Algorithm to improve accuracy of energy expended in a room calorimeter.

Haiying Quan1,2, Wenrui Hao1,2, Lu Li1,2, Ming Sun3, Kuan Zhang4,5.   

Abstract

The whole-room indirect calorimeter is considered as important equipment for human energy expenditure measurement, but noise reduction in the system remains a challenge. A selective filtering method (SFM) was designed to improve the accuracy of the computation of O2 consumption rate ([Formula: see text]) and CO2 production rate ([Formula: see text]), based on two facts: (1) the rapid changes of [Formula: see text], [Formula: see text] and respiratory quotient (RQ) in human should be accompanied by physical activity; (2) the oxygen consumption and the carbon dioxide production should not be negative because living humans do not generate oxygen, nor consume carbon dioxide. The performance of SFM was compared with the moving average method, the central difference method and the wavelet de-noising method. The range of [Formula: see text] and [Formula: see text] in the empty room (the background noise) is reduced from -130.00-146.00 ml/min to -26.00-24.00 ml/min, and from -20.50-12.50 ml/min to -3.99-4.19 ml/min, by SFM. The background noise was added to simulated rectangular and sinusoidal signals that were used to evaluate the four methods over different time periods (64, 32, 16 and 8 min). The highest signal-to-noise ratio and the lowest deviation were achieved by SFM. Abnormal metabolic rates and RQs were corrected and compensated with measurement accuracy of 98.51 ± 0.3 % for 24-h alcohol burning tests. The results of the study showed that SFM can significantly improve [Formula: see text] and [Formula: see text] measurements.

Entities:  

Keywords:  Energy expenditure; Metabolic rate; Respiratory quotient; Selective filtering method; Whole-room indirect calorimeter

Mesh:

Substances:

Year:  2016        PMID: 27766518     DOI: 10.1007/s11517-016-1583-9

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  21 in total

1.  Measurement of energy expenditure.

Authors:  James P DeLany
Journal:  Pediatr Blood Cancer       Date:  2011-10-18       Impact factor: 3.167

2.  Description of a direct-indirect room-sized calorimeter.

Authors:  J L Seale; W V Rumpler; P W Moe
Journal:  Am J Physiol       Date:  1991-02

3.  Chamber for indirect calorimetry with improved transient response.

Authors:  B Henning; R Löfgren; L Sjöström
Journal:  Med Biol Eng Comput       Date:  1996-05       Impact factor: 2.602

4.  Introducing a new generation indirect calorimeter for estimating energy requirements in adult intensive care unit patients: feasibility, practical considerations, and comparison with a mathematical equation.

Authors:  Elisabeth De Waele; Herbert Spapen; Patrick M Honoré; Sabrina Mattens; Viola Van Gorp; Marc Diltoer; Luc Huyghens
Journal:  J Crit Care       Date:  2013-04-03       Impact factor: 3.425

5.  On-line computer analysis and breath-by-breath graphical display of exercise function tests.

Authors:  W L Beaver; K Wasserman; B J Whipp
Journal:  J Appl Physiol       Date:  1973-01       Impact factor: 3.531

6.  Analysis of gaseous exchange in open-circuit indirect calorimetry.

Authors:  D Brown; T J Cole; M J Dauncey; R W Marrs; P R Murgatroyd
Journal:  Med Biol Eng Comput       Date:  1984-07       Impact factor: 2.602

Review 7.  Indirect calorimetry: a guide for optimizing nutritional support in the critically ill child.

Authors:  Racheli Sion-Sarid; Jonathan Cohen; Zion Houri; Pierre Singer
Journal:  Nutrition       Date:  2013-09       Impact factor: 4.008

8.  Repeatability of 24-h energy expenditure measurements in humans by indirect calorimetry.

Authors:  W V Rumpler; J L Seale; J M Conway; P W Moe
Journal:  Am J Clin Nutr       Date:  1990-02       Impact factor: 7.045

9.  Closed-loop control of carbon dioxide concentration and pressure improves response of room respiration calorimeters.

Authors:  J K Moon; F A Vohra; O S Valerio Jimenez; M R Puyau; N F Butte
Journal:  J Nutr       Date:  1995-02       Impact factor: 4.798

10.  Chamber for indirect calorimetry with accurate measurement and time discrimination of metabolic plateaus of over 20 min.

Authors:  T Nguyen; L de Jonge; S R Smith; G A Bray
Journal:  Med Biol Eng Comput       Date:  2003-09       Impact factor: 3.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.