79 | <0.01 |
| Right Parotid | 21.58 ± 6.16 | 13.29 ± 4.17 | <0.01 |
Figure 3DVH comparisons between replanning based on the initial planning CT with propagated new contours from the rCT and replanning based on the rCT.
Detailed dosimetric comparison between replanning based on the initial CT with propagated contours from the repeat CT and replanning based on the repeat CT
| PTVGTV Dmax | 1343.5 ± 65.9 | 1335.2 ± 29.7 | 0.66 |
| PTVGTV Dmean | 1253.5 ± 40.3 | 1244.4 ± 13.7 | 0.42 |
| PTVGTV V93 | 96.9 ± 3.0 | 97.2 ± 2.6 | 0.79 |
| PTVGTV V95 | 91.8 ± 7.2 | 93.0 ± 5.4 | 0.66 |
| PTVCTV Dmax | 1349.4 ± 61.3 | 1338.2 ± 29.0 | 0.55 |
| PTVCTV Dmean | 1083.4 ± 31.5 | 1078.3 ± 23.7 | 0.67 |
| PTVCTV V93 | 98.9 ± 1.1 | 99.1 ± 0.9 | 0.12 |
| PTVCTV V95 | 98.4 ± 1.3 | 98.7 ± 1.2 | 0.24 |
| Brainstem Dmax | 859.7 ± 89.1 | 852.8 ± 89.7 | 0.5 |
| Brainstem Dmean | 457.2 ± 92.3 | 436.9 ± 110.5 | 0.15 |
| Cord Dmax | 692.0 ± 29.8 | 690.6 ± 31.3 | 0.9 |
| Cord Dmean | 413.3 ± 64.2 | 418.7 ± 63.5 | 0.17 |
| Mandible Dmax | 1146.9 ± 99.5 | 1131.6 ± 84.5 | 0.38 |
| Mandible Dmean | 691.7 ± 70.2 | 689.8 ± 60.5 | 0.24 |
| Lt parotid Dmax | 1134.0 ± 143.2 | 1139.1 ± 103.6 | 0.97 |
| Lt parotid Dmean | 519.4 ± 61.3 | 513.2 ± 57.2 | 0.17 |
| Lt parotid V5 | 42.1 ± 8.5 | 41.8 ± 7.8 | 0.6 |
| Rt parotid Dmax | 1144.0 ± 95.8 | 1162.1 ± 107.3 | 0.21 |
| Rt parotid Dmean | 514.9 ± 32.8 | 511.8 ± 40.8 | 0.63 |
| Rt parotid V5 | 39.5 ± 4.0 | 39.3 ± 4.8 | 0.82 |
Note: the unit for dose is cGy; V5 = the volume receiving 5 Gy.
Figure 4Typical dose distributions. A typical dose distribution of (a) the original VMAT plan delivered to the new contours projected in the initial planning CT, and (b) the re-optimized VMAT plan dose distribution based on the new contours projected in the iCT.
Dosimetric comparisons of without replanning in the initial CT, replanning in the initial CT
| PTVGTV Dmax | 1332.6 ± 42.4 | 1343.5 ± 65.9 | 0.77 |
| PTVGTV Dmean | 1257.2 ± 30.1 | 1253.5 ± 40.3 | 0.97 |
| PTVGTV V93 | 98.2 ± 1.7 | 96.9 ± 3.0 | 0.12 |
| PTVGTV V95 | 94.8 ± 4.7 | 91.8 ± 7.2 | 0.26 |
| PTVCTV Dmax | 1341.8 ± 40.7 | 1349.4 ± 61.3 | 0.85 |
| PTVCTV Dmean | 1087.2 ± 20.7 | 1083.4 ± 31.5 | 0.91 |
| PTVCTV V93 | 99.2 ± 0.7 | 98.9 ± 1.1 | 0.27 |
| PTVCTV V95 | 98.8 ± 0.9 | 98.4 ± 1.3 | 0.62 |
| Brainstem Dmax | 844.7 ± 95.4 | 859.7 ± 89.1 | 0.94 |
| Brainstem Dmean | 448.5 ± 76.4 | 457.2 ± 92.3 | 0.84 |
| Cord Dmax | 695.5 ± 44.2 | 692.0 ± 29.8 | 0.87 |
| Cord Dmean | 409.0 ± 54.7 | 413.3 ± 64.2 | 0.82 |
| Mandible Dmax | 1128.2 ± 95.0 | 1146.9 ± 99.5 | 0.76 |
| Mandible Dmean | 705.4 ± 70.9 | 691.7 ± 70.2 | 0.91 |
| Lt parotid Dmax | 1138.7 ± 114.6 | 1134.0 ± 143.2 | 0.99 |
| Lt parotid Dmean | 581.9 ± 87.1 | 519.4 ± 61.3 | 0.05 |
| Lt parotid V5 | 49.9 ± 12.3 | 42.1 ± 8.5 | 0.01 |
| Rt parotid Dmax | 1151.8 ± 108.1 | 1144.0 ± 95.8 | 0.75 |
| Rt parotid Dmean | 581.2 ± 79.8 | 514.9 ± 32.8 | 0.02 |
| Rt parotid V5 | 50.7 ± 8.3 | 39.5 ± 4.0 | 0.001 |