| Literature DB >> 24099061 |
Michael Hombach1, Reinhard Zbinden, Erik C Böttger.
Abstract
BACKGROUND: Standardisation of disk diffusion readings could improve reproducibility and accuracy of antibiotic susceptibility testing (AST). This study evaluated accuracy, reproducibility, and precision of automated inhibition zone reading using the "Sirscan automatic" zone reader (i2a, Perols Cedex, France).Entities:
Mesh:
Substances:
Year: 2013 PMID: 24099061 PMCID: PMC3852248 DOI: 10.1186/1471-2180-13-225
Source DB: PubMed Journal: BMC Microbiol ISSN: 1471-2180 Impact factor: 3.605
Mean differences of zone diameters measurements as determined by calliper and Sirscan on-screen adjusted
| | | ||
|---|---|---|---|
| Penicillins | 0.9 | 1.4 | 2.5 |
| Cephalosporins | 1 | | |
| Carbapenems | 1.4 | | |
| Aminoglycosides | 0.6 | 1.3 | |
| Quinolones | 0.9 | 1.4 | |
| Trimethoprim-Sulfamethoxazole | 0.8 | 0.9 | |
| Rifampicin | | 1.1 | |
| Glycopeptides | | | 0.8 |
| Cefoxitin | | 0.7 | |
| Clindamycin | | 1.6 | |
| All antibiotics | 0.9 | 1.2 | 1.7 |
Antibiotic drug classes / drugs tested for Gram-negative rods comprised ampicillin, amoxicillin/clavulanic acid, piperacillin/tazobactam (Penicillins), cefuroxime, cefpodoxime, ceftriaxone, ceftazidime, cefotaxime, cefepime, cefoxitin, (cephalosporins), ertapenem, imipenem, meropenem (carbapenems), amikacin, gentamicin, tobramycin (aminoglycosides), nalidixic acid, ciprofloxacin, levofloxacin (quinolones), and trimethoprim-sulfamethoxazole. Antibiotic drug classes / drugs tested for Staphylococcus spp. comprised penicillin (penicillins), cefoxitin, amikacin, gentamicin, tobramycin (aminoglycosides), ciprofloxacin, levofloxacin (quinolones), rifampicin, erythromycin, clindamycin, and trimethoprim-sulfamethoxazole. Antibiotic drugs tested for Enterococcus spp. comprised ampicillin (penicillins) and vancomycin (glycopeptides).
Relative deviation of zone diameter values and resulting discrepancies of the Sirscan (on-screen adjusted) and manual calliper measurements
| | ||||||
|---|---|---|---|---|---|---|
| Gram-negative rods | 19 | 45 | 36 | 1.27 | 0 | 0 |
| 27 | 37 | 36 | 0.94 | 0 | 0 | |
| 53 | 35 | 12 | 0 | 0 | 0 | |
For discrepancy analysis manual calliper measurements were regarded as the gold standard. Sirscan values were on-screen adjusted by an experienced person as recommended by the manufacturer.
Comparison of standard deviations of measurements with calliper, the Sirscan system adaped on-screen by the human eye and the Sirscan fully automated mode
| | | | | | | | | | | | | | | | | | | | | ||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 20-26 | 18-24 | 19-25 | 21-27 | 23-29 | 12-18 | 24-30 | 23-29 | 23-29 | 29-32 | 30-36 | | | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | 23.2 | 23.4 | 23.6 | 27.8 | 28.1 | 15.9 | 25.0 | 27.8 | 29.4 | 29.4 | 32.7 | | | | | | | | | | |
| | 0.8 | 0.5 | 0.8 | 0.9 | 1.3 | 0.3 | 1.2 | 0.8 | 0.7 | 0.9 | 1.0 | | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | 24.7 | 25.5 | 25.2 | 27.8 | 29.5 | 15.8 | 26.1 | 30.4 | 29.9 | 29.8 | 33.6 | | | | | | | | | | |
| | 1.2 | 0.7 | 1.4 | 1.7 | 1 | 0.4 | 0.9 | 2.2 | 3 | 1.3 | 0.7 | | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | 23.4 | 23.2 | 23.8 | 22.7 | 27.2 | 17.2 | 26.1 | 26.2 | 26.7 | 26.2 | 32.9 | | | | | | | | | | |
| | 1 | 1.6 | 1.2 | 1.1 | 2 | 0.6 | 0.8 | 1.1 | 1.2 | 4.7 | 2.1 | | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | | | ||
| | |||||||||||||||||||||
| 18-26 | 19-26 | 19-26 | 22-28 | 28-35 | 30-40 | 29-37 | 16-22 | 18-24 | 21-27 | 20-26 | 23-29 | 25-31 | 23-28 | 29-35 | 31-37 | 28-34 | 29-36 | 23-29 | 17-23 | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | 22.4 | 20.6 | 20.4 | 25.9 | 28.1 | 28.4 | 28.4 | 20.6 | 21.8 | 23.3 | 24.8 | 26.0 | 27.5 | 24.7 | 29.3 | 31.0 | 29.2 | 34.0 | 26.3 | 17.5 | |
| | 1.1 | 0.8 | 0.5 | 0.8 | 0.6 | 1.0 | 0.7 | 0.9 | 0.6 | 0.4 | 0.4 | 0.0 | 0.5 | 0.5 | 0.7 | 0.7 | 0.9 | 0.8 | 0.4 | 0.5 | |
| | | | | | | | | | | | | | | | | | | | | | |
| | 23.2 | 24.5 | 25.1 | 25.9 | 34 | 37.3 | 35.4 | 25.9 | 23.3 | 27.6 | 26.1 | 27.8 | 31.1 | 29.3 | 32.3 | 35 | 35.9 | 34.3 | 28.4 | 24.8 | |
