Literature DB >> 24039458

Validation and ease of use of a new pen device for self-administration of recombinant human growth hormone: results from a two-center usability study.

Robert Rapaport1, Paul Saenger, Heinrich Schmidt, Yukihiro Hasegawa, Michel Colle, Sandro Loche, Sandra Marcantonio, Walter Bonfig, Markus Zabransky, Fima Lifshitz.   

Abstract

Close adherence to the recommended treatment regimen is important for the success of recombinant human growth hormone therapy, although nonadherence can be common. Ease of use and safety during use/storage are among several important factors in the design of a growth hormone injection device intended for long-term use. This study was performed to validate the usability and assess the ease of use of a new pen device (SurePal™) that has been developed to support daily administration of the recombinant human growth hormone product, Omnitrope® (somatropin). The primary objectives of the study were to assess if study participants, representing intended users of the pen in clinical practice, were able to perform an injection procedure into an injection pad effectively and safely and disassemble the pen without receiving a needlestick injury. A total of 106 participants (61 adults and 45 children/adolescents) were enrolled at two study centers (one in the US, one in Germany). Results for both primary usability tasks met the predefined acceptance criteria, with >85% of participants successfully performing each task. All of the other tasks/handling steps assessed were also successfully performed by most participants, with high success rates reflected in the high proportion of participants who classified each task as "very easy" or "easy". After a second use of the device, 87%-97% of participants rated it as "very easy" or "easy" to use. In summary, the new pen device is safe and easy to use for both adults and children, and will help to support effective, long-term daily administration of the recombinant human growth hormone product, Omnitrope®.

Entities:  

Keywords:  Omnitrope®; SurePal™; ease of use; growth hormone; pen device

Year:  2013        PMID: 24039458      PMCID: PMC3770891          DOI: 10.2147/MDER.S50088

Source DB:  PubMed          Journal:  Med Devices (Auckl)        ISSN: 1179-1470


Introduction

Recombinant human growth hormone (rhGH) is used in the treatment of several pediatric conditions characterized by poor growth, including growth hormone deficiency, Turner syndrome, short children born small for gestational age, chronic renal insufficiency, and Prader–Willi syndrome.1–9 It is also used to treat metabolic dysfunction in adult growth hormone deficiency.10 Continuing rhGH treatment in young adults treated for childhood-onset growth hormone deficiency has a beneficial effect on bone mineral density in adult life.11 In order to achieve maximum benefit, treatment with rhGH should begin as soon as possible after diagnosis, and be continued uninterrupted (usually for many years).3,4 An important factor in the success of rhGH therapy is the ability of the patient to comply with the recommended treatment regimen, and failure to comply is associated with reduced linear growth.12 Ensuring compliance can be challenging, given that subcutaneous administration of rhGH is required over the long term. Indeed, noncompliance is fairly common among children and adults receiving this therapy.12–14 Factors affecting adherence include the physical discomfort of injections, the delay before beneficial effects are noticed, and, importantly, the ease of administration for patients or carers.5,15 Many studies have assessed interventions to help improve patient compliance with prescribed medications, with inconsistent results.16 Methods for improving compliance with long-term treatment appear to be mostly complex and not very effective, and research concerning innovations to assist compliance has been identified as a high priority.16 In the setting of rhGH therapy, manufacturers have focused on improvements in injection devices as a means to improve compliance. Studies have identified reliability, ease of use, lack of pain, safety during use/storage, and the number of steps involved in the procedure to be important in the design of a growth hormone injection device intended for long-term use.17–20 Growth hormone injection devices continue to evolve, with conventional syringes and needles having been replaced by more user-friendly pen injection devices. Various types of pen devices have been developed,18,21–25 each with distinct features designed for specific patient needs. The current study was performed to validate the usability and assess the ease of use of a new pen device (SurePal™, Kundl, Austria). This is a semi-reusable needle-based injection system (the pen itself is reusable, while the drug-containing cartridges are replaced when empty), which has been developed to support daily administration of the rhGH product, Omnitrope® (somatropin, Sandoz, Kundl, Austria).

