Test-retest reliability of three dimensional gait analysis: including a novel approach to visualising agreement of gait cycle waveforms with Bland and Altman plots.

Dara Meldrum1, Ciara Shouldice, Ronan Conroy, Kim Jones, Malcolm Forward.   

Abstract

BACKGROUND/AIM: Estimating the measurement error (reliability) of three dimensional gait analysis (3DGA) is crucial to interpretation of gait data. The purpose of this study was to investigate the intra-rater reliability of 3DGA and apply a novel method of visualising reliability of gait cycle waveforms.
METHODS: A test re-test design was employed. A convenience sample of 30 healthy adults (18F; 12 M: mean age 30 ± 6.8 years) participated. Subjects walked along a 10 m walkway at their preferred gait speed and 3DGA data were collected using a VICON(®) 3DGA system. Testing was performed by the same investigator on two separate days within two weeks. Data from 10 trials were averaged and analysed. ANALYSIS: The intraclass correlation coefficient (ICC), the standard error of measurement (SEM), minimal detectable change (MDC) and limits of agreement were calculated for kinetic and kinematic data. Bland and Altman plots were applied to gait cycle waveforms.
RESULTS: Spatio-temporal parameters such as cadence, step length, velocity, step time and step width were highly repeatable generating ICC's of 0.90 and above and low SEM/MDC. Range of joint movement across the gait cycle was generally more reliable than either minimum or maximum values and higher ICCs were obtained for movement in the sagittal plane. For kinematic data the standard error of measurement was low (≤ 5) for the majority of parameters. Transverse plane measurements showed poor reliability with lowest ICC's. ICCs for kinetic data ranged from 0.51 to 0.81.
CONCLUSIONS: Reliability of 3DGA has been estimated for our gait laboratory. Bland and Altman plots of gait cycle waveforms provide a useful addition to reliability analysis.
Copyright © 2013 Elsevier B.V. All rights reserved.

Keywords:  Bland and Altman plots; Gait; Gait analysis; Measurement error; Reliability

Mesh:

Year:  2013        PMID: 24139682     DOI: 10.1016/j.gaitpost.2013.07.130

Source DB:  PubMed          Journal:  Gait Posture        ISSN: 0966-6362            Impact factor:   2.840


  23 in total

1.  Influence of Posture-Cuing Shirt on Tennis Serve Kinematics in Division III Tennis Players.

Authors:  Joseph Zappala; Carolina Orrego; Emily Boe; Heather Fechner; Derek Salminen; Daniel J Cipriani
Journal:  J Chiropr Med       Date:  2016-11-18

Review 2.  [Biomechanical screening for injury prevention : The importance of 3D-motion analysis in high performance sports].

Authors:  H Dewitz; B Yildirim; P Klein
Journal:  Unfallchirurg       Date:  2018-06       Impact factor: 1.000

3.  Multiple modes of assessment of gait are better than one to predict incident falls.

Authors:  Gilles Allali; Emmeline I Ayers; Joe Verghese
Journal:  Arch Gerontol Geriatr       Date:  2015-03-06       Impact factor: 3.250

4.  Absolute Reliability of Gait Parameters Acquired With Markerless Motion Capture in Living Domains.

Authors:  Sherveen Riazati; Theresa E McGuirk; Elliott S Perry; Wandasun B Sihanath; Carolynn Patten
Journal:  Front Hum Neurosci       Date:  2022-06-16       Impact factor: 3.473

5.  Verification of gait analysis method fusing camera-based pose estimation and an IMU sensor in various gait conditions.

Authors:  Masataka Yamamoto; Koji Shimatani; Yuto Ishige; Hiroshi Takemura
Journal:  Sci Rep       Date:  2022-10-21       Impact factor: 4.996

6.  Reliability of upper limb and trunk joint angles in healthy adults during activities of daily living.

Authors:  Susannah M Engdahl; Deanna H Gates
Journal:  Gait Posture       Date:  2017-11-08       Impact factor: 2.840

7.  Walking improvements with nabiximols in patients with multiple sclerosis.

Authors:  G Coghe; M Pau; F Corona; J Frau; L Lorefice; G Fenu; G Spinicci; E Mamusa; L Musu; S Massole; R Massa; M G Marrosu; E Cocco
Journal:  J Neurol       Date:  2015-08-05       Impact factor: 4.849

8.  Test-Retest Reliability of a Conventional Gait Model for Registering Joint Angles during Initial Contact and Toe-Off in Healthy Subjects.

Authors:  Francisco Molina-Rueda; Pilar Fernández-González; Alicia Cuesta-Gómez; Aikaterini Koutsou; María Carratalá-Tejada; Juan Carlos Miangolarra-Page
Journal:  Int J Environ Res Public Health       Date:  2021-02-02       Impact factor: 3.390

9.  Inertial Measurement Units for Clinical Movement Analysis: Reliability and Concurrent Validity.

Authors:  Mohammad Al-Amri; Kevin Nicholas; Kate Button; Valerie Sparkes; Liba Sheeran; Jennifer L Davies
Journal:  Sensors (Basel)       Date:  2018-02-28       Impact factor: 3.576

10.  Effects of Simulated Marker Placement Deviations on Running Kinematics and Evaluation of a Morphometric-Based Placement Feedback Method.

Authors:  Sean T Osis; Blayne A Hettinga; Shari Macdonald; Reed Ferber
Journal:  PLoS One       Date:  2016-01-14       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.