Erin E E Fowler, Beibei Lu, John J Heine1.
Abstract
BACKGROUND: Breast density is a significant breast cancer risk factor measured from mammograms. The most appropriate method for measuring breast density for risk applications is still under investigation. Calibration standardizes mammograms to account for acquisition technique differences prior to making breast density measurements. We evaluated whether a calibration methodology developed for an indirect x-ray conversion full field digital mammography (FFDM) technology applies to direct x-ray conversion FFDM systems.Entities:
Mesh:
Year: 2013 PMID: 24207013 PMCID: PMC3829208 DOI: 10.1186/1475-925X-12-114
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Figure 1Breast tissue equivalent phantom positioning, detector field of view, and regions of interest. This shows the large field of view (largest rectangle) for the Hologic Selenia unit, the phantom (gray rectangle) placement on the detector (18 cm × 24 cm or approximately 2500 × 3400 pixels), and the regions of interest (ROIs) used for the analysis outlined with narrow light borders. The size of the large ROI is 2000 × 2500 pixels, and the size of the small ROI is 1000 × 1250 pixels.
Exposure and pixel value response analysis by target/filter and select kV combination
| W/Rh | | | | | | | | | |
| kV = | 26 | 26 - 150 | 63.86 (0.15) | 0.65 | 0.99 | 26 - 150 | 66.26 (0.15) | 38.69 | 0.99 |
| | 28 | 10 - 120 | 82.37 (0.38) | 3.06 | 0.99 | 10 - 120 | 85.24 (0.09) | 43.44 | 0.99 |
| | 30 | 4 - 85 | 101.37 (0.66) | 16.52 | 0.99 | 4 - 85 | 105.26 (0.17) | 48.54 | 0.99 |
| W/Ag | | | | | | | | | |
| kV = | 28 | 4 - 85 | 113.52 (0.57) | 19.31 | 0.99 | 4 - 85 | 108.44 (0.21) | 52.47 | 0.99 |
| | 30 | 4 - 75 | 146.00 (1.04) | 2.51 | 0.99 | 4 - 75 | 137.99 (0.51) | 54.89 | 0.99 |
| Mo/Mo | | | | | | | | | |
| kV = | 26 | 6 - 80 | 103.86 (0.30) | 72.34 | 0.99 | 10 - 80 | 183.37 (0.89) | 56.65 | 0.99 |
| | 27 | 6 - 80 | 122.71 (0.36) | 72.29 | 0.99 | 10 - 65 | 219.59 (1.09) | 29.72 | 0.99 |
| Mo/Rh | | | | | | | | | |
| kV = | 28 | 6 - 65 | 123.56 (0.60) | 76.19 | 0.99 | 7 - 56 | 248.42 (0.78) | 12.03 | 0.99 |
| 29 | 6 - 65 | 142.01 (0.73) | 69.55 | 0.99 | 10 - 45 | 288.13 (1.32) | 3.09 | 0.99 | |
H1: data acquired on 07/27/2012.
H2: data acquired on 07/20/2012.
H3: data acquired on 04/08/2013.
GE: data acquired on 10/13/2006.
GE: quantities referenced with the ′0′ subscript.
The detector exposure and pixel value response is modeled as a linear function of mAs for select kV settings: (1) the W/Rh and W/Ag target/filter combinations for the similar Hologic Selenia systems H1 and H2 (top); and (2) the Mo/Mo and Mo/Rh combinations for the third Hologic Selenia H3 system and General Electric Senographe 2000D (GE) system (bottom), designated by the subscript 0. The slopes (mi) and standard error (SEi), estimated as unit increase in pixel value per mAs, intercept (bi) measured in pixel value, and coefficient of determination (Ri2) were derived from linear regression analysis. The mAs columns give the range of settings used for the acquisition techniques. The upper mAs value for a given technique was set close to where detector saturation occurs.
Figure 2Exposure response for the Hand Hunits. This shows exposure measured in milliampere seconds (mAs), the pixel value response, and fitted curves (solid) of the detector without attenuation for the H1 (left) and H2 (right) units. The measured values are averages represented by squares. The top plots show the W/Rh (target/filter) combinations acquired with 28 kV and the bottom plots show the W/Ag combinations acquired with 30 kV.
Figure 3Exposure response for the Hand GE units. This shows the exposure measured in milliampere seconds (mAs), the pixel value response, and fitted curves (solid) of the detector for the H3 (left) and GE (right) units. The measured values are averages represented by squares. The Mo/Mo (target/filter) combinations were acquired with 27 kV (top) and the Mo/Rh combinations (bottom) were acquired with 28 kV.
