Inaccuracy of estimated resting oxygen uptake in the clinical setting.

Nikhil Narang1, Jennifer T Thibodeau, Benjamin D Levine, M Odette Gore, Colby R Ayers, Richard A Lange, Joaquin E Cigarroa, Aslan T Turer, James A de Lemos, Darren K McGuire.   

Abstract

BACKGROUND: The Fick principle (cardiac output = oxygen uptake ( O2)/systemic arterio-venous oxygen difference) is used to determine cardiac output in numerous clinical situations. However, estimated rather than measured O2 is commonly used because of complexities of the measurement, though the accuracy of estimation remains uncertain in contemporary clinical practice. METHODS AND
RESULTS: From 1996 to 2005, resting O2 was measured via the Douglas bag technique in adult patients undergoing right heart catheterization. Resting O2 was estimated by each of 3 published formulae. Agreement between measured and estimated O2 was assessed overall, and across strata of body mass index, sex, and age. The study included 535 patients, with mean age 55 yrs, mean body mass index 28.4 kg/m2; 53% women; 64% non-white. Mean (±standard deviation) measured O2 was 241 ± 57 ml/min. Measured O2 differed significantly from values derived from all 3 formulae, with median (interquartile range) absolute differences of 28.4 (13.1, 50.2) ml/min, 37.7 (19.4, 63.3) ml/min, and 31.7 (14.4, 54.5) ml/min, for the formulae of Dehmer, LaFarge, and Bergstra, respectively (P<0.0001 for each). The measured and estimated values differed by >25% in 17% to 25% of patients depending on the formula used. Median absolute differences were greater in severely obese patients (body mass index > 40 kg/m2), but were not affected by sex or age.
CONCLUSIONS: Estimates of resting O2 derived from conventional formulae are inaccurate, especially in severely obese individuals. When accurate hemodynamic assessment is important for clinical decision-making, O2 should be directly measured.

Entities:  

Keywords:  cardiac output; catheterization; hemodynamics

Mesh:

Year:  2013        PMID: 24077170     DOI: 10.1161/CIRCULATIONAHA.113.003334

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  23 in total

1.  Discordance Between Clinical Assessment and Invasive Hemodynamics in Patients With Advanced Heart Failure.

Authors:  Nikhil Narang; Ben Chung; Ann Nguyen; Rohan J Kalathiya; Luke J Laffin; Luise Holzhauser; Imo A Ebong; Stephanie A Besser; Teruhiko Imamura; Bryan A Smith; Sara Kalantari; Jayant Raikhelkar; Nitasha Sarswat; Gene H Kim; Valluvan Jeevanandam; Daniel Burkhoff; Gabriel Sayer; Nir Uriel
Journal:  J Card Fail       Date:  2019-08-21       Impact factor: 5.712

2.  Effects of Hollow Fiber Membrane Oscillation on an Artificial Lung.

Authors:  Ryan A Orizondo; Guy Gino; Garret Sultzbach; Shalv P Madhani; Brian J Frankowski; William J Federspiel
Journal:  Ann Biomed Eng       Date:  2018-02-20       Impact factor: 3.934

3.  Thermodilution vs Estimated Fick Cardiac Output Measurement in Clinical Practice: An Analysis of Mortality From the Veterans Affairs Clinical Assessment, Reporting, and Tracking (VA CART) Program and Vanderbilt University.

Authors:  Alexander R Opotowsky; Edward Hess; Bradley A Maron; Evan L Brittain; Anna E Barón; Thomas M Maddox; Laith I Alshawabkeh; Bradley M Wertheim; Meng Xu; Tufik R Assad; Jonathan D Rich; Gaurav Choudhary; Ryan J Tedford
Journal:  JAMA Cardiol       Date:  2017-10-01       Impact factor: 14.676

4.  The Impact of Direct Cardiac Output Determination On Using A Widely Available Direct Continuous Oxygen Consumption Measuring Device On The Hemodynamic Assessment of Aortic Valve.

Authors:  Zaher Fanari; Matthew Grove; Anitha Rajamanickam; Sumaya Hammami; Cassie Walls; Paul Kolm; Mitchell Saltzberg; William S Weintraub; Andrew J Doorey
Journal:  Del Med J       Date:  2016-09

5.  Geographic Variation in the Treatment of U.S. Adult Heart Transplant Candidates.

Authors:  William F Parker; Allen S Anderson; Donald Hedeker; Elbert S Huang; Edward R Garrity; Mark Siegler; Matthew M Churpek
Journal:  J Am Coll Cardiol       Date:  2018-04-24       Impact factor: 24.094

6.  Resting Oxygen Consumption and Heart Failure: Importance of Measurement for Determination of Cardiac Output With the Use of the Fick Principle.

Authors:  Gillian Grafton; Thomas M Cascino; Daniel Perry; Carmel Ashur; Todd M Koelling
Journal:  J Card Fail       Date:  2019-02-10       Impact factor: 5.712

7.  Hemodynamic Effects of Weight Loss in Obesity: A Systematic Review and Meta-Analysis.

Authors:  Yogesh N V Reddy; Mahesh Anantha-Narayanan; Masaru Obokata; Katlyn E Koepp; Patricia Erwin; Rickey E Carter; Barry A Borlaug
Journal:  JACC Heart Fail       Date:  2019-07-10       Impact factor: 12.035

8.  Cardiac output determination using a widely available direct continuous oxygen consumption measuring device: a practical way to get back to the gold standard.

Authors:  Zaher Fanari; Matthew Grove; Anitha Rajamanickam; Sumaya Hammami; Cassie Walls; Paul Kolm; Mitchell Saltzberg; William S Weintraub; Andrew J Doorey
Journal:  Cardiovasc Revasc Med       Date:  2016-02-27

9.  In Vitro Characterization of the Pittsburgh Pediatric Ambulatory Lung.

Authors:  Ryan A Orizondo; Alexandra G May; Shalv P Madhani; Brian J Frankowski; Greg W Burgreen; Peter D Wearden; William J Federspiel
Journal:  ASAIO J       Date:  2018 Nov/Dec       Impact factor: 2.872

Review 10.  Pulmonary Vascular Disease: Hemodynamic Assessment and Treatment Selection-Focus on Group II Pulmonary Hypertension.

Authors:  Bhavadharini Ramu; Brian A Houston; Ryan J Tedford
Journal:  Curr Heart Fail Rep       Date:  2018-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.