Abstract

OBJECTIVES: The aim of this in vivo study was to investigate how reproducible individuals find their habitual intercuspation.
METHODS: A new type of highly accurate measuring procedure was involved using an intraoral optical scanning device (Cerec Bluecam). First, a pilot test investigated the accuracy of the procedure within a standardized setting. With regard to the in vivo study, fifteen participants with full dentition were selected. For each of these individuals, eight scans were taken both in the morning and in the afternoon. Furthermore, during each session, scans were taken both in horizontal and upright position. In order to compare this to ubiquitous used processes, plaster replicas of five individuals were investigated as well. The scans were analysed, and the differences in the position of the lower jaw were calculated by a specialized superimposition program (n=570 comparisons/OraCheck).
RESULTS: The results showed that there was no significant difference between the time of day and the position of the patient. The overall mean ± SD value for locating the habitual intercuspation was 42 ± 34 μm, however ranging from 22 ± 9 μm to 77 ± 58 μm for single individuals. On the other hand, the differences in positioning plaster replicas reached a mean of 135 ± 77 μm.
CONCLUSIONS: The reproducibility of the habitual intercuspation can be obtained under in vivo conditions by a newly developed and highly accurate measuring procedure. Individuals with full dentition show values in average of 42 μm. CLINICAL SIGNIFICANCE: Determining the occlusal jaw relation is an important precondition in restorative dentistry and many methods are proposed for a proper occlusal registration. Although much is known about in vitro accuracy of these techniques, little is known how reproducible the habitual occlusal position itself is found between individuals.
Copyright © 2013 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Bite registration; Buccal scan; Centric occlusion; Habitual intercuspation; Reproducibility; Static occlusion

Mesh:

Year:  2013        PMID: 24189225     DOI: 10.1016/j.jdent.2013.09.010

Source DB:  PubMed          Journal:  J Dent        ISSN: 0300-5712            Impact factor:   4.379


  4 in total

1.  In vivo precision of conventional and digital methods for obtaining quadrant dental impressions.

Authors:  Andreas Ender; Moritz Zimmermann; Thomas Attin; Albert Mehl
Journal:  Clin Oral Investig       Date:  2015-11-07       Impact factor: 3.573

2.  Precision of the virtual occlusal record.

Authors:  Keyan P Botsford; Michael C Frazier; Ahmed A M Ghoneima; Achint Utreja; Surya S Bhamidipalli; Kelton T Stewart
Journal:  Angle Orthod       Date:  2019-03-28       Impact factor: 2.079

3.  In vivo evaluation of three-dimensional of volumetric changes using a CAD/CAM chair-side system: Technical procedure.

Authors:  Rubén Agustín-Panadero; Alberto Ferreiroa; Agustín Pascual-Moscardó; Antonio Fons-Font; María-Fernanda Solá-Ruíz
Journal:  J Clin Exp Dent       Date:  2017-03-01

4.  Individualized 3D-Printed Tissue Retraction Devices for Head and Neck Radiotherapy.

Authors:  Christopher Herpel; Franz Sebastian Schwindling; Thomas Held; Leo Christ; Kristin Lang; Martha Schwindling; Julius Moratin; Karim Zaoui; Tracy Moutsis; Peter Plinkert; Klaus Herfarth; Christian Freudlsperger; Peter Rammelsberg; Jürgen Debus; Sebastian Adeberg
Journal:  Front Oncol       Date:  2021-03-23       Impact factor: 6.244

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.