Literature DB >> 24136316

Value of routine plain x-ray position checks after cochlear implantation.

Franziska Dirr1, John Martin Hempel, Eike Krause, Joachim Müller, Alexander Berghaus, Birgit Ertl-Wagner, Thomas Braun.   

Abstract

OBJECTIVE: To analyze the value of a routine x-ray position check after cochlear implantation and to assess if an increased resistance during electrode insertion is a sufficient predictor of electrode misplacement. STUDY
DESIGN: Retrospective data collection.
SETTING: University hospital.
METHODS: Plain x-rays (Stenvers' projection) and the respective surgery reports of 218 patients having received cochlear implantation (243 ears) were analyzed for possible electrode misplacements and intraoperative conspicuities during electrode insertion.
RESULTS: Electrode misplacement (tip-over, loop, kinking, scalar transition, and incomplete insertion) was observed in 8% of the entire study cohort, but only in 5% if cases with inner ear dysplasia or labyrinthine ossification (n = 28) were excluded from analysis. Intraoperatively, an increased resistance during electrode insertion was found in 16% but only in 8% when cases with inner ear dysplasia or labyrinthine ossification were excluded. The intraoperative finding of an increased resistance during electrode insertion had a sensitivity of 55% and a specificity of 88% for predicting radiographically confirmed electrode misplacements (positive predictive value, 29%; negative predicting value, 96%).
CONCLUSION: Nearly half of the cases of electrode misplacement would have been overlooked if radiographic position checks would have been done only in patients with intraoperative conspicuities during electrode insertion. This finding advocates routine radiographic position checks, although electrode misplacements are relatively rare in patients with regular inner ear anatomy.

Entities:  

Mesh:

Year:  2013        PMID: 24136316     DOI: 10.1097/MAO.0b013e3182a09cc3

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  8 in total

1.  Auditory Brainstem Implant Array Position Varies Widely Among Adult and Pediatric Patients and Is Associated With Perception.

Authors:  Samuel R Barber; Elliott D Kozin; Aaron K Remenschneider; Sidharth V Puram; Max Smith; Barbara S Herrmann; Mary E Cunnane; M Christian Brown; Daniel J Lee
Journal:  Ear Hear       Date:  2017 Nov/Dec       Impact factor: 3.570

2.  Incomplete and false tract insertions in cochlear implantation: retrospective review of surgical and auditory outcomes.

Authors:  Ashish Vashishth; Andrea Fulcheri; Maurizio Guida; Antonio Caruso; Mario Sanna
Journal:  Eur Arch Otorhinolaryngol       Date:  2018-03-05       Impact factor: 2.503

3.  Evaluation of a minimally invasive surgical fixation technique for young children with the Concerto Pin cochlear implant system.

Authors:  Johannes Schnabl; Astrid Wolf-Magele; Stefan Marcel Pok; Christoph Url; Patrick Zorowka; Georg Sprinzl
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-08       Impact factor: 2.503

4.  A Protocol for Imaging of Cochlear Implantation.

Authors:  Stephanie J Wong; Alexis M Strohl-Bryan; Benjamin T Crane
Journal:  Otol Neurotol       Date:  2021-07-01       Impact factor: 2.619

Review 5.  Review on cochlear implant electrode array tip fold-over and scalar deviation.

Authors:  Anandhan Dhanasingh; Claude Jolly
Journal:  J Otol       Date:  2019-01-09

6.  Suitable Electrode Choice for Robotic-Assisted Cochlear Implant Surgery: A Systematic Literature Review of Manual Electrode Insertion Adverse Events.

Authors:  Paul Van de Heyning; Peter Roland; Luis Lassaletta; Sumit Agrawal; Marcus Atlas; Wolf-Dieter Baumgartner; Kevin Brown; Marco Caversaccio; Stefan Dazert; Wolfgang Gstoettner; Rudolf Hagen; Abdulrahman Hagr; Greg Eigner Jablonski; Mohan Kameswaran; Vladislav Kuzovkov; Martin Leinung; Yongxin Li; Andreas Loth; Astrid Magele; Robert Mlynski; Joachim Mueller; Lorne Parnes; Andreas Radeloff; Chris Raine; Gunesh Rajan; Joachim Schmutzhard; Henryk Skarzynski; Piotr H Skarzynski; Georg Sprinzl; Hinrich Staecker; Timo Stöver; Dayse Tavora-Viera; Vedat Topsakal; Shin-Ichi Usami; Vincent Van Rompaey; Nora M Weiss; Wilhelm Wimmer; Mario Zernotti; Javier Gavilan
Journal:  Front Surg       Date:  2022-03-24

7.  Vertigo Associated With Cochlear Implant Surgery: Correlation With Vertigo Diagnostic Result, Electrode Carrier, and Insertion Angle.

Authors:  Charlotte Weinmann; Uwe Baumann; Martin Leinung; Timo Stöver; Silke Helbig
Journal:  Front Neurol       Date:  2021-06-11       Impact factor: 4.003

8.  Intra-operative radiological diagnosis of a tip roll-over electrode array displacement using fluoroscopy, when electrophysiological testing is normal: the importance of both techniques in cochlear implant surgery.

Authors:  Octavio Garaycochea; Raquel Manrique-Huarte; Manuel Manrique
Journal:  Braz J Otorhinolaryngol       Date:  2017-06-01
  8 in total