Biomechanical strength analysis of mini anchors for the temporomandibular joint.

Andrezza Lauria1, Danillo Costa Rodrigues, Fábio Ricardo Loureiro Sato, Roger William Fernandes Moreira.   

Abstract

INTRODUCTION: The usage of mini anchors for the treatment of internal derangements of the temporomandibular joint (TMJ) is considered an innovative treatment, and it has presented excellent clinical results in the stabilization of the articular disc. This study aimed to evaluate, through mechanical tensile testing, the resistance of mini anchors for TMJ articular disc repositioning.
MATERIALS AND METHODS: Ten thermoplastic polymer poly-ether-ether-ketone (PEEK) mini anchors were tested in artificial polyurethane blocks with a pullout test in a mechanical Instron™ machine, model 4411. The mini anchors were pulled out until failure and the force and displacement were recorded for each specimen. A small standard deviation was noted, which indicated reproducibility.
RESULTS: The findings indicate that Cillen™ mini anchors presented a mean force at failure up to 46.1 N with similar values in the scientific literature. Nevertheless, further studies should be conducted for a comparison of PEEK mini anchors with other commercially available anchors, in addition to studies related to PEEK, a scarcely known material in the dentistry field.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24062014     DOI: 10.1007/s10006-013-0431-4

Source DB:  PubMed          Journal:  Oral Maxillofac Surg        ISSN: 1865-1550


  15 in total

1.  Mechanical properties evaluation of an orthodontic miniscrew system for skeletal anchorage.

Authors:  Daniele Francioli; Grazia Ruggiero; Roberto Giorgetti
Journal:  Prog Orthod       Date:  2010-05-15       Impact factor: 2.750

2.  Use of the Mitek anchor in temporomandibular joint disc-repositioning surgery.

Authors:  P Mehra; L M Wolford
Journal:  Proc (Bayl Univ Med Cent)       Date:  2001-01

3.  What do you do when you have a loose screw?

Authors:  Paul C Brady; Paolo Arrigoni; Stephen S Burkhart
Journal:  Arthroscopy       Date:  2006-09       Impact factor: 4.772

Review 4.  Nonsurgical management of disc-interference disorders.

Authors:  J P Okeson
Journal:  Dent Clin North Am       Date:  1991-01

5.  The pullout force for Mitek mini and micro suture anchor systems in human mandibular condyles.

Authors:  R T Fields; L E Cardenas; L M Wolford
Journal:  J Oral Maxillofac Surg       Date:  1997-05       Impact factor: 1.895

6.  The Mitek mini anchor for TMJ disc repositioning: surgical technique and results.

Authors:  P Mehra; L M Wolford
Journal:  Int J Oral Maxillofac Surg       Date:  2001-12       Impact factor: 2.789

7.  The osseointegration of Mitek mini anchors in the mandibular condyle.

Authors:  R T Fields; L M Wolford
Journal:  J Oral Maxillofac Surg       Date:  2001-12       Impact factor: 1.895

Review 8.  Temporomandibular joint devices: treatment factors and outcomes.

Authors:  L M Wolford
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  1997-01

9.  Mechanical evaluation of different techniques for symphysis fracture fixation--an in vitro polyurethane mandible study.

Authors:  Thalita Regina Vieira e Oliveira; Luis Augusto Passeri
Journal:  J Oral Maxillofac Surg       Date:  2011-03-12       Impact factor: 1.895

10.  Biomechanical evaluation of plating techniques for fixing mandibular angle fractures: the introduction of a new 3D plate approach.

Authors:  Evagelos F Kalfarentzos; Despoina Deligianni; Georgios Mitros; Minos Tyllianakis
Journal:  Oral Maxillofac Surg       Date:  2009-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.