Accuracy of acoustic respiration rate monitoring in pediatric patients.

Mario Patino1, Daniel T Redford, Thomas W Quigley, Mohamed Mahmoud, C Dean Kurth, Peter Szmuk.   

Abstract

BACKGROUND: Rainbow acoustic monitoring (RRa) utilizes acoustic technology to continuously and noninvasively determine respiratory rate from an adhesive sensor located on the neck.
OBJECTIVE: We sought to validate the accuracy of RRa, by comparing it to capnography, impedance pneumography, and to a reference method of counting breaths in postsurgical children.
METHODS: Continuous respiration rate data were recorded from RRa and capnography. In a subset of patients, intermittent respiration rate from thoracic impedance pneumography was also recorded. The reference method, counted respiratory rate by the retrospective analysis of the RRa, and capnographic waveforms while listening to recorded breath sounds were used to compare respiration rate of both capnography and RRa. Bias, precision, and limits of agreement of RRa compared with capnography and RRa and capnography compared with the reference method were calculated. Tolerance and reliability to the acoustic sensor and nasal cannula were also assessed.
RESULTS: Thirty-nine of 40 patients (97.5%) demonstrated good tolerance of the acoustic sensor, whereas 25 of 40 patients (62.5%) demonstrated good tolerance of the nasal cannula. Intermittent thoracic impedance produced erroneous respiratory rates (>50 b·min(-1) from the other methods) on 47% of occasions. The bias ± SD and limits of agreement were -0.30 ± 3.5 b·min(-1) and -7.3 to 6.6 b·min(-1) for RRa compared with capnography; -0.1 ± 2.5 b·min(-1) and -5.0 to 5.0 b·min(-1) for RRa compared with the reference method; and 0.2 ± 3.4 b·min(-1) and -6.8 to 6.7 b·min(-1) for capnography compared with the reference method.
CONCLUSIONS: When compared to nasal capnography, RRa showed good agreement and similar accuracy and precision but was better tolerated in postsurgical pediatric patients.
© 2013 John Wiley & Sons Ltd.

Entities:  

Keywords:  MeSH; apnea; capnography; monitoring physiologic; oximetry; postanesthesia nursing; respiration rate

Mesh:

Year:  2013        PMID: 24033591     DOI: 10.1111/pan.12254

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  12 in total

1.  Long-term tolerability of capnography and respiratory inductance plethysmography for respiratory monitoring in pediatric patients treated with patient-controlled analgesia.

Authors:  Karen M Miller; Andrew Y Kim; Myron Yaster; Sapna R Kudchadkar; Elizabeth White; James Fackler; Constance L Monitto
Journal:  Paediatr Anaesth       Date:  2015-06-03       Impact factor: 2.556

2.  Characterizing soundscapes across diverse ecosystems using a universal acoustic feature set.

Authors:  Sarab S Sethi; Nick S Jones; Ben D Fulcher; Lorenzo Picinali; Dena Jane Clink; Holger Klinck; C David L Orme; Peter H Wrege; Robert M Ewers
Journal:  Proc Natl Acad Sci U S A       Date:  2020-07-07       Impact factor: 11.205

3.  Comparison of Nellcor™ PM1000N and Masimo Radical-7® for detecting apnea in volunteers.

Authors:  Chanatthee Kitsiripant; Tomoko Fukada; Hiroko Iwakiri; Yuri Tsuchiya; Makoto Ozaki; Minoru Nomura
Journal:  J Anesth       Date:  2017-07-20       Impact factor: 2.078

4.  Acoustic method respiratory rate monitoring is useful in patients under intravenous anesthesia.

Authors:  Kentaro Ouchi; Shigeki Fujiwara; Kazuna Sugiyama
Journal:  J Clin Monit Comput       Date:  2016-01-12       Impact factor: 2.502

5.  Listening to breathing again!

Authors:  Michael Ramsay
Journal:  J Clin Monit Comput       Date:  2014-12       Impact factor: 2.502

6.  Influences of environmental noise level and respiration rate on the accuracy of acoustic respiration rate monitoring.

Authors:  Shizuha Yabuki; Hiroaki Toyama; Yusuke Takei; Toshihiro Wagatsuma; Hiroshi Yabuki; Masanori Yamauchi
Journal:  J Clin Monit Comput       Date:  2017-02-07       Impact factor: 2.502

7.  Evaluation of a wireless, portable, wearable multi-parameter vital signs monitor in hospitalized neurological and neurosurgical patients.

Authors:  Robert S Weller; Kristina L Foard; Timothy N Harwood
Journal:  J Clin Monit Comput       Date:  2017-12-06       Impact factor: 2.502

8.  Respiration pattern recognition by wearable mask device.

Authors:  Vishal Varun Tipparaju; Di Wang; Jingjing Yu; Fang Chen; Francis Tsow; Erica Forzani; Nongjian Tao; Xiaojun Xian
Journal:  Biosens Bioelectron       Date:  2020-09-03       Impact factor: 10.618

9.  Electrical Bioimpedance Spectroscopy on Acute Unilateral Stroke Patients: Initial Observations regarding Differences between Sides.

Authors:  Fernando Seoane; Seyed Reza Atefi; Jens Tomner; Konstantinos Kostulas; Kaj Lindecrantz
Journal:  Biomed Res Int       Date:  2015-10-18       Impact factor: 3.411

10.  Estimating respiratory and heart rates from the correntropy spectral density of the photoplethysmogram.

Authors:  Ainara Garde; Walter Karlen; J Mark Ansermino; Guy A Dumont
Journal:  PLoS One       Date:  2014-01-22       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.