Ion trajectories in an electrostatic ion guide for external ion source fourier transform ion cyclotron resonance mass spectrometry.

J A Marto1, A G Marshall, M A May, P A Limbach.   

Abstract

An electrostatic ion guide (EIG) that consists of concentric cylinder and central wire electrodes can transport ions efficiently from an external ion source to an ion cyclotron resonance (ICR) ion trap for mass analysis, with several advantages over current injection methods. Because the electrostatic force of the EIG captures ions in a stable orbit about the wire electrode, ions with initially divergent trajectories may be redirected toward the ICR ion trap for improved ion transmission efficiency. SIMION trajectory calculations (ion kinetic energy, 1-200 eV; elevation angle, 0.30 °; azimuthal angle, 0.360°) predict that ions of m/z 1000 may be transmitted through a strong (0.01 → 3.0-T) magnetic field gradient. Judicious choice of ion source position and EIG potential minimizes the spread in ion axial kinetic energy at the ICR ion trap. Advantages of the EIG include large acceptance angle, even for ions that have large initial kinetic energy and large radial displacement with respect to the central z-axis, low ion extraction voltage (5-20 V), and efficient trapping because ions need not be accelerated to high velocity to pass through the magnetic field gradient.

Year:  1995        PMID: 24214038     DOI: 10.1016/1044-0305(95)94194-X

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  6 in total

1.  A pulsed ion deflection system for background reduction in (252)CF-plasma desorption mass spectrometry.

Authors:  B Wolf; R D Macfarlane
Journal:  J Am Soc Mass Spectrom       Date:  1992-10       Impact factor: 3.109

2.  Excitation modes for fourier transform-ion cyclotron resonance mass spectrometry.

Authors:  L Schweikhard; A G Marshall
Journal:  J Am Soc Mass Spectrom       Date:  1993-06       Impact factor: 3.109

3.  A low voltage ion transport system for external ionization fourier transform mass spectrometry.

Authors:  C F Ijames; S P Markey
Journal:  J Am Soc Mass Spectrom       Date:  1994-05       Impact factor: 3.109

4.  Fourier-transform electrospray instrumentation for tandem high-resolution mass spectrometry of large molecules.

Authors:  S C Beu; M W Senko; J P Quinn; F M Wampler; F W McLafferty
Journal:  J Am Soc Mass Spectrom       Date:  1993-07       Impact factor: 3.109

Review 5.  Fourier transform mass spectrometry of high-mass biomolecules.

Authors:  M V Buchanan; R L Hettich
Journal:  Anal Chem       Date:  1993-03-01       Impact factor: 6.986

6.  Wide-mass-range axialization for high-resolution Fourier-transform ion cyclotron resonance mass spectrometry of externally generated ions.

Authors:  J A Marto; S Guan; A G Marshall
Journal:  Rapid Commun Mass Spectrom       Date:  1994-08       Impact factor: 2.419

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.