OBJECTIVES: This study aimed to investigate the impact of lead diameter and design on implantable cardioverter-defibrillator (ICD) lead survival in children and young adults.
BACKGROUND: Recent reports have called attention to high rates of lead failure in adults with small-diameter ICD leads, but data in the pediatric population is limited.
METHODS: We reviewed lead performance in consecutive subjects ≤30 years with transvenous right ventricular ICD leads implanted at our center between January 1995 and October 2011. Lead failure was defined as fracture, perforation, or sensing failure necessitating revision.
RESULTS: A total of 120 ICD leads were implanted in 101 patients at a mean age of 15.5 ± 4.9 years. There were 47 small-diameter (≤8-F) and 73 standard-diameter (>8-F) leads. During a median follow-up of 28.7 months (interquartile range: 14.4 to 59.2 months), there were 25 lead failures (21% prevalence), with an incidence of 5.6%/year (95% confidence interval: 3.4 to 7.8). Sprint Fidelis (SF) (Medtronic, Inc., Minneapolis, Minnesota) leads had lower 3-year (69% vs. 92%, p < 0.01) and 5-year (44% vs. 86%, p < 0.01) survival probabilities than standard-diameter leads. In multivariate Cox regression, SF design conferred the greatest hazard ratio for lead failure (hazard ratio: 4.42, 95% confidence interval: 1.73 to 11.29, p < 0.01). Age and linear growth were not significantly associated with lead failure.
CONCLUSIONS: In this single-center pediatric study that evaluated lead diameter, lead design, and patient factors, the SF design conferred the highest risk of lead failure, suggesting that design rather than diameter is the critical issue in ICD lead performance.
Copyright © 2014 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  CHD; CI; HR; ICD; IQR; RV; SF; Sprint Fidelis; confidence interval; congenital heart disease; hazard ratio; implantable cardioverter-defibrillator; interquartile range; lead failure; pediatrics; right ventricle/ventricular; small-caliber

Mesh:

Year:  2013        PMID: 24140671     DOI: 10.1016/j.jacc.2013.09.033

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  4 in total

1.  Longitudinal Outcomes of Epicardial and Endocardial Pacemaker Leads in the Adult Fontan Patient.

Authors:  Geoffrey D Huntley; Abhishek J Deshmukh; Carole A Warnes; Suraj Kapa; Alexander C Egbe
Journal:  Pediatr Cardiol       Date:  2018-06-14       Impact factor: 1.655

2.  Initial experience with the subcutaneous implantable cardioverter-defibrillator with the real costs of hospitalization analysis in a single Polish center.

Authors:  Marcin Grabowski; Monika Gawałko; Marcin Michalak; Andrzej Cacko; Michał Kowara; Agnieszka Kołodzińska; Łukasz Januszkiewicz; Paweł Balsam; Laura Vitali Serdoz; Joachim Winter; Grzegorz Opolski
Journal:  Cardiol J       Date:  2018-04-03       Impact factor: 2.737

Review 3.  Channelopathies - emerging trends in the management of inherited arrhythmias.

Authors:  Priya Chockalingam; Yuka Mizusawa; Arthur Am Wilde
Journal:  Indian Pacing Electrophysiol J       Date:  2015-04-01

4.  2021 PACES expert consensus statement on the indications and management of cardiovascular implantable electronic devices in pediatric patients.

Authors:  Maully J Shah; Michael J Silka; Jennifer N Avari Silva; Seshadri Balaji; Cheyenne M Beach; Monica N Benjamin; Charles I Berul; Bryan Cannon; Frank Cecchin; Mitchell I Cohen; Aarti S Dalal; Brynn E Dechert; Anne Foster; Roman Gebauer; M Cecilia Gonzalez Corcia; Prince J Kannankeril; Peter P Karpawich; Jeffery J Kim; Mani Ram Krishna; Peter Kubuš; Martin J LaPage; Douglas Y Mah; Lindsey Malloy-Walton; Aya Miyazaki; Kara S Motonaga; Mary C Niu; Melissa Olen; Thomas Paul; Eric Rosenthal; Elizabeth V Saarel; Massimo Stefano Silvetti; Elizabeth A Stephenson; Reina B Tan; John Triedman; Nicholas H Von Bergen; Philip L Wackel
Journal:  Indian Pacing Electrophysiol J       Date:  2021-07-29
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.