Materials and methods

The primary objectives of the study were to assess if study participants, representing intended users of the pen in clinical practice, were able to perform an injection procedure into an injection pad effectively and safely and disassemble the pen without receiving a needlestick injury. Both tasks had to be successfully executed by >85% of participants for the usability of the pen to be validated. The required success rate for acceptance was calculated based on the limits of the confidence intervals of the defined sample size. These acceptance criteria were agreed to by the relevant body in the US (US Food and Drug Administration [FDA]) during device development and accepted by the relevant body in Europe (BSI) as part of the CE marking process. Additional objectives of the study included assessment of: success rates for specific tasks/handling steps required to prepare the pen for use; participants’ viewpoints on the ease with which each of these tasks can be performed; and participants’ opinions on the usefulness of the written instructions for use (IFU) that accompany the pen.

Study participants

The study was conducted in two centers, one in the US and one in Germany, and recruited children (or their adult caregivers), adolescents (or their adult caregivers), and adults being treated for growth hormone deficiency. To ensure enough participants, adults with diabetes experienced in the use of pen devices were also included at the German center (six of 15 adults). Participants were subdivided into naïve users (never used an injection device before) and experienced users (patients or caregivers experienced in administering injections, either through use of another growth hormone or insulin injection device). All study participants (or their caregivers) were volunteers and provided written informed consent.

Study device

The pen evaluated in the study was a multidose, semi-reusable needle-based injection system (Figure 1). This new device contains a number of safety features, including a system to ensure that only cartridges of the correct dose can be used with each device, and an optional needle guard. There is also a dose memory function that enables the correct dose to be preset by a doctor or nurse and locked into the device, thereby reducing the risk of incorrect dosing. In addition, there are several features designed to make the device easy and convenient to use, including autopriming and a sliding injection button that reduces the force needed to perform an injection. The device is also designed to minimize drug wastage; if a cartridge in the device does not contain sufficient drug for injection of a complete dose, the device automatically administers the correct amount of additional drug once a new cartridge is inserted (without the need for priming or adjustment of the dose setting). Information on whether the complete dose has been administered is provided to the user through a dose display window; if it does not show zero, the cartridge holds less than the complete dose and a new cartridge should be inserted. The automated nature of this feature makes it more convenient to administer a second injection compared with most other rhGH injection devices, for which the user must calculate the amount of additional dose required and reset their device accordingly.
Figure 1

The pen device.

Usability tests

Training and test sessions were conducted at both centers as follows. In the US center, 75% of participants received a 15–20-minute training session then waited for at least 2 hours before returning for their test session, while 25% received no training (this was a requirement of the FDA, and participants who were to receive no training were selected at random). In the German center, all participants received training and then immediately commenced the test session. In the test session, participants were asked to perform the tasks/handling steps outlined in Table 1. All participants were allowed to read the IFU document as a supportive measure.
Table 1

Outline of the usability testing strategy

StepUS centerGerman center
1Introduction and background interview
275% received a 15–20-minute training session, 25% received no training100% received training and then immediately commenced the test session
3Trained patients waited for at least 2 hours before commencing the test session
4Set dose memory
5Preparation of pen (insertion of cartridge)
6Preparation of pen (attach needle and needle hider)
7Injection procedure
8Disassembly
9No drug wastage system including cartridge exchange
10Closing interview
The moderator evaluated the performance of each participant by observing the participant conducting different tasks. Participants were asked to comment on the various steps and to rate the ease of performing each step as “very easy”, “easy”, “undecided”, “difficult”, or “very difficult”. Participants were also instructed to rate the helpfulness of the IFU as “very helpful”, “helpful”, “undecided”, “not helpful”, or “not at all helpful”. A note-taker was present during the test session to document all observations and answers by means of a session protocol and a laptop computer. All sessions were recorded on video. Data were summarized using descriptive statistics.

