| Literature DB >> 25022449 |
Volker Rudat1, Alaa Nour, Abdul Aziz Alaradi, Adel Mohamed, Saleh Altuwaijri.
Abstract
PURPOSE: The purpose of this study was to compare the surface dose of 7-field IMRT (7 F-IMRT), tangential beam IMRT (TB-IMRT), and tangential beam 3D-CRT (3D-CRT) of breast cancer patients receiving adjuvant radiotherapy by means of in vivo GafChromic film dosimetry.Entities:
Mesh:
Year: 2014 PMID: 25022449 PMCID: PMC4120005 DOI: 10.1186/1748-717X-9-156
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
PTV, fractionation regimen, radiotherapy technique and timing of the in vivo Gafchromic film dosimetry
| | | | | |||||
|---|---|---|---|---|---|---|---|---|
| Patient series 1: 3D-CRT versus TB-IMRT | ||||||||
| Whole breast | 17 (7) | 1.80 | 28 | 3DCRT | 1-25 | TB-IMRT | 26-28 | 23-28 |
| Whole breast | 1 (1) | 2.67 | 15 | 3DCRT | 1-12 | TB-IMRT | 13-15 | 10-15 |
| Chest wall | 14 (7) | 2.00 | 25 | 3DCRT | 1-22 | TB-IMRT | 23-25 | 20-25 |
| Patient series 2: TB-IMRT versus 7 F-IMRT | ||||||||
| Whole breast | 7 (0) | 1.80 | 28 | TB-IMRT | 1-25 | 7 F-IMRT | 26-28 | 23-28 |
| Whole breast | 1 (0) | 2.67 | 15 | TB-IMRT | 1-12 | 7 F-IMRT | 13-15 | 10-15 |
| Chest wall | 5 (4) | 2.00 | 25 | TB-IMRT | 1-22 | 7 F-IMRT | 23-25 | 20-25 |
Abbreviations: PTV planning target volume, n/p Total number of patients (number of patients with left-sided breast cancer), D/f (Gy) Dose per fraction (Gy), n/f Number of fractions, RT technique Radiotherapy technique, In vivo FD In vivo Gafchromic film dosimetry, 3D-CRT Conformal 3-dimensional radiotherapy, TB-IMRT Tangential beam intensity modulated radiotherapy, 7 F-IMRT 7-field intensity modulated radiotherapy.
Figure 1Illustration of the OAR skin and the position of the in vivo Gafchromic dosimetry films using the image of a 7 F-IMRT plan in the isocenter plane. The OAR skin was defined as a 3 mm strip from the surface of the skin to the PTV. PTV for the IMRT plans included the same PTV used for the 3D-CRT plans plus an extension into the air anteriorly of the skin of 1.5 cm to ensure appropriate opening of the 160 MLC Multileaf Collimator. The numbers denote the position of the in vivo Gafchromic dosimetry films (1 = lateral position, 2 = central position, 3 = medial position).
Figure 2EBT3 film calibration curve.
Figure 3Illustration of the beam configuration used for 3D-CRT and TB-IMRT (A) and 7 F-IMRT (B).
Dose-volume constraints for IMRT plans
| PTV | Target | 1 | Maximum | 5200 | 0 | 100 |
| PTV | Target | 1 | Minimum | 4900 | 100 | 100 |
| Ipsilateral lung | Organ at risk | 2 | Maximum | 2000 | 20 | 100 |
| Ipsilateral lung | Organ at risk | 2 | Minimum | 1200 | 30 | 100 |
| Heart | Organ at risk | 3 | Maximum | 4500 | 0 | 100 |
| Unspecified tissue | Organ at risk | 4 | Maximum | 4500 | 0 | 100 |
Figure 4Comparison of two radiotherapy plans of the same patient. (A), 3D-CRT versus TB-IMRT; (B), TB-IMRT versus 7 F-IMRT. The yellow lines represent the 95% isodose lines, and the red line the 90% isodose lines.
Surface dose by means of in vivo Gafchromic film dosimetry and calculated skin dose in dependence of the PTV and radiation technique
| | ||||||||
|---|---|---|---|---|---|---|---|---|
| | | |||||||
| | ||||||||
| Patient series 1: 3D-CRT versus TB-IMRT | | | | | | | | |
| Whole breast | | | | | | | | |
| 3D-CRT | 53.8 | 6.0 | 42.2 | 5.0 | 48.5 | 7.1 | 91.8 | 13.3 |
| TB-IMRT | 53.4 | 5.4 | 41.2 | 6.5 | 49.1 | 8.4 | 90.8 | 8.7 |
| Difference (%) | -0.7 | | -2.4 | | 1.2 | | -1.1 | |
| p-value* | n.s. | | n.s. | | n.s. | | n.s. | |
| Chest wall | | | | | | | | |
| 3D-CRT | 58.1 | 5.7 | 47.2 | 9.0 | 49.8 | 5.1 | 92.2 | 6.0 |
| TB-IMRT | 58.3 | 6.3 | 45.4 | 7.9 | 51.9 | 6.2 | 92.0 | 8.8 |
| Difference (%) | 0.3 | | -3.8 | | 4.2 | | -0.2 | |
| p-value* | n.s. | | n.s. | | n.s. | | n.s. | |
| Patient series 2: TB-IMRT versus 7 F-IMRT | | | | | | | | |
| Whole breast | | | | | | | | |
| TB-IMRT | 54.0 | 3.9 | 45.0 | 5.6 | 41.7 | 4.5 | 86.8 | 6.9 |
| 7 F-IMRT | 45.6 | 2.3 | 43.6 | 7.0 | 35.8 | 7.4 | 80.6 | 10.8 |
| Difference (%) | -15.6 | | -3.1 | | -14.1 | | -7.1 | |
| p-value* | <0.01 | | n.s. | | 0.05 | | 0.03 | |
| Chest wall | | | | | | | | |
| TB-IMRT | 60.1 | 5.3 | 39.4 | 12.0 | 59.2 | 3.5 | 95.2 | 2.3 |
| 7 F-IMRT | 53.0 | 5.9 | 38.2 | 17.6 | 48.1 | 8.4 | 90.5 | 2.0 |
| Difference (%) | -11.8 | | -3.0 | | -18.8 | | -4.9 | |
| p-value* | 0.02 | n.s. | 0.05 | <0.01 | ||||
Abbreviations:†Proportion of the prescribed dose (%), SD Standard deviation, *Paired t-test, n.s. not significant, other abbreviations: see Table 1.