Baseline (BL) calibration dataset summary for the Hologic Selenia (Hand H) units
| | | | | | | | ||
| W/Rh | | | | | | | | |
| kV = | 26 | 2 - 6 | 0.464 (0.008) | 4.03 | 0.99 | 0.624 (0.022) | 3.80 | 0.99 |
| | 27 | 2 - 6 | 0.461 (0.007) | 4.17 | 0.99 | 0.620 (0.020) | 3.93 | 0.99 |
| | 28 | 2 - 7 | 0.448 (0.008) | 4.26 | 0.99 | 0.595 (0.020) | 4.00 | 0.99 |
| | 29 | 3 - 7 | 0.432 (0.009) | 4.31 | 0.99 | 0.560 (0.018) | 3.94 | 0.99 |
| | 30 | 3 - 7 | 0.432 (0.007) | 4.44 | 0.99 | 0.560 (0.016) | 4.08 | 0.99 |
| W/Ag | | | | | | |||
| kV = | 27 | 3 - 7 | 0.412 (0.006) | 4.40 | 0.99 | 0.539 (0.013) | 4.09 | 0.99 |
| | 28 | 3 - 7 | 0.405 (0.006) | 4.55 | 0.99 | 0.539 (0.013) | 4.29 | 0.99 |
| | 29 | 3 - 7 | 0.403 (0.006) | 4.71 | 0.99 | 0.538 (0.010) | 4.45 | 0.99 |
| | 30 | 3 - 7 | 0.400 (0.005) | 4.83 | 0.99 | 0.537 (0.010) | 4.61 | 0.99 |
| | 31a | 3 - 7 | 0.406 (0.006) | 5.04 | 0.99 | 0.533 (0.011) | 4.76 | 0.99 |
| | 32b | 4 - 7 | 0.392 (0.004) | 5.08 | 0.99 | 0.512 (0.011) | 4.79 | 0.99 |
| | | | | | | | ||
| W/Rh | | | | | | |||
| kV = | 26 | 2 - 6 | 0.464 (0.008) | 4.06 | 0.99 | 0.616 (0.026) | 3.82 | 0.99 |
| | 27 | 2 - 6 | 0.462 (0.008) | 4.21 | 0.99 | 0.621 (0.019) | 3.97 | 0.99 |
| | 28 | 2 - 7 | 0.442 (0.008) | 4.27 | 0.99 | 0.581 (0.023) | 4.00 | 0.99 |
| | 29 | 3 - 7 | 0.431 (0.006) | 4.35 | 0.99 | 0.536 (0.026) | 3.90 | 0.99 |
| | 30 | 3 - 7 | 0.422 (0.007) | 4.42 | 0.99 | 0.540 (0.020) | 4.06 | 0.99 |
| W/Ag | | | | | | | ||
| kV = | 27 | 3 - 7 | 0.404 (0.005) | 4.34 | 0.99 | 0.518 (0.019) | 4.01 | 0.99 |
| | 28 | 3 - 7 | 0.401 (0.004) | 4.49 | 0.99 | 0.522 (0.015) | 4.23 | 0.99 |
| | 29 | 3 - 7 | 0.402 (0.004) | 4.66 | 0.99 | 0.520 (0.014) | 4.37 | 0.99 |
| | 30 | 3 - 7 | 0.395 (0.006) | 4.77 | 0.99 | 0.534 (0.009) | 4.58 | 0.99 |
| | 31 | 3 - 7 | 0.395 (0.005) | 4.91 | 0.99 | 0.524 (0.009) | 4.67 | 0.99 |
| 32b | 4 - 7 | 0.388 (0.005) | 5.01 | 0.99 | 0.506 (0.011) | 4.72 | 0.99 | |
H1: W/Rh data acquired on 07/06/2012; W/Ag 27-30 kV data acquired on 07/06/2012.
H2: W/Rh data acquired on 07/10/2012; W/Ag 27-31 kV data acquired on 07/10/2012.
a: data acquired on 07/09/2012.
b: data acquired on 07/17/2012.
