BACKGROUND & AIMS: Indirect calorimetry was performed for a long time with the DeltatracII(®) device (Datex, Finland), considered as a reference but no longer produced. This study aims at comparing the energy expenditure (EE), the volume of oxygen (VO2) and carbon dioxide (VCO2) measured by two new available indirect calorimeters, the QuarkRMR(®) (Cosmed, Italy) and the CCMexpress(®) (MedGraphic,USA), using three different methods of gas collection, with the DeltatracII(®) in healthy subjects.
METHODS: Twenty-four healthy subjects (15 women and 9 men, age 53 ± 15 yrs, mean BMI 25.5 ± 7.1 kg/m(2)) underwent measurements of EE with DeltatracII(®) using canopy, QuarkRMR(®) using canopy and CCMexpress(®) using canopy, face tent and facemask. All measurements were performed for 10 min in random order. Results are presented as mean ± SD and compared by linear regression, repeated measure one-way ANOVA with Bonferroni's post hoc test and Bland & Altman test.
RESULTS: EE was 1630 ± 340 kcal for DeltatracII(®) and 1607 ± 307 kcal, 1741 ± 360 kcal, 1666 ± 315 and 1626 ± 336 kcal for QuarkRMR(®) and CCMexpress(®) with canopy, face tent and facemask, respectively (p = 0.001). Compared to DeltatracII(®), Bland & Altman test showed a mean EE difference (2SD) of 24(220)kcal for QuarkRMR(®), and -111(260) kcal, -36(304) kcal, 5(402) kcal for CCMexpress(®) with canopy, face tent and facemask, respectively. There was no systematic over- or underestimation with any device or gas collection method.
CONCLUSION: Mean EE was similar between QuarkRMR(®) and DeltatracII(®) but not between CCMexpress(®), in any mode of gas collection, and DeltatracII(®). Bland & Altman test shows a large variability in EE differences with both devices compared to DeltatracII(®), highlighting the need for refining their accuracy.
Copyright © 2013 Elsevier Ltd and European Society for Clinical Nutrition and Metabolism. All rights reserved.

Entities:  

Keywords:  Energy expenditure; Indirect calorimetry; Validation

Mesh:

Substances:

Year:  2013        PMID: 24064252     DOI: 10.1016/j.clnu.2013.08.012

Source DB:  PubMed          Journal:  Clin Nutr        ISSN: 0261-5614            Impact factor:   7.324


  13 in total

1.  Validity Test of a New Open-Circuit Indirect Calorimeter.

Authors:  Christine M Ashcraft; David C Frankenfield
Journal:  JPEN J Parenter Enteral Nutr       Date:  2014-03-10       Impact factor: 4.016

2.  Algorithm to improve accuracy of energy expended in a room calorimeter.

Authors:  Haiying Quan; Wenrui Hao; Lu Li; Ming Sun; Kuan Zhang
Journal:  Med Biol Eng Comput       Date:  2016-10-20       Impact factor: 2.602

3.  Methods for Estimating Energy Expenditure in Critically Ill Adults.

Authors:  Makayla Cordoza; Lingtak-Neander Chan; Elizabeth Bridges; Hilaire Thompson
Journal:  AACN Adv Crit Care       Date:  2020-09-15

4.  Separate and combined effects of 21-day bed rest and hypoxic confinement on body composition.

Authors:  Tadej Debevec; Tarsi C Bali; Elizabeth J Simpson; Ian A Macdonald; Ola Eiken; Igor B Mekjavic
Journal:  Eur J Appl Physiol       Date:  2014-08-05       Impact factor: 3.078

5.  Comparison of the GEM and the ECAL indirect calorimeters against the Deltatrac for measures of RMR and diet-induced thermogenesis.

Authors:  S Kennedy; L Ryan; A Fraser; M E Clegg
Journal:  J Nutr Sci       Date:  2014-11-07

6.  Validity of Predictive Equations for Resting Energy Expenditure Developed for Obese Patients: Impact of Body Composition Method.

Authors:  Najate Achamrah; Pierre Jésus; Sébastien Grigioni; Agnès Rimbert; André Petit; Pierre Déchelotte; Vanessa Folope; Moïse Coëffier
Journal:  Nutrients       Date:  2018-01-10       Impact factor: 5.717

7.  VO2 and VCO2 variabilities through indirect calorimetry instrumentation.

Authors:  Miguel Cadena-Méndez; Boris Escalante-Ramírez; Joaquín Azpiroz-Leehan; Oscar Infante-Vázquez
Journal:  Springerplus       Date:  2013-12-23

8.  Measuring energy expenditure in the intensive care unit: a comparison of indirect calorimetry by E-sCOVX and Quark RMR with Deltatrac II in mechanically ventilated critically ill patients.

Authors:  Martin Sundström Rehal; Erik Fiskaare; Inga Tjäder; Åke Norberg; Olav Rooyackers; Jan Wernerman
Journal:  Crit Care       Date:  2016-03-05       Impact factor: 9.097

Review 9.  The influence of metabolic imbalances and oxidative stress on the outcome of critically ill polytrauma patients: a review.

Authors:  Alexandru Florin Rogobete; Dorel Sandesc; Marius Papurica; Emil Robert Stoicescu; Sonia Elena Popovici; Lavinia Melania Bratu; Corina Vernic; Adriana Mariana Sas; Adrian Tudor Stan; Ovidiu Horea Bedreag
Journal:  Burns Trauma       Date:  2017-03-07

10.  Acute Effects of High Intensity, Resistance, or Combined Protocol on the Increase of Level of Neurotrophic Factors in Physically Inactive Overweight Adults: The BrainFit Study.

Authors:  María A Domínguez-Sanchéz; Rosa H Bustos-Cruz; Gina P Velasco-Orjuela; Andrea P Quintero; Alejandra Tordecilla-Sanders; Jorge E Correa-Bautista; Héctor R Triana-Reina; Antonio García-Hermoso; Katherine González-Ruíz; Carlos A Peña-Guzmán; Enrique Hernández; Jhonatan C Peña-Ibagon; Luis A Téllez-T; Mikel Izquierdo; Robinson Ramírez-Vélez
Journal:  Front Physiol       Date:  2018-06-27       Impact factor: 4.566

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.