| Literature DB >> 24169688 |
Ömer Yılmaz1, Esma Dilek Üstün, Mustafa Kayan, Fatmanur Kayan, Aykut Recep Aktaş, Elif Nisa Unlü, Bumin Değirmenci, Meltem Cetin.
Abstract
BACKGROUND: Our purpose was to evaluate the effectiveness of different kilovolt (kV) uses in computed tomography pulmonary angiography (CTPA) in the diagnosis of pulmonary thromboembolism (PTE). We also aimed to establish the optimal kV value and investigate the possibility of obtaining appropriate imaging quality with minimal radiation dose.Entities:
Mesh:
Year: 2013 PMID: 24169688 PMCID: PMC3816752 DOI: 10.12659/MSM.889578
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
Image quality characteristics was assessed subjectively using a five-point scales.
| Scale and Score | Description |
|---|---|
| 1 | Very bad, no diagnosis possible |
| 2 | Low, confidence in making diagnosis degraded |
| 3 | Moderate but sufficient for diagnosis |
| 4 | Good |
| 5 | Excellent, enabling excellent differentiation of even small structures |
| 1 | None, no diagnosis of PE possible |
| 2 | Slight |
| 3 | Sufficient |
| 4 | Good |
| 5 | Excellent |
| 1 | Massive, no diagnosis possible |
| 2 | Definite, establishment of diagnosis impeded |
| 3 | Definite but image sufficient for establishment of diagnosis |
| 4 | Slight |
| 5 | None |
Mean age and gender distribution of groups.
| A Group | B Group | C Group | |
|---|---|---|---|
| Subjects, n | 30 | 30 | 30 |
|
| |||
| Gender, n (%) | |||
| Female | 13 (43.3) | 16 (53.3) | 18 (60) |
| Male | 17 (56.7) | 14 (46.7) | 12 (40) |
|
| |||
| Age (years), mean±SD | 66.1±12.1 | 60.8±18.5 | 62.8±14.8 |
Figure 1Axial thorax CT images show the pulmonary artery embolism (same window with and level value) (A) with 120 kV protocol, (B) with 100 kV protocol, (C) with 80 kV protocol.
Figure 2Axial thorax CT images show the image quality of main pulmonary artery (A) with 120 kV protocol, (B) with 100 kV protocol, (C) with 80 kV protocol.
Figure 4Axial thorax CT images show the density of the right-sided basal segment of pulmonary artery (A) with 120 kV protocol, (B) with 100 kV protocol, (C) with 80 kV protocol.
Comparison of pulmonary artery density, characteristics of image quality, radiation doses between groups.
| A Group | B Group | C Group | p-Value | |
|---|---|---|---|---|
| Pulmonary artery density (HU) | ||||
| Main pulmonary artery | 309.5±79.1 | 381.7±124.0 | 477.3±193.3 | A–C→p<0.001, B–C→p 0.026 |
| Apical pulmonary artery | 369.7±88.7 | 463.9±127.0 | 578.1±271.1 | A–C→p 0.001, B–C→p=0.042 |
| Basal pulmonary artery | 357.9±88.1 | 449.6±123.0 | 565.3±248.7 | A–C→p<0.001, B–C→p 0.025 |
| Characteristics of image quality | ||||
| Image quality | 4.8±0.2 | 4.7±0.4 | 4.6±0.6 | p>0.05 (between all groups) |
| Enhancement | 4.5±0.6 | 4.7±0.4 | 4.6±0.7 | p>0.05 (between all groups) |
| Motion artefact | 4.9±0.1 | 5±0 | 5±0 | p>0.05 (between all groups) |
| Radiation doses | ||||
| DLP (mGy.cm) | 277.4±67.7 | 144.4±34.0 | 68.4±10.3 | p<0.001 (between all groups) |
| Effective dose (mSv) | 4.7±0.5 | 2.5±0.3 | 1.2±0.1 | p<0.001 (between all groups) |