Comparative evaluation of the accuracy of linear measurements between cone beam computed tomography and 3D microtomography.

Francesca Mangione1, Deborah Meleo, Marco Talocco, Raffaella Pecci, Luciano Pacifici, Rossella Bedini.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the influence of artifacts on the accuracy of linear measurements estimated with a common cone beam computed tomography (CBCT) system used in dental clinical practice, by comparing it with microCT system as standard reference.
MATERIALS AND METHODS: Ten bovine bone cylindrical samples containing one implant each, able to provide both points of reference and image quality degradation, have been scanned by CBCT and microCT systems. Thanks to the software of the two systems, for each cylindrical sample, two diameters taken at different levels, by using implants different points as references, have been measured. Results have been analyzed by ANOVA and a significant statistically difference has been found. RESULTS AND DISCUSSION: Due to the obtained results, in this work it is possible to say that the measurements made with the two different instruments are still not statistically comparable, although in some samples were obtained similar performances and therefore not statistically significant.
CONCLUSION: With the improvement of the hardware and software of CBCT systems, in the near future the two instruments will be able to provide similar performances.

Entities:  

Mesh:

Year:  2013        PMID: 24071605     DOI: 10.4415/ANN_13_03_05

Source DB:  PubMed          Journal:  Ann Ist Super Sanita        ISSN: 0021-2571            Impact factor:   1.663


  3 in total

1.  Accuracy of linear measurements around dental implants by means of cone beam computed tomography with different exposure parameters.

Authors:  Lauren O L Bohner; Pedro Tortamano; Juliana Marotti
Journal:  Dentomaxillofac Radiol       Date:  2017-03-23       Impact factor: 2.419

2.  Peri-implant osseointegration after low-level laser therapy: micro-computed tomography and resonance frequency analysis in an animal model.

Authors:  Luciano Mayer; Fernando Vacilotto Gomes; Marília Gerhardt de Oliveira; João Feliz Duarte de Moraes; Lennart Carlsson
Journal:  Lasers Med Sci       Date:  2016-08-17       Impact factor: 3.161

3.  Effect of Dental Implant Metal Artifacts on Accuracy of Linear Measurements by Two Cone-Beam Computed Tomography Systems Before and After Crown Restoration.

Authors:  Hoorieh Bashizadeh Fakhar; Roxana Rashtchian; Milad Parvin
Journal:  J Dent (Tehran)       Date:  2017-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.