The effective x-ray attenuation coefficients, μa and μg, logarithmic intercepts, LIa and LIg, and coefficient of determination (R2) were derived with regression analysis for the similar Hologic (H1 and H2 ) units for the W/Rh and W/Ag target/filter combinations are provided as mean values. The mean μl and the associated mean standard error (SEl) are cited. The x-ray tube voltage (kV) and phantom height ranges are provided. For each kV setting, the images were acquired by incrementing the phantom heights from the lower range to the upper range in 1 cm increments. The absolute value of the slope is equivalent to the effective x-ray attenuation coefficient for each acquisition technique and BTE material.
Figure 4Calibration curves for Hand Hunits. This shows the fitted adipose (solid) and glandular (dotted) calibration curves for the H1 (left-side plots) and H2 (right-side plots) units. The measured logarithmic relative response (LRE) quantities are averages represented by squares. The top plots show the W/Rh target/filter combinations acquired with 28 kV and the bottom plots show the W/Ag combinations acquired with 30 kV.
Baseline (BL) calibration data summary for two different FFDM technologies
| | | | | | | | ||
| Mo/Mo | | | | | | | | |
| kV = | 25 | 2 - 6 | 0.561 (0.014) | 4.18 | 0.99 | 0.705 (0.047) | 3.69 | 0.99 |
| | 26 | 2 - 6 | 0.557 (0.011) | 4.40 | 0.99 | 0.706 (0.043) | 3.92 | 0.99 |
| | 27 | 2 - 6 | 0.551 (0.012) | 4.57 | 0.99 | 0.709 (0.036) | 4.18 | 0.99 |
| | 28 | 2 - 7 | 0.526 (0.012) | 4.70 | 0.99 | 0.639 (0.039) | 4.15 | 0.99 |
| | 29 | 3 - 7 | 0.495 (0.010) | 4.74 | 0.99 | 0.586 (0.029) | 4.09 | 0.99 |
| | 30 | 3 - 7 | 0.486 (0.009) | 4.89 | 0.99 | 0.568 (0.030) | 4.22 | 0.99 |
| | 31 | 3 - 7 | 0.470 (0.011) | 4.99 | 0.99 | 0.560 (0.026) | 4.42 | 0.99 |
| Mo/Rh | | | | | | | | |
| kV = | 27 | 3 - 7 | 0.476 (0.009) | 4.35 | 0.99 | 0.586 (0.031) | 3.80 | 0.99 |
| | 28 | 3 - 7 | 0.482 (0.008) | 4.55 | 0.99 | 0.592 (0.026) | 4.01 | 0.99 |
| | 29 | 3 - 7 | 0.475 (0.008) | 4.67 | 0.99 | 0.595 (0.024) | 4.20 | 0.99 |
| | 30 | 3 - 7 | 0.471 (0.008) | 4.83 | 0.99 | 0.585 (0.025) | 4.32 | 0.99 |
| | 31 | 3 - 7 | 0.463 (0.008) | 4.93 | 0.99 | 0.582 (0.021) | 4.49 | 0.99 |
| | 32 | 3 - 7 | 0.459 (0.009) | 5.07 | 0.99 | 0.564 (0.022) | 4.57 | 0.99 |
| | 33 | 3 - 7 | 0.445 (0.008) | 5.14 | 0.99 | 0.554 (0.021) | 4.71 | 0.99 |
| | 34 | 3 - 7 | 0.443 (0.008) | 5.28 | 0.99 | 0.533 (0.020) | 4.77 | 0.99 |
| | | | | | | | ||
| Mo/Mo | | | | | | | | |
| kV = | 25* | 2 - 6 | 0.584 (0.004) | 4.93 | 0.99 | 0.861 (0.014) | 4.74 | 0.99 |
| | 26 | 2 - 6 | 0.572 (0.005) | 5.09 | 0.99 | 0.833 (0.016) | 4.89 | 0.99 |
| | 27 | 2 - 6 | 0.560 (0.005) | 5.26 | 0.99 | 0.805 (0.017) | 5.04 | 0.99 |
| | 28* | 2 - 7 | 0.548 (0.005) | 5.43 | 0.99 | 0.777 (0.019) | 5.19 | 0.99 |
| | 29* | 3 - 7 | 0.536 (0.006) | 5.60 | 0.99 | 0.749 (0.021) | 5.34 | 0.99 |
| | 30* | 3 - 7 | 0.524 (0.006) | 5.77 | 0.99 | 0.721 (0.023) | 5.49 | 0.99 |
| Mo/Rh | | | | | | | | |
| kV = | 27* | 3 - 7 | *0.512 (0.005) | 5.21 | 0.99 | 0.736 (0.011) | 5.03 | 0.99 |
| | 28 | 3 - 7 | 0.504 (0.005) | 5.34 | 0.99 | 0.718 (0.013) | 5.15 | 0.99 |
| | 29 | 3 - 7 | 0.495 (0.004) | 5.47 | 0.99 | 0.700 (0.014) | 5.27 | 0.99 |
| | 30* | 3 - 7 | *0.487 (0.004) | 5.60 | 0.99 | 0.682 (0.016) | 5.39 | 0.99 |
| 31* | 3 - 7 | *0.479 (0.004) | 5.74 | 0.99 | 0.665 (0.017) | 5.51 | 0.99 | |
H3: data acquired on 04/08/2013.