| | 1.5 | 1.3 | 1.4 | 1.3 | 2.4 | 1.6 | 1.5 | 1.8 | 1 | 1 | 1.2 | 1.2 | 1.9 | 1.2 | 1.3 | 1 | 1.2 | 1.2 | 1.4 | 1 | |
| | | | | | | | | | | | | | | | | | | | | | |
| | 22.4 | 24.4 | 23.3 | 27.6 | 31.1 | 34.1 | 31.5 | 22.3 | 25.1 | 25.7 | 24.8 | 25.5 | 28.5 | 27.6 | 30.2 | 33.7 | 23.9 | 34.3 | 27.3 | 17.4 | |
| | 1.1 | 2.5 | 1.2 | 1.3 | 1 | 1.6 | 1.5 | 1.2 | 0.7 | 1.1 | 1.2 | 1 | 1.5 | 2.4 | 1.4 | 1.4 | 1.3 | 2.4 | 0.9 | 1.6 | |
| | | | | | | | | | | | | | | | | | | | | ||
| | | | | | | | | | | | |||||||||||
| 19-25 | 18-26 | 16-21 | 19-26 | 24-30 | 21-27 | 23-29 | 20-28 | 27-33 | | | | | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | 24 | 24.9 | 21.5 | 28 | 27.8 | 22.9 | 25.3 | 23.6 | 29.9 | | | | | | | | | | | | |
| | 0.8 | 0.7 | 1.4 | 0.6 | 0.4 | 0.3 | 0.7 | 0.5 | 0.3 | | | | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | 23.2 | 25.2 | 22 | 27.8 | 26.6 | 22.2 | 24.5 | 25 | 26.5 | | | | | | | | | | | | |
| | 0.8 | 1 | 0.9 | 1.3 | 1.4 | 0.9 | 1.2 | 0.5 | 0.6 | | | | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | 23.5 | 25.0 | 21.6 | 25.9 | 25.8 | 22.2 | 23.9 | 24.9 | 26.4 | | | | | | | | | | | | |
| | 0.6 | 0.7 | 0.5 | 1.2 | 0.9 | 0.8 | 1.1 | 0.6 | 0.9 | ||||||||||||
Standard deviations of repeat measurements of S. aureus ATCC 29213 and E. coli ATCC 25922 were significantly lower with fully automated Sirscan readings as compared to manual calliper measurements indicating better reproducibility and precision of Sirscan readings. Asterisks indicate statistically significant differences (p<0.05) of mean standard deviations using the paired t-test. Measurements were done independently and double-blinded by 19 experienced persons (technicians and laboratory physicians) with the same disk diffusion plates of EUCAST quality control strains of S. aureus ATCC 29213, E. coli ATCC 25922, and P. aeruginosa ATCC 27853. Measurements of the Sirscan fully automated mode comprise 19 independent measurements of the panels. QC, quality control; AM, ampicillin; AMC, amoxicillin-clavulanic acid; AK, amikacin; CAZ, ceftazidime; CIP, ciprofloxacin; CN, gentamicin; CPD, cefpodoxime; CRO, ceftriaxone; CTX, cefotaxime; CXM, cefuroxime; DA, clindamycin; E, erythromycin; ETP, ertapenem; FEP, cepefim; FOX, cefoxitin; IPM, imipenem; LEV, levofloxacin; MEM, meropenem; NA, nalidixic acid; NF, nitrofuratoine; NOR, norfloxacin; P, penicillin G; RA, rifampicin; SXT, trimethoprim-sulfamethoxazole.
Examples of scattergrams of inhibition zone measurements with calliper, the Sirscan system adaped on-screen by the human eye and the Sirscan fully automated mode
| | ||||||||||||||||||||
| Diameter (mm) | | | | | | | | | | | ||||||||||
| Sirscan fully automated | | | 9 | 10 | | | | | | | | | | | | | | | | |
| Sirscan on-screen adjusted | | 6 | 4 | 5 | 2 | 2 | | | | | | | | | | | | | | |
| Calliper | 3 | 3 | 4 | 3 | 5 | 1 | | | | | | | | | | | | | | |
| | ||||||||||||||||||||
| Diameter (mm) | | | | | | | | | | |||||||||||
| Sirscan fully automated | | | | | | 3 | 7 | 9 | | | | | | | | | | | | |
| Sirscan on-screen adjusted | | | | | 1 | | 4 | 6 | 2 | 3 | 3 | | | | | | | | | |
| Calliper | 1 | | | 1 | 1 | 4 | 3 | 1 | 5 | 3 | | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | | | |
| | ||||||||||||||||||||
| Diameter (mm) | ||||||||||||||||||||
| Sirscan fully automated | | | | | | | | | | | 4 | 6 | 7 | 2 | | | | | | |
| Sirscan on-screen adjusted | | | | | | | | | | | 1 | 4 | 3 | 7 | 4 | | | | | |
| Calliper | 2 | 1 | 1 | | | 1 | | 1 | 2 | 3 | 2 | 1 | 2 | 1 | 1 | | 1 | | | |
| | ||||||||||||||||||||
| Diameter (mm) | | | | | | | | | | | | |||||||||
| Sirscan fully automated | | | | | | 7 | 12 | | | | | | | | | | | | | |
| Sirscan on-screen adjusted | | | | | | 1 | 6 | 8 | 4 | | | | | | | | | | | |
| Calliper | 1 | 5 | 3 | 7 | 3 | |||||||||||||||
Measurements were done independently and double-blinded by 19 experienced persons (technicians and laboratory physicians) with the same disk diffusion plates of EUCAST quality control strains of S. aureus ATCC 29213, and E. coli ATCC 25922. Measurements of the Sirscan fully automated mode comprise 19 independent measurements of the panels. EUCAST quality control ranges are indicated in italics.