Accuracy of drug store blood pressure monitors: an observational study.

Katherine L Ross1, Sanchit Bhasin, Merne P Wilson, Samuel A Stewart, Thomas W Wilson.   

Abstract

OBJECTIVES: Publically accessible blood pressure monitors are widely used, but little information is available on their accuracy. We compared blood pressure readings of 17 drug store monitors with those obtained using a validated home monitor (Omron BP742CAN) and both with those taken at home using the Canadian Hypertension Education Program protocol.
MATERIALS AND METHODS: Duplicate readings were taken using the drug store monitor (VitaStat, n=6, and PharmaSmart, n=11) on the left arm and the Omron on the right in three participants: two normal and one untreated hypertensive patient. We used Bland-Altman methods for comparison. We explored the correlation with average home blood pressure readings.
RESULTS: Home average blood pressure for our three participants was 121±6/73±5, 106±6/62±4, and 142±8/81±7 mmHg. The mean systolic blood pressure difference (drug store-Omron) was -1.8±8.2 mmHg. Diastolic pressure difference was 1.7±5.6. Individual paired systolic differences varied from -19 to 14 mmHg. For the participant who required a large cuff, drug store systolic readings tended to be higher (4.1±6.7). In our three participants, drug store monitors as a group read higher than home systolic blood pressure: 7.5 [95% confidence interval (CI) 1.5-13.4], 1.2 (95% CI -4.0 to 6.4), and 1.0 (95% CI -2.5 to 4.4) mmHg. Diastolic blood pressure and heart rate differences were similar in magnitude.
CONCLUSION: On average, drug store monitors recorded lower systolic blood pressures and higher diastolic blood pressures than a validated monitor, but the difference was neither statistically nor clinically significant. Single reading comparisons showed a much broader range. In three participants, drug store monitors did reflect the average home blood pressure.

Entities:  

Mesh:

Year:  2013        PMID: 24192848     DOI: 10.1097/MBP.0000000000000003

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  2 in total

1.  Accuracy of monitors used for blood pressure checks in English retail pharmacies: a cross-sectional observational study.

Authors:  James Hodgkinson; Constantinos Koshiaris; Una Martin; Jonathan Mant; Carl Heneghan; Fd Richard Hobbs; Richard J McManus
Journal:  Br J Gen Pract       Date:  2016-03-29       Impact factor: 5.386

Review 2.  Physician-pharmacist collaborative practice and telehealth may transform hypertension management.

Authors:  Stefano Omboni; Mauro Tenti; Claudio Coronetti
Journal:  J Hum Hypertens       Date:  2018-12-13       Impact factor: 3.012