Literature DB >> 24082311

Remotely controlled mandibular protrusion during sleep predicts therapeutic success with oral appliances in patients with obstructive sleep apnea.

John Remmers1, Shouresh Charkhandeh, Joshua Grosse, Zbigniew Topor, Rollin Brant, Peter Santosham, Sabina Bruehlmann.   

Abstract

STUDY
OBJECTIVES: The present study addresses the need for a validated tool that prospectively identifies favorable candidates for oral appliance therapy in treatment of obstructive sleep apnea. The objective of the study was to evaluate the ability of a mandibular titration study, performed with a remotely controlled mandibular positioner (RCMP), to predict treatment outcome with a mandibular repositioning appliance (MRA) and to predict an effective target protrusive position (ETPP).
DESIGN: A prospective, blinded, outcome study.
SETTING: Standard clinical care with tests performed in the polysomnographic laboratory. PARTICIPANTS: Consecutive patients (n = 67) recruited from a sleep center or a dental practice using broad inclusion criteria (age 21-80 years; AHI > 10/h; BMI < 40 kg/m(2)).
INTERVENTIONS: Therapeutic outcome with a mandibular protruding oral appliance was predicted following a mandibular protrusive titration study in the polysomnographic laboratory using a remotely controlled positioner and prospectively established predictive rules. An ETPP was also prospectively determined for participants predicted to be therapeutically successful with MRA therapy. All participants were blindly treated with a MRA, at either the predicted ETPP or a sham position, and therapeutic outcome was compared against prediction. MEASUREMENTS AND
RESULTS: At the final protrusive position, standard predictive parameters (sensitivity, specificity, positive and negative predictive values) showed statistically significant predictive accuracy (P < 0.05) in the range of 83% to 94%. The predicted ETPP provided an efficacious protrusive position in 87% of participants predicted to be therapeutically successful with MRA therapy (P < 0.05).
CONCLUSIONS: Using prospectively established rules for interpreting the polysomnographic data, the mandibular titration study predicted mandibular repositioning appliance therapeutic outcome with significant accuracy, particularly with regard to accurately predicting therapeutic success. As well, among the participants predicted to be therapeutically successful with mandibular repositioning appliance therapy, the effective target protrusive position provided efficacious mandibular protrusion in the majority.

Entities:  

Keywords:  Obstructive sleep apnea; mandibular repositioning titration; oral appliance therapy; sleep disordered breathing

Mesh:

Year:  2013        PMID: 24082311      PMCID: PMC3773201          DOI: 10.5665/sleep.3048

Source DB:  PubMed          Journal:  Sleep        ISSN: 0161-8105            Impact factor:   5.849


  29 in total

Review 1.  Improving CPAP use by patients with the sleep apnoea/hypopnoea syndrome (SAHS).

Authors:  Heather M Engleman; Matt R Wild
Journal:  Sleep Med Rev       Date:  2003-02       Impact factor: 11.609

2.  Mandibular advancement and obstructive sleep apnoea: a method for determining effective mandibular protrusion.

Authors:  L C Dort; E Hadjuk; J E Remmers
Journal:  Eur Respir J       Date:  2006-05       Impact factor: 16.671

3.  Advancement of the mandible improves velopharyngeal airway patency.

Authors:  S Isono; A Tanaka; Y Sho; A Konno; T Nishino
Journal:  J Appl Physiol (1985)       Date:  1995-12

4.  A randomized crossover study of an oral appliance vs nasal-continuous positive airway pressure in the treatment of mild-moderate obstructive sleep apnea.

Authors:  K A Ferguson; T Ono; A A Lowe; S P Keenan; J A Fleetham
Journal:  Chest       Date:  1996-05       Impact factor: 9.410

5.  Sleepiness-related accidents in sleep apnea patients.

Authors:  S Horstmann; C W Hess; C Bassetti; M Gugger; J Mathis
Journal:  Sleep       Date:  2000-05-01       Impact factor: 5.849

6.  Effect of a titration polysomnogram on treatment success with a mandibular repositioning appliance.

Authors:  Fernanda R Almeida; Jonathan A Parker; James S Hodges; Alan A Lowe; Kathleen A Ferguson
Journal:  J Clin Sleep Med       Date:  2009-06-15       Impact factor: 4.062

Review 7.  Oral appliances for snoring and obstructive sleep apnea: a review.

