A case of anastomosis malposition of the Blalock-Taussig shunt diagnosed using perioperative transesophageal echocardiography monitoring.

Tomohiro Yamamoto1, Ehrenfried Schindler.   

Abstract

The perioperative transesophageal echocardiography (TEE) is a very useful and minimal invasive monitoring not only for the cardiac anesthesia management, but also for the anesthesia management of the non-cardiac operations of the heart high risk patients. In this case report, we report a case of the urgent Blalock-Taussig shunt (BT shunt) operation for a small patient of Fallot's tetralogy with anoxic spell, which showed an atypical change of hemodynamics and SpO2 with the shunt opening. After the BT shunt anastomosis, the diastolic blood pressure decreased with the shunt opening, however, the expected rise of SpO2 was not provided. By the perioperative TEE monitoring with the single plane TEE transducer for neonate; UST-52110S (Hitachi Aloka Medical, Tokyo, Japan) with 4.5 mm in diameter, the cause of this atypical change of hemodynamics and SpO2 was diagnosed to be an accidental anastomosis malposition of the BT shunt to the right pulmonary vein and reported to the operator during the operation, and the operation was performed correctly.

Entities:  

Mesh:

Year:  2013        PMID: 24046098     DOI: 10.1007/s10877-013-9509-7

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  6 in total

1.  The safety of intraoperative transesophageal echocardiography: a case series of 7200 cardiac surgical patients.

Authors:  I J Kallmeyer; C D Collard; J A Fox; S C Body; S K Shernan
Journal:  Anesth Analg       Date:  2001-05       Impact factor: 5.108

2.  Transesophageal echocardiography in myocardial revascularization: II. Influence on intraoperative decision making.

Authors:  B D Bergquist; W H Bellows; J M Leung
Journal:  Anesth Analg       Date:  1996-06       Impact factor: 5.108

3.  Intraoperative transoesophageal echocardiography: impact on adult cardiac surgery.

Authors:  D C Sutton; R Kluger
Journal:  Anaesth Intensive Care       Date:  1998-06       Impact factor: 1.669

4.  Impact of routine use of intraoperative transesophageal echocardiography during cardiac surgery.

Authors:  P Couture; A Y Denault; S McKenty; D Boudreault; F Plante; R Perron; D Babin; L Normandin; N Poirier
Journal:  Can J Anaesth       Date:  2000-01       Impact factor: 5.063

5.  Real-time intraoperative transesophageal echocardiography--how useful? Experience of 5,016 cases.

Authors:  M Mishra; R Chauhan; K K Sharma; A Dhar; M Bhise; S Dhole; A Omar; R R Kasliwal; N Trehan
Journal:  J Cardiothorac Vasc Anesth       Date:  1998-12       Impact factor: 2.628

6.  Safety of transesophageal echocardiography. A multicenter survey of 10,419 examinations.

Authors:  W G Daniel; R Erbel; W Kasper; C A Visser; R Engberding; G R Sutherland; E Grube; P Hanrath; B Maisch; K Dennig
Journal:  Circulation       Date:  1991-03       Impact factor: 29.690

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.