GE: data acquired on 10/13/2006.
* Interpolated values.
The effective x-ray attenuation coefficients, μa and μg, logarithmic intercepts LIa and LIg, and coefficient of determination (R2), derived with regression analysis, for the Hologic (H3) and the General Electric Senographe 2000D (GE) FFDM systems for the Mo/Mo and Mo/Rh target/filter combinations are provided as mean values. The mean μl and the associated mean standard error (SEl) are cited. The x-ray tube voltage (kV) and phantom height (thickness) ranges are provided. For each kV sample, images were acquired by incrementing the phantom heights from the lower range limit to the upper range limit in 1 cm increments. The absolute value of the slope, as cited, is equivalent to the effective attenuation coefficient for a given BTE material.
Figure 5Calibration curves for Hand GE units. This shows the fitted adipose (solid) and glandular (dotted) calibration curves for the H3 (left) and GE (right) units. The measured logarithmic relative response (LRE) quantities are averages represented by squares. The top plots show the Mo/Mo target/filter combinations acquired with 28 kV and the bottom plots show the Mo/Rh combinations acquired with 28 kV.
Calibration accuracy for the Hand Hunits
| W/Rh | | | | | | | | | | | | | |
| kV = | 26 | 50.6(0.44) | 51.9(0.47) | 54.0(0.47) | 48.8(0.48) | 51.6(0.44) | 52.9(0.40) | 55.7(0.44) | 57.1(0.47) | 59.2(0.47) | 43.6(0.48) | 46.5(0.44) | 47.8(0.40) |
| | 27 | 50.2(0.44) | 51.8(0.46) | 53.5(0.44) | 49.2(0.43) | 50.6(0.43) | 52.3(0.42) | 54.2(0.44) | 55.9(0.46) | 57.6(0.44) | 45.2(0.43) | 46.5(0.43) | 48.2(0.42) |
| | 28 | 50.8(0.46) | 52.3(0.46) | 53.5(0.46) | 50.8(0.37) | 51.8(0.33) | 53.2(0.35) | 55.5(0.47) | 57.0(0.47) | 58.2(0.46) | 46.2(0.36) | 47.2(0.33) | 48.5(0.35) |
| | 29 | 50.3(0.47) | 51.4(0.50) | 52.7(0.48) | 51.2(0.31) | 52.6(0.30) | 53.3(0.31) | 55.9(0.47) | 57.0(0.50) | 58.3(0.49) | 45.7(0.31) | 47.1(0.29) | 47.8(0.31) |
| | 30 | 50.9(0.61) | 52.1(0.63) | 53.0(0.62) | 51.5(0.39) | 52.5(0.40) | 53.1(0.42) | 55.8(0.62) | 57.0(0.65) | 57.9(0.63) | 46.7(0.38) | 47.8(0.39) | 48.3(0.41) |
| W/Ag | | | | | | | | | | | | | |
| kV = | 27 | 51.0(0.77) | 52.4(0.78) | 52.9(0.80) | 52.3(0.43) | 53.3(0.44) | 54.0(0.44) | 50.9(0.80) | 52.4(0.81) | 52.9(0.83) | 52.3(0.42) | 53.3(0.42) | 53.9(0.43) |
| | 28 | 51.1(0.90) | 52.1(0.92) | 52.8(0.93) | 52.2(0.39) | 53.5(0.41) | 52.5(0.43) | 50.1(0.92) | 51.1(0.94) | 51.8(0.95) | 53.1(0.38) | 54.4(0.39) | 53.5(0.42) |
| | 29 | 50.8(0.94) | 52.1(0.96) | 52.2(0.97) | 51.5(0.40) | 52.8(0.41) | 52.9(0.40) | 49.5(0.99) | 50.7(0.98) | 50.9(0.99) | 52.8(0.39) | 54.1(0.40) | 54.2(0.39) |
| | 30 | 51.5(0.98) | 52.5(0.99) | 52.6(1.00) | 52.5(0.44) | 53.0(0.44) | 53.4(0.45) | 48.9(0.99) | 50.0(1.00) | 50.0(1.01) | 55.0(0.43) | 55.6(0.44) | 55.9(0.45) |
| | 31 | 55.9(0.91) | 56.7(0.90) | 56.8(0.93) | 52.3(0.51) | 53.2(0.51) | 53.6(0.52) | 48.3(0.94) | 49.1(0.93) | 49.3(0.97) | 59.7(0.49) | 60.6(0.49) | 61.0(0.50) |
| 32 | 54.0(0.97) | 54.8(0.97) | 55.9(1.00) | 54.0(0.64) | 54.6(0.64) | 55.8(0.65) | 51.2(0.98) | 52.1(0.97) | 53.2(1.00) | 56.7(0.63) | 57.3(0.64) | 58.5(0.65) | |
Within-unit Calibration: 50/50 mixtures acquired and calibrated with the same FFDM unit.
