| Literature DB >> 24144803 |
Laurens W Wollersheim1, Wilson W Li, Jan van der Meulen, Bas A de Mol.
Abstract
We describe a case of a 76-year old male who presented with progressive dyspnoea. He underwent an aortic valve replacement with a Freedom SOLO bioprosthesis 6 years ago. Transthoracic echocardiography showed a moderate-to-severe leakage of the Freedom SOLO bioprosthesis. During surgical reintervention, a partial tear of the left coronary cusp was seen from the commissure of the right coronary cusp to its base. After radiographic and microscopic examination, no clear cause was found for the failure of this Freedom SOLO bioprosthesis. To our knowledge, this is the third failure of a Freedom SOLO bioprosthesis reported in the literature. When the long-term follow-up of the Freedom SOLO bioprosthesis is available, it has to be compared with other bioprosthesis for long-term durability.Entities:
Keywords: Aortic valve replacement; Sorin Freedom SOLO bioprosthesis; Torn bioprosthesis
Mesh:
Year: 2013 PMID: 24144803 PMCID: PMC3867049 DOI: 10.1093/icvts/ivt439