Results

A total of 106 participants (61 adults and 45 children/adolescents) were enrolled at the two study centers. Key characteristics of the study participants are summarized in Table 2. Overall, 49% of children/adolescents and 48% of adults had previous experience with either a rhGH or insulin pen.
Table 2

Key characteristics of study participants

Children/adolescents (n = 45)*
 Gender
  Male19 (42)
  Female26 (58)
 Experienced with injections22 (49)
 Received training42 (93)
Adults (n = 61)
 Gender
  Male22 (36)
  Female39 (64)
 Experienced with injections29 (48)
 Received training55 (90)

Notes: Data are n (%).

Aged 8–16 years in Germany, 8–19 years in the US;

aged ≥18 years in Germany, ≥22 years in the US.

Primary usability tasks

Results for both primary usability tasks met the predefined acceptance criteria, with >85% of participants successfully performing each task (Figure 2). Overall, 92% of participants (92% of adults and 93% of children/adolescents) were able to perform the injection procedure into the injection pad successfully; 92% of participants (95% of adults and 89% of children/adolescents) rated this procedure as very easy or easy. The vast majority of participants (97% of adults and 91% of children/adolescents) were able to complete the task with no help at all, by referring to the IFU or with just verbal hints. The success rate was higher in the US center (98%) than in the German center (89%).
Figure 2

Primary usability tasks. (A) Proportion of patients who successfully performed task. (B) Proportion of patients who rated completing the task as very easy/easy.

In total, 99% of participants (98% of adults and all children/adolescents) were able to disassemble the pen successfully without receiving a needlestick injury; 84% (82% of adults and 87% of children/adolescents) rated disassembly of the pen as very easy or easy. The majority of participants (93% of adults and 89% of children/adolescents) were able to complete the task with no help at all, by referring to the IFU or with just verbal hints. Only one participant (a naïve adult user) received a needlestick injury. The success rate was 98% in the US center and 100% in the German center.

Additional study assessments

Overall, most participants found the pen device easier to use the more they used it. For example, among children/adolescents at the US center, 73% and 87% rated the first and second injection, respectively, as very easy/easy. Similarly, among the German children/adolescents, 70% rated the device as easy to use at first, rising to 97% on subsequent use. Data for the other tasks/handling steps that were assessed during the study are summarized in Table 3. Overall, 92% of participants (89% of US participants, 93% of German participants) were able to set the dose memory (program the correct dose). The majority of adults (85%) and children/adolescents (84%) were able to complete the task with no help at all, by referring to the IFU or with just verbal hints.
Table 3

Success rates for additional tasks/handling steps

TaskParticipants successfully completing task, n (%)
Adults (n = 61)Children/adolescents* (n = 45)All participants (n = 106)
Set dose memory58 (95)39 (87)97 (92)
Preparation of pen (insertion of cartridge)60 (98)42 (93)102 (96)
Preparation of pen (attach needle and needle hider)61 (100)43 (96)104 (98)
Low drug-waste system including cartridge exchange54 (89)37 (82)91 (86)

Notes:

Aged 8–16 years in Germany, 8–19 years in the US;

aged ≥18 years in Germany, ≥22 years in the US.

Overall, 96% of participants were able to reinsert the preassembled cartridge and attach it to the pen correctly (96% of US participants, 97% of German participants). Again, the majority of adults (97%) and children/adolescents (84%) were able to complete the task with no help at all, by referring to the IFU or with just verbal hints. Ninety-eight percent of participants (100% of US participants, 97% of German participants) were able to attach the needle and needle hider correctly, with 93% of adults and 93% of children/adolescents able to complete the task with no help at all, by referring to the IFU or with just verbal hints; 88% of the participants rated the procedure as being very easy/easy. Participants’ ability to manage the low drug-wastage system (including cartridge exchange) was also assessed. Overall, 86% successfully managed the system (80% of US participants, 90% of German participants); most (89% of adults and 82% of children/adolescents) were able to complete the task with no help at all, by referring to the IFU or with just verbal hints. Eighty-six per cent rated the procedure as very easy/easy. Some differences were observed between children/adolescents and adults in the usage of the pen. Many younger participants had less difficulty with the pen than the adults, appearing to remember more from the training sessions and following instructions more carefully. Children/adolescents were less likely to try to perform a task without reading the IFU but were more likely to omit the disinfection step. Some minor differences were also observed between naïve and experienced users, with experienced users generally requiring more assistance than those with no previous experience of injection devices. Participants at both centers generally had a good impression of the new device. In the US, 100% of the children liked the look of the device, and 90% of adults had a good impression of it. Among the German participants, 90% of children and 74% of adults had a good first impression of the device.

Discussion

This usability study met its primary objectives, in that more than 90% of participants were able to successfully execute the injection procedure and safely disassemble the pen. When analyzed separately, both the adult group and the children/adolescents met the predefined acceptance criteria (>85% success rate) for the two primary usability tasks. Despite the training and test sessions being conducted slightly differently, the success rates for the two primary usability tasks were similar in the US and German centers. All of the other tasks/handling steps were also successfully performed by most participants. The high success rates for the usability tasks were reflected in the percentage of participants classifying each one as “very easy” or “easy”. Both naïve and experienced participants found the pen easy to use, a property associated with successful adherence to treatment.19,26–28 Experienced participants generally required more assistance than naive subjects; this may be due to the fact that experienced participants had to “unlearn” how to use their old device before learning about the new system. Participants’ perception of ease of use progressed rapidly in line with experience in use of the new device. An important aspect of this study was the inclusion of caregivers in addition to patients, given that these groups may require/value different features in the device.29 For example, caregivers are more concerned with needlestick injuries than patients.30 Use of the pen was associated with an extremely successful disassembly rate, and only one participant (<1%) received a needlestick injury. Safety during use is an important feature of a growth hormone injection device intended for long-term use.18 While the study met its primary objectives, establishing the validity of this new pen device for the administration of rhGH, several potential limitations do exist; the injection was into a pad rather than a patient, and the study design was open-label and uncontrolled. In conclusion, the new pen device is safe and easy to use for both adults and children, and will help to support effective, long-term daily administration of the rhGH product, Omnitrope®.
  28 in total

1.  Consensus guidelines for the diagnosis and treatment of growth hormone (GH) deficiency in childhood and adolescence: summary statement of the GH Research Society. GH Research Society.

Authors: 
Journal:  J Clin Endocrinol Metab       Date:  2000-11       Impact factor: 5.958

2.  Management of the child born small for gestational age through to adulthood: a consensus statement of the International Societies of Pediatric Endocrinology and the Growth Hormone Research Society.

Authors:  P E Clayton; S Cianfarani; P Czernichow; G Johannsson; R Rapaport; A Rogol
Journal:  J Clin Endocrinol Metab       Date:  2007-01-02       Impact factor: 5.958

3.  Care of girls and women with Turner syndrome: a guideline of the Turner Syndrome Study Group.

Authors:  Carolyn A Bondy
Journal:  J Clin Endocrinol Metab       Date:  2006-10-17       Impact factor: 5.958

4.  Compliance with growth hormone treatment--are they getting it?

Authors:  S L Smith; P C Hindmarsh; C G Brook
Journal:  Arch Dis Child       Date:  1993-01       Impact factor: 3.791

Review 5.  Human growth hormone treatment: synthesis of literature on product delivery systems and administration practices.

Authors:  Monali Bhosle; David Klingman; Mark Aagren; Tami Wisniewski; Won Chan Lee
Journal:  J Spec Pediatr Nurs       Date:  2011-01       Impact factor: 1.260

6.  Multicentre survey on compliance with growth hormone therapy: what can be improved?

Authors:  M Oyarzabal; M Aliaga; M Chueca; G Echarte; A Ulied
Journal:  Acta Paediatr       Date:  1998-04       Impact factor: 2.299

7.  Risk of needlestick injuries by injection pens.

Authors:  G Pellissier; B Miguéres; A Tarantola; D Abiteboul; I Lolom; E Bouvet
Journal:  J Hosp Infect       Date:  2006-03-15       Impact factor: 3.926

8.  Treatment for 24 months with recombinant human GH has a beneficial effect on bone mineral density in young adults with childhood-onset GH deficiency.

Authors:  G S Conway; M Szarras-Czapnik; K Racz; A Keller; P Chanson; M Tauber; M Zacharin
Journal:  Eur J Endocrinol       Date:  2009-03-26       Impact factor: 6.664

9.  Compliance and persistence in pediatric and adult patients receiving growth hormone therapy.

Authors:  Ron G Rosenfeld; Bert Bakker
Journal:  Endocr Pract       Date:  2008-03       Impact factor: 3.443

10.  Patient Acceptance, Ease of Use, and Preference for Norditropin NordiFlex with NordiFlex PenMate: Results from an Open-Label, User Survey of Everyday Use.

Authors:  Anita Hokken-Koelega; Alexandra Keller; Viatcheslav Rakov; Stefan Kipper; Jovanna Dahlgren
Journal:  ISRN Endocrinol       Date:  2011-09-20
View more
  8 in total

1.  Patients' perceptions on the usability of the SurePal™ self-injection device for Omnitrope®: a questionnaire-based observational study conducted in paediatric patients in France.

Authors:  Régis Coutant; Clémentine Dupuis; Patricia Pigeon; Phillipe Rebaud
Journal:  Ther Adv Endocrinol Metab       Date:  2017-10-02       Impact factor: 3.565

Review 2.  Ten years of biosimilar recombinant human growth hormone in Europe.

Authors:  Paul Saenger
Journal:  Drug Des Devel Ther       Date:  2017-05-16       Impact factor: 4.162

3.  Formative usability evaluation of a fixed-dose pen-injector platform device.

Authors:  Jakob Lange; Tobias Nemeth
Journal:  Med Devices (Auckl)       Date:  2018-04-04

4.  Acceptability of the reusable SurePal™ self-injection device for Omnitrope(®) among pediatric patients: results from a questionnaire-based, cross-sectional, multicenter observational study.

Authors:  Carl-Joachim Partsch; Dirk Schnabel; Sarah Ehtisham; Helen C Johnstone; Markus Zabransky; Wieland Kiess
Journal:  Med Devices (Auckl)       Date:  2015-09-15

5.  Results from an international multicenter trial evaluating the ease-of-use of and preference for a newly developed disposable injection pen for the treatment of growth hormone deficiency in treatment-naïve children and adults.

Authors:  Andreas M Pleil; Feyza Darendeliler; Helmuth G Dörr; Katherine Hutchinson; Hartmut A Wollmann
Journal:  Med Devices (Auckl)       Date:  2014-04-11

6.  Usability of devices for self-injection: results of a formative study on a new disposable pen injector.

Authors:  Jakob Lange; Philipp Richard; Nick Bradley
Journal:  Med Devices (Auckl)       Date:  2014-06-12

7.  Acceptance of a reusable self-injection device for recombinant human growth hormone: final data from a questionnaire-based, cross-sectional, international, multicenter, observational study in pediatric patients.

Authors:  Dirk Schnabel; Carl-Joachim Partsch; Muriel Houang; Sarah Ehtisham; Helen Johnstone; Markus Zabransky; Wieland Kiess
Journal:  Med Devices (Auckl)       Date:  2016-09-07

8.  Ease of Use, Preference, and Safety of the Recombinant Human Growth Hormone Disposable Pen Compared with the Reusable Device: A Multicenter, Single-Arm, Open-Label, Switch-Over, Prospective, Phase IV Trial.

Authors:  Ji-Eun Lee; Se Young Kim; Jae-Ho Yoo; Il Tae Hwang; Jung Sub Lim; Kyung Hee Yi; Young-Jun Rhie; Gyung Min Lee; Hyo-Kyoung Nam; Hyun Wook Chae; Eun Young Kim; Chong Kun Cheon; Jieun Lee; Young Suk Shim; YuJin Lee; Eun Young Kim; Jin Soon Hwang
Journal:  Patient Prefer Adherence       Date:  2019-12-20       Impact factor: 2.711

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.