Cross-unit Calibration: 50/50 mixtures acquired with each FFDM unit were calibrated with the other unit's calibration data.
H1: phantom data acquired on 07/27/2012 and calibration data on 07/06/2012.
H2: phantom data acquired on 07/20/2012 and calibration data on 07/10/2012.
This table summarizes the results from the within-unit calibration (left side) and cross-unit calibration (right side) for H1 and H2 units. For each target/filter combination and kV setting, the 50/50 mixture phantom images were acquired with three mAs settings including the reference denoted by mAsr. The mean and the standard deviation (SD) of the respective distributions are provided in PG units.
Within-unit calibration accuracy for the H
| Mo/Mo | | | | |
| kV = | 25 | 49.4(0.53) | 54.5(0.49) | 56.1(0.48) |
| | 26 | 50.3(0.47) | 53.7(0.40) | 55.3(0.41) |
| | 27 | 51.6(0.36) | 54.6(0.34) | 55.1(0.35) |
| | 28 | 52.2(0.24) | 54.6(0.23) | 55.0(0.25) |
| | 29 | 53.4(0.16) | 55.4(0.17) | 55.6(0.19) |
| | 30 | 52.8(0.16) | 55.1(0.20) | 55.3(0.21) |
| | 31 | 53.3(0.27) | 55.7(0.32) | 55.7(0.32) |
| Mo/Rh | | | | |
| kV = | 27 | 51.8(1.44) | 53.6(1.52) | 54.2(1.51) |
| | 28 | 52.6(1.45) | 53.8(1.52) | 54.0(1.53) |
| | 29 | 52.8(1.44) | 54.0(1.52) | 54.4(1.51) |
| | 30 | 53.7(1.35) | 54.5(1.42) | 54.8(1.42) |
| | 31 | 53.0(1.30) | 54.6(1.37) | 54.7(1.35) |
| | 32 | 53.3(1.16) | 54.6(1.24) | 54.2(1.19) |
| | 33 | 54.3(1.09) | 55.0(1.15) | 55.0(1.14) |
| 34 | 55.6(0.94) | 55.6(1.00) | 55.8(0.98) | |
Within-unit Calibration: 50/50 mixtures acquired and calibrated with the same FFDM unit.
For each target/filter and kV setting, the 50/50 mixture phantom images were acquired with three mAs settings, including the reference denoted by mAsr. The mean and the standard deviation (SD) of the respective distributions are provided in PG units.
