| Literature DB >> 24258033 |
F Cappellani1, G Melandrone, G Restelli.
Abstract
The use of a second harmonic diode laser system for monitoring gaseous ammonia in tropospheric air is discussed. Using a 96 m White cell operated at 13 torr pressure in flow mode, a minimum detectable concentration lower than 1 ppbv was obtained.The NH3 mixing ratio was determined by simultaneous spectroscopic analysis of an atmospheric CO2 line used as a built-in calibration standard.Entities:
Year: 1985 PMID: 24258033 DOI: 10.1007/BF00394068
Source DB: PubMed Journal: Environ Monit Assess ISSN: 0167-6369 Impact factor: 2.513