Relevant plan parameters of 3D-CRT versus TB-IMRT or TB-IMRT versus 7 F-IMRT of the breast or chest wall
| Skin | ||||||||
| Mean Dose (percent of the prescribed dose) | ||||||||
| | 3D-CRT | 92.0 | (88.1; 95.9) | n.s. | TB-IMRT | 90.0 | (85.8; 94.2) | <0.01 |
| | TB-IMRT | 91.4 | (88.1; 94.8) | | 7 F-IMRT | 84.4 | (78.6; 90.3) | |
| | Difference (%) | -0.7 | | | Difference (%) | -6.2 | | |
| PTV | ||||||||
| Homogeneity index | ||||||||
| | 3D-CRT | 0.73 | (0.65; 0.80) | n.s. | TB-IMRT | 0.75 | (0.68; 0.81) | n.s. |
| | TB-IMRT | 0.73 | (0.67; 0.78) | | 7 F-IMRT | 0.72 | (0.63; 0.81) | |
| | Difference (%) | 0 | | | Difference (%) | -4.0 | | |
| Conformity index | ||||||||
| | 3D-CRT | 0.76 | (0.71; 0.81) | <0.001 | TB-IMRT | 0.88 | (0.85; 0.91) | n.s. |
| | TB-IMRT | 0.86 | (0.82; 0.89) | | 7 F-IMRT | 0.88 | (0.85; 0.91) | |
| | Difference (%) | 13.2 | | | Difference (%) | 0 | | |
| Heart (left-sided PTV) | ||||||||
| Mean Dose (cGy) | ||||||||
| | 3D-CRT | 506 | (359; 654) | n.s. | TB-IMRT | 824 | (423; 1224) | 0.01 |
| | TB-IMRT | 504 | (380; 628) | | 7 F-IMRT | 1647 | (937; 2357) | |
| | Difference (%) | -0.4 | | | Difference (%) | 99.9 | | |
| V70% | | | | | | | | |
| | 3D-CRT | 5.12 | (2.49; 7.75) | 0.03 | TB-IMRT | 9.25 | (1.7; 16.8) | n.s. |
| | TB-IMRT | 2.57 | (1.06; 4.09) | | 7 F-IMRT | 10.00 | (1.5; 18.5) | |
| | Difference (%) | -49.8 | | | Difference (%) | 8.1 | | |
| Heart (right-sided PTV) | ||||||||
| Mean Dose (cGy) | ||||||||
| | 3D-CRT | 157 | (33; 282) | n.s. | TB-IMRT | 78 | (60; 97) | <0.001 |
| | TB-IMRT | 149 | (42; 282) | | 7 F-IMRT | 704 | (603; 805) | |
| | Difference (%) | -5.1 | | | Difference (%) | 802.6 | | |
| Ipsilateral lung | ||||||||
| Mean Dose (cGy) | ||||||||
| | 3D-CRT | 1276 | (1179; 1374) | <0.001 | TB-IMRT | 1031 | (842; 1221) | <0.001 |
| | TB-IMRT | 1036 | (943; 1129) | | 7 F-IMRT | 1535 | (1382; 1689) | |
| | Difference (%) | -18.8 | | | Difference (%) | 48.9 | | |
| D60% | | | | | | | | |
| | 3D-CRT | 194 | (174; 214) | n.s. | TB-IMRT | 139 | (101; 177) | <0.001 |
| | TB-IMRT | 187 | (168; 206) | | 7 F-IMRT | 686 | (458; 913) | |
| | Difference (%) | -3.6 | | | Difference (%) | 393.5 | | |
| D30% | | | | | | | | |
| | 3D-CRT | 1327 | (1036; 1618) | <0.001 | TB-IMRT | 934 | (535; 1334) | <0.001 |
| | TB-IMRT | 922 | (725; 1119) | | 7 F-IMRT | 1775 | (1459; 2091) | |
| | Difference (%) | -30.5 | | | Difference (%) | 90.0 | | |
| Contralateral lung | ||||||||
| Mean Dose (cGy) | ||||||||
| | 3D-CRT | 27 | (24; 31) | n.s. | TB-IMRT | 23 | (18; 28) | <0.001 |
| | TB-IMRT | 26 | (23; 29) | | 7 F-IMRT | 328 | (259; 397) | |
| Difference (%) | -3.7 | Difference (%) | 1326.1 | |||||
Abbreviations:95% CI 95% confidence interval, Vx% percentage of tissue volume encompassed by the x% isodose line, Dx% dose to x% of the volume of the organ at risk, *Paired t-test, n.s. Not significant, other abbreviations: see Table 1.