Calibration accuracy for the Hand Hunits using the regression parameters
| W/Rh | | | | | | | |
| kV = | 26 | 55.2 (0.46) | 56.6 (0.49) | 58.8 (0.48) | 53.4 (0.50) | 56.3 (0.46) | 57.7 (0.42) |
| | 27 | 54.0 (0.45) | 55.7 (0.47) | 57.4 (0.45) | 53.2 (0.44) | 54.6 (0.44) | 56.3 (0.43) |
| | 28 | 56.1 (0.48) | 57.7 (0.48) | 58.9 (0.48) | 56.8 (0.39) | 57.9 (0.35) | 59.4 (0.37) |
| | 29 | 51.8 (0.47) | 53.0 (0.50) | 54.2 (0.48) | 52.7 (0.32) | 54.1 (0.30) | 54.8 (0.32) |
| | 30 | 52.4 (0.61) | 53.6 (0.64) | 54.5 (0.62) | 52.8 (0.39) | 53.9 (0.40) | 54.5 (0.43) |
| W/Ag | | | | | | | |
| kV = | 27 | 52.0 (0.78) | 53.4 (0.79) | 53.9 (0.81) | 53.2 (0.44) | 54.2 (0.44) | 55.0 (0.44) |
| | 28 | 52.3 (0.89) | 53.3 (0.91) | 54.0 (0.93) | 53.0 (0.40) | 54.4 (0.41) | 53.4 (0.43) |
| | 29 | 52.0 (0.94) | 53.2 (0.96) | 53.3 (0.97) | 52.3 (0.40) | 53.6 (0.41) | 53.8 (0.41) |
| | 30 | 52.2 (0.98) | 53.2 (0.99) | 53.3 (1.00) | 53.4 (0.44) | 54.0 (0.45) | 54.4 (0.45) |
| | 31 | 57.1 (0.91) | 57.9 (0.90) | 58.0 (0.93) | 53.2 (0.51) | 54.1 (0.51) | 54.5 (0.52) |
| 32 | 54.8 (0.97) | 55.6 (0.97) | 56.8 (1.00) | 54.9 (0.64) | 55.5 (0.65) | 56.7 (0.66) | |
Within-unit Calibration: 50/50 mixtures acquired and calibrated with a same FFDM unit.
H1: Phantom data acquired on 07/27/2012 and calibration data on 07/06/2012.
H2: Phantom data acquired on 07/20/2012 and calibration data on 07/10/2012.
This table gives the calibration findings for the two similar H1 and H2 FFDM systems determined with the regression parameters. The mean and the standard deviation (SD) of the respective distributions are provided in PG units. For each target/filter and kV setting, the 50/50 mixture phantom images were acquired with three mAs settings including the reference denoted as mAsr.
Calibration accuracy using the cubic-spline height interpolation
| | | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| W/Rh | | | | | | Mo/Mo | | | | ||
| kV = | 25 | -- | -- | -- | -- | -- | kV = | 25 | 4.2 | 54.5 (0.49) | 54.9 (0.84) |
| | 26 | 4.2 | 46.5 (0.36) | 48.6 (0.67) | 51.1 (0.41) | 52.8 (0.76) | | 26 | 4.4 | 53.7 (0.40) | 54.5 (0.47) |
| | 27 | 4.4 | 48.6 (0.67) | 48.8 (0.42) | 51.4 (0.43) | 50.9 (0.82) | | 27 | 4.2 | 54.6 (0.34) | 54.7 (0.70) |
| | 28 | 4.2 | 48.4 (0.39) | 49.6 (0.65) | 51.8 (0.68) | 53.2 (0.91) | | 28 | 4.4 | 54.6 (0.23) | 54.7 (0.32) |
| | 29 | 4.4 | 47.8 (0.49) | 48.3 (0.53) | 51.3 (0.56) | 52.9 (0.63) | | 29 | 4.2 | 55.4 (0.17) | 55.2 (0.56) |
| | 30 | 4.2 | 48.2 (0.70) | 48.5 (0.80) | 51.9 (0.51) | 51.7 (0.88) | | 30 | 4.4 | 55.1 (0.20) | 54.8 (0.31) |
| | 31 | -- | -- | -- | -- | -- | | 31 | 4.2 | 55.7 (0.32) | 55.9 (0.52) |
| W/Ag | | | | | | Mo/Rh | | | | ||
| kV = | 27 | 4.2 | 46.5 (0.73) | 47.8 (0.91) | 52.0 (0.54) | 53.5 (0.89) | kV = | 27 | 4.2 | 53.6 (1.52) | 53.9 (1.68) |
| | 28 | 4.4 | 47.3 (0.85) | 48.0 (0.96) | 52.1 (0.52) | 51.8 (0.65) | | 28 | 4.4 | 53.8 (1.52) | 54.2 (1.38) |
| | 29 | 4.2 | 47.5 (0.90) | 48.4 (1.10) | 50.9 (0.49) | 52.4 (0.73) | | 29 | 4.2 | 54.0 (1.52) | 54.2 (1.68) |
| | 30 | 4.4 | 47.8 (0.95) | 48.0 (0.99) | 51.6 (0.52) | 51.5 (0.54) | | 30 | 4.4 | 54.5 (1.42) | 54.3 (1.23) |
| | 31 | 4.2 | 52.4 (0.86) | 52.8 (1.13) | 52.0 (0.50) | 52.0 (0.56) | | 31 | 4.2 | 54.6 (1.37) | 54.7 (1.52) |
| | 32 | 6.4 | 51.9 (0.75) | 52.4 (0.78) | 53.6 (0.57) | 54.5 (0.61) | | 32 | 4.4 | 54.6 (1.24) | 54.1 (0.98) |
| | 33 | -- | -- | -- | -- | -- | | 33 | 4.2 | 55.0 (1.15) | 55.0 (1.29) |
| 34 | -- | -- | -- | -- | -- | 34 | 4.4 | 55.6 (1.00) | 52.9 (0.76) | ||
Within-unit Calibration: 50/50 mixtures acquired and calibrated with a same FFDM unit.
H1: data acquired on 08/10/2012.
H2: data acquired on 08/06/2012.
H3: data acquired on 04/08/2013.
32 kV W/Ag BL acquired at 6.0 cm.
50/50 mixture phantom images were acquired with the 160 mAs reference setting for heights (right most column) not considered in the baseline sampling (Non-BL) and calibrated to evaluate the cubic-spline height interpolation. The findings are provided as the mean and the standard deviation (parenthetically) for the respective distributions in PG units. The calibration findings are also provided for 50/50 mixture phantom images acquired with the standard BL heights close to (4 cm or 6 cm) the non-BL heights, which were acquired on the same day for comparison purposes.
Calibration Accuracy using the regression parameters
| | | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| W/Rh | | | | | | Mo/Mo | | | | ||
| kV = | 25 | -- | -- | -- | -- | -- | kV = | 25 | 4.2 | 62.8 (0.53) | 63.3 (0.92) |
| | 26 | 4.2 | 50.9 (0.37) | 53.1 (0.70) | 55.8 (0.43) | 57.8 (0.80) | | 26 | 4.4 | 60.8 (0.43) | 61.5 (0.51) |
| | 27 | 4.4 | 51.6 (0.36) | 52.7 (0.44) | 55.4 (0.43) | 54.8 (0.84) | | 27 | 4.2 | 61.2 (0.36) | 61.2 (0.74) |
| | 28 | 4.2 | 53.6 (0.40) | 55.2 (0.68) | 57.9 (0.72) | 59.8 (0.97) | | 28 | 4.4 | 63.5 (0.25) | 64.5 (0.36) |
| | 29 | 4.4 | 49.3 (0.49) | 51.4 (0.54) | 52.7 (0.57) | 55.7 (0.65) | | 29 | 4.2 | 58.0 (0.18) | 59.2 (0.58) |
| | 30 | 4.2 | 49.6 (0.71) | 50.8 (0.81) | 53.2 (0.52) | 54.0 (0.90) | | 30 | 4.4 | 57.7 (0.20) | 59.7 (0.33) |
| | 31 | -- | -- | -- | -- | -- | | 31 | 4.2 | 58.2 (0.33) | 59.7 (0.53) |
| W/Ag | | | | | | Mo/Rh | | | | ||
| kV = | 27 | 4.2 | 47.5 (0.73) | 49.5 (0.93) | 52.9 (0.54) | 55.3 (0.90) | kV = | 27 | 4.2 | 55.8 (1.55) | 57.5 (1.74) |
| | 28 | 4.4 | 48.5 (0.84) | 50.4 (0.97) | 53.0 (0.52) | 54.3 (0.66) | | 28 | 4.4 | 55.8 (1.54) | 58.1 (1.43) |
| | 29 | 4.2 | 48.6 (0.90) | 50.2 (1.11) | 51.7 (0.49) | 54.0 (0.74) | | 29 | 4.2 | 55.7 (1.55) | 57.0 (1.73) |
| | 30 | 4.4 | 48.5 (0.94) | 50.0 (1.00) | 52.6 (0.52) | 53.6 (0.55) | | 30 | 4.4 | 56.3 (1.45) | 58.1 (1.28) |
| | 31 | 4.2 | 53.6 (0.87) | 54.8 (1.14) | 52.9 (0.50) | 53.5 (0.56) | | 31 | 4.2 | 56.9 (1.39) | 58.0 (1.56) |
| | 32 | 6.4 | 52.7 (0.76) | 52.5 (0.78) | 54.5 (0.57) | 54.6 (0.62) | | 32 | 4.4 | 57.1 (1.26) | 58.4 (1.01) |
| | 33 | -- | -- | -- | -- | -- | | 33 | 4.2 | 57.4 (1.18) | 58.5 (1.33) |
| 34 | -- | -- | -- | -- | -- | 34 | 4.4 | 57.6 (1.02) | 56.9 (0.79) | ||
Within-unit Calibration: 50/50 mixtures acquired and calibrated with a same FFDM unit.
H1: data acquired on 08/10/2012.
H2: data acquired on 08/06/2012.
H3: data acquired on 04/08/2013.
32 kV W/Ag BL acquired at 6.0 cm.
50/50 mixture phantom images were acquired with the 160 mAs reference setting for heights (right most column) not considered in the BL sampling (Non-BL) and calibrated to assess the regression parameter height interpolation. The findings are provided as the mean and the standard deviation (parenthetically) for the respective distributions in PG units. The calibration findings are also provided for 50/50 images acquired with the standard BL heights close to (4 cm or 6 cm) the non-BL heights, which were acquired on the same day for comparison proposes.
Calibration accuracy for images acquired with the small FOV calibrated with data acquired with the large FOV
| W/Rh | | | | | Mo/Mo | | | ||
| kV = | 25 | -- | -- | -- | -- | kV = | 25 | 53.0 (0.37) | 54.5 (0.49) |
| | 26 | 54.5 (0.38) | 51.9 (0.47) | 51.8 (0.28) | 51.6 (0.44) | | 26 | 52.3 (0.30) | 53.7 (0.40) |
| | 27 | 53.0 (0.38) | 51.8 (0.46) | 48.2 (0.34) | 50.6 (0.43) | | 27 | 53.1 (0.24) | 54.6 (0.34) |
| | 28 | 53.2 (0.44) | 52.3 (0.46) | 51.5 (0.42) | 50.6 (0.43) | | 28 | 53.3 (0.17) | 54.6 (0.23) |
| | 29 | 52.5 (0.50) | 51.4 (0.50) | 51.5 (0.55) | 51.8 (0.33) | | 29 | 54.2 (0.14) | 55.4 (0.17) |
| | 30 | 53.0 (0.64) | 52.1 (0.63) | 51.6 (0.72) | 52.6 (0.30) | | 30 | 54.1 (0.17) | 55.1 (0.20) |
| | 31 | -- | -- | -- | -- | | 31 | 54.7 (0.25) | 55.7 (0.32) |
| W/Ag | | | | | Mo/Rh | | | ||
| kV = | 27 | 52.2 (0.39) | 52.4 (0.78) | 52.7 (0.70) | 53.3 (0.44) | kV = | 27 | 51.2 (1.15) | 53.6 (1.52) |
| | 28 | 52.1 (0.42) | 52.1 (0.92) | 52.2 (0.72) | 53.5 (0.41) | | 28 | 51.7 (1.15) | 53.8 (1.52) |
| | 29 | 50.9 (0.42) | 52.1 (0.96) | 52.1 (0.75) | 52.8 (0.41) | | 29 | 51.6 (1.15) | 54.0 (1.52) |
| | 30 | 52.1 (0.43) | 52.5 (0.99) | 51.9 (0.86) | 53.0 (0.44) | | 30 | 52.3 (1.08) | 54.5 (1.42) |
| | 31 | 56.3 (0.41) | 56.7 (0.90) | 52.2 (0.96) | 53.2 (0.51) | | 31 | 52.3 (1.08) | 54.6 (1.37) |
| | 32 | 54.9 (0.38) | 54.8 (0.97) | 54.5 (1.00) | 54.6 (0.64) | | 32 | 52.7 (1.01) | 54.6 (1.24) |
| | 33 | -- | -- | -- | -- | | 33 | 53.3 (0.99) | 55.0 (1.15) |
| 34 | -- | -- | -- | -- | 34 | 53.8 (0.91) | 55.6 (1.00) | ||
H1: data acquired on 07/27/2012.
H2: data acquired on 07/20/2012.
H3: data acquired on 04/08/2012.
50/50 mixture phantom images were acquired with small field of view (FOV) using the standard BL heights and the 160 mAs reference setting for and calibrated with data acquired with the large FOV for the H1, H2, and H3 FFDM units (Hologic units). The large FOV calibration results are provided for reference. The findings are provided as the mean and the standard deviation (parenthetically) of the respective distributions in PG units.