Authors:  Kathleen A Ferguson; Rosalind Cartwright; Robert Rogers; Wolfgang Schmidt-Nowara
Journal:  Sleep       Date:  2006-02       Impact factor: 5.849

8.  A comparison of responders and nonresponders to oral appliance therapy for the treatment of obstructive sleep apnea.

Authors:  Ryo Otsuka; Fernanda Ribeiro de Almeida; Alan A Lowe; Frank Ryan
Journal:  Am J Orthod Dentofacial Orthop       Date:  2006-02       Impact factor: 2.650

Review 9.  Obstructive sleep apnea: implications for cardiac and vascular disease.

Authors:  Abu S M Shamsuzzaman; Bernard J Gersh; Virend K Somers
Journal:  JAMA       Date:  2003-10-08       Impact factor: 56.272

10.  Initial Evaluation of a Titration Appliance for Temporary Treatment of Obstructive Sleep Apnea.

Authors:  Daniel J Levendowski; Todd Morgan; Philip Westbrook
Journal:  J Sleep Disord Ther       Date:  2011-12-24
View more
  35 in total

Review 1.  Sleep medicine care under one roof: a proposed model for integrating dentistry and medicine.

Authors:  Sunil Sharma; Greg Essick; David Schwartz; Amy J Aronsky
Journal:  J Clin Sleep Med       Date:  2013-08-15       Impact factor: 4.062

2.  Another advance in oral appliance therapy?

Authors:  Patrick J Strollo
Journal:  Sleep       Date:  2013-10-01       Impact factor: 5.849

3.  Reply: Objective measurement of the therapeutic effectiveness of continuous positive airway pressure versus oral appliance therapy for the treatment of obstructive sleep apnea.

Authors:  David P White
Journal:  Am J Respir Crit Care Med       Date:  2013-11-01       Impact factor: 21.405

4.  Oral Appliance Treatment Response and Polysomnographic Phenotypes of Obstructive Sleep Apnea.

Authors:  Kate Sutherland; Hisashi Takaya; Jin Qian; Peter Petocz; Andrew T Ng; Peter A Cistulli
Journal:  J Clin Sleep Med       Date:  2015-08-15       Impact factor: 4.062

5.  Think before sinking your teeth into oral appliance therapy.

Authors:  Joachim Ngiam; Peter A Cistulli
Journal:  J Clin Sleep Med       Date:  2014-12-15       Impact factor: 4.062

6.  A Feedback-Controlled Mandibular Positioner Identifies Individuals With Sleep Apnea Who Will Respond to Oral Appliance Therapy.

Authors:  John E Remmers; Zbigniew Topor; Joshua Grosse; Nikola Vranjes; Erin V Mosca; Rollin Brant; Sabina Bruehlmann; Shouresh Charkhandeh; Seyed Abdolali Zareian Jahromi
Journal:  J Clin Sleep Med       Date:  2017-07-15       Impact factor: 4.062

7.  Performance of Remotely Controlled Mandibular Protrusion Sleep Studies for Prediction of Oral Appliance Treatment Response.

Authors:  Kate Sutherland; Joachim Ngiam; Peter A Cistulli
Journal:  J Clin Sleep Med       Date:  2017-03-15       Impact factor: 4.062

8.  CPAP pressure for prediction of oral appliance treatment response in obstructive sleep apnea.

Authors:  Kate Sutherland; Craig L Phillips; Amanda Davies; Vasanth K Srinivasan; Oyku Dalci; Brendon J Yee; M Ali Darendeliler; Ronald R Grunstein; Peter A Cistulli
Journal:  J Clin Sleep Med       Date:  2014-09-15       Impact factor: 4.062

9.  Remotely Controlled Mandibular Positioning During Drug-Induced Sleep Endoscopy Toward Mandibular Advancement Device Therapy: Feasibility and Protocol.

Authors:  Chloé Kastoer; Marijke Dieltjens; Sara Op de Beeck; Marc J Braem; Paul H Van de Heyning; Olivier M Vanderveken
Journal:  J Clin Sleep Med       Date:  2018-08-15       Impact factor: 4.062

Review 10.  The Use of Remotely Controlled Mandibular Positioner as a Predictive Screening Tool for Mandibular Advancement Device Therapy in Patients with Obstructive Sleep Apnea through Single-Night Progressive Titration of the Mandible: A Systematic Review.

Authors:  Chloé Kastoer; Marijke Dieltjens; Eline Oorts; Evert Hamans; Marc J Braem; Paul H Van de Heyning; Olivier M Vanderveken
Journal:  J Clin Sleep Med       Date:  2016-10-15       Impact factor: 4.062

View more