Literature DB >> 24108364

A new nasopharyngeal dynamic reference frame improves accuracy in navigated skull base targets.

Florian Kral1, Matthew DiFranco2, Johanna Puschban2, Romed Hoermann3, Herbert Riechelmann2, Wolfgang Freysinger2.   

Abstract

OBJECTIVE: We questioned whether the position of the dynamic reference frame (DRF) influences the application accuracy in electromagnetically navigated cranial procedures. A carrier for an electromagnetic DRF was developed, which could be fixed at the posterior edge of the vomer near the center of the head. This nasopharyngeal DRF was compared with a standard DRF fixed to the surface of the forehead.
METHODS: Image coordinates and real-world coordinates were co-registered and the total target error (TTE) was measured in the frontal and the lateral skull base of formalin fixed human head. At each anatomical site, 10 targets served for TTE determinations and 5 different fiducial combinations were used for registration.
RESULTS: With the nasopharyngeal DRF, lower TTE values (2.8 ± 1.4 mm; mean ± SD) were observed when compared with the forehead DRF (3.7 ± 2.8 mm; P = .004). TTEs of both anatomical sites investigated were significantly lower when using the nasopharyngeal DRF (frontal skull base 3.4 vs 2.1 mm, P = .005 and lateral skull base 3.9 vs 3.5 mm, P = .013) than with the standard forehead mounted one.
CONCLUSION: Positioning the DRF in the center of the head significantly improved the application accuracy of targets in the skull base with electromagnetic navigation by 25%.
© The Author(s) 2013.

Entities:  

Keywords:  frontal skull base; lateral skull base surgery; navigation; otosurgery; registration

Mesh:

Year:  2013        PMID: 24108364     DOI: 10.1177/1553350613503738

Source DB:  PubMed          Journal:  Surg Innov        ISSN: 1553-3506            Impact factor:   2.058


  1 in total

1.  Coffee: the key to safer image-guided surgery-a granular jamming cap for non-invasive, rigid fixation of fiducial markers to the patient.

Authors:  Patrick S Wellborn; Neal P Dillon; Paul T Russell; Robert J Webster
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-03-30       Impact factor: 2.924

  1 in total