Literature DB >> 24227878

Prospective evaluation of the VITEK MS for the routine identification of bacteria and yeast in the clinical microbiology laboratory: assessment of accuracy of identification and turnaround time.

Angella Charnot-Katsikas1, Vera Tesic1, Sue Boonlayangoor1, Cindy Bethel1, Karen M Frank1.   

Abstract

This study assessed the accuracy of bacterial and yeast identification using the VITEK MS, and the time to reporting of isolates before and after its implementation in routine clinical practice. Three hundred and sixty-two isolates of bacteria and yeast, consisting of a variety of clinical isolates and American Type Culture Collection strains, were tested. Results were compared with reference identifications from the VITEK 2 system and with 16S rRNA sequence analysis. The VITEK MS provided an acceptable identification to species level for 283 (78 %) isolates. Considering organisms for which genus-level identification is acceptable for routine clinical care, 315 isolates (87 %) had an acceptable identification. Six isolates (2 %) were identified incorrectly, five of which were Shigella species. Finally, the time for reporting the identifications was decreased significantly after implementation of the VITEK MS for a total mean reduction in time of 10.52 h (P<0.0001). Overall, accuracy of the VITEK MS was comparable or superior to that from the VITEK 2. The findings were also comparable to other studies examining the accuracy of the VITEK MS, although differences exist, depending on the diversity of species represented as well as on the versions of the databases used. The VITEK MS can be incorporated effectively into routine use in a clinical microbiology laboratory and future expansion of the database should provide improved accuracy for the identification of micro-organisms.

Entities:  

Mesh:

Year:  2013        PMID: 24227878     DOI: 10.1099/jmm.0.063636-0

Source DB:  PubMed          Journal:  J Med Microbiol        ISSN: 0022-2615            Impact factor:   2.472


  7 in total

1.  Evaluation of FilmArray and Verigene systems for rapid identification of positive blood cultures.

Authors:  M M Bhatti; S Boonlayangoor; K G Beavis; V Tesic
Journal:  J Clin Microbiol       Date:  2014-07-16       Impact factor: 5.948

2.  MALDI-TOF MS versus VITEK 2 ANC card for identification of anaerobic bacteria.

Authors:  Yang Li; Bing Gu; Genyan Liu; Wenying Xia; Kun Fan; Yaning Mei; Peijun Huang; Shiyang Pan
Journal:  J Thorac Dis       Date:  2014-05       Impact factor: 2.895

3.  Rapid identification of positive blood cultures by matrix-assisted laser desorption ionization-time of flight mass spectrometry using prewarmed agar plates.

Authors:  M M Bhatti; S Boonlayangoor; K G Beavis; V Tesic
Journal:  J Clin Microbiol       Date:  2014-09-17       Impact factor: 5.948

4.  Effect of Black Grape Seed Extract (Vitis vinifera) on Biofilm Formation of Methicillin-Resistant Staphylococcus aureus and Staphylococcus haemolyticus.

Authors:  Alia Hussein Al-Mousawi; Siham Jasim Al-Kaabi; Ahmad J H Albaghdadi; Abbas F Almulla; Ameer Raheem; Ali Abbas Abo Algon
Journal:  Curr Microbiol       Date:  2019-12-05       Impact factor: 2.188

5.  Alterations of lung microbiota in a mouse model of LPS-induced lung injury.

Authors:  Valeriy Poroyko; Fanyong Meng; Angelo Meliton; Taras Afonyushkin; Alexander Ulanov; Ekaterina Semenyuk; Omar Latif; Vera Tesic; Anna A Birukova; Konstantin G Birukov
Journal:  Am J Physiol Lung Cell Mol Physiol       Date:  2015-05-08       Impact factor: 5.464

6.  Efficacy of an Enrichment Media for Increasing Threshold for Carbapenem-Resistant Enterobacteriaceae Screening.

Authors:  Avi Peretz; Nina Pastukh; Natlya Isakovich; Anna Koifman; Diana Brodsky; Hila Mizrahi; Ilana Aharon; Kozitta Labay
Journal:  J Clin Lab Anal       Date:  2015-12-14       Impact factor: 2.352

7.  Matrix-assisted laser desorption ionization-time of flight mass spectrometry for identification of bacteria isolated from pharmaceutical clean rooms.

Authors:  Laíse de Oliveira Andrade; Rajendra Awasthi; Kamal Dua; Terezinha de Jesus Andreoli Pinto
Journal:  Interv Med Appl Sci       Date:  2018-03
  7 in total