Melanie Fischbach, Roger A Hälg, Matthias Hartmann, Jürgen Besserer, Günther Gruber, Uwe Schneider1.
Abstract
BACKGROUND: For patients with high risk breast cancer and mastectomy, radiotherapy is the treatment of choice to improve survival and local control. Target dose is mainly limited due to skin reactions. The feasibility of using 4 MV beams for chest wall treatment was studied and compared to standard 6 MV bolus radiotherapy.Entities:
Mesh:
Year: 2013 PMID: 24238366 PMCID: PMC3842796 DOI: 10.1186/1748-717X-8-270
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Illustration of the Alderson phantom with the TLD measurement positions marked. Please note that the ‘+’-marks are not relevant for this experiment.
Figure 2Transversal slice at 0.1 cm cranial to the isocenter with the TLDs (in blue) in 3 mm depth. For the 4 MV treatment plan the 85%, 95%, 100% and 105% isodoses are marked in yellow, orange, green and red, respectively. The geometry of the two tangential fields is shown.
Results of the TLD measurements at the four positions in 3 mm depth (3.1 – 3.4) and 1 mm depth (O.1 – O.4) in percentage of 2 Gy (prescribed dose per fraction) for the 4 MV and the 6 MV treatment plan
| O.1 | 69.4 | 84.1 |
| O.2 | 70.5 | 76.7 |
| O.3 | 73.1 | 84.7 |
| O.4 | 78.7 | 81.0 |
| 3.1 | 95.1 | 95.7 |
| 3.2 | 94.6 | 95.2 |
| 3.3 | 93.1 | 94.8 |
| 3.4 | 94.9 | 93.0 |
The 6 MV treatment without/with bolus was weighted 40/60.
Figure 3Measured dose at the 4 locations in 3 mm depth as a percentage of the prescribed dose. The light grey bars indicate the results from the 4 MV measurement and the dark grey bars from the 6 MV measurements without/with bolus using a 40/60 weighting.
Figure 4Measured dose at the 4 locations in 1 mm depth as a percentage of the prescribed dose. The light grey bars indicate the results from the 4 MV measurement and the dark grey bars from the 6 MV measurements without/with bolus using a 40/60 weighting.
Results of the TLD measurements and TPS calculations at the four positions in 3 mm depth (3.1 – 3.4) and 1 mm depth (O.1 – O.4) in Gy for the 4 MV and the 6 MV (without flab) treatment plan
| O.1 | 1.39 ± 0.03 | 1.41 | 1.4 | 1.23 ± 0.01 | 1.23 | 0.0 |
| O.2 | 1.41 ± 0.03 | 1.38 | -2.1 | 1.16 ± 0.01 | 1.25 | 7.8 |
| O.3 | 1.46 ± 0.03 | 1.42 | -2.7 | 1.35 ± 0.01 | 1.35 | 0.0 |
| O.4 | 1.57 ± 0.02 | 1.55 | -1.3 | 1.30 ± 0.02 | 1.37 | 5.4 |
| 3.1 | 1.90 ± 0.04 | 1.87 | -1.6 | 1.73 ± 0.01 | 1.64 | -5.2 |
| 3.2 | 1.89 ± 0.03 | 1.88 | -0.5 | 1.78 ± 0.02 | 1.71 | -3.9 |
| 3.3 | 1.86 ± 0.05 | 1.88 | 1.1 | 1.81 ± 0.02 | 1.74 | -3.9 |
| 3.4 | 1.90 ± 0.06 | 1.95 | 2.6 | 1.74 ± 0.02 | 1.83 | 5.2 |
The 6 MV treatment without/with bolus was weighted 40/60. The last column indicates the difference between